Literature DB >> 12406379

Surface analysis of ProFile instruments by scanning electron microscopy and X-ray energy-dispersive spectroscopy: a preliminary study.

R C Martins1, M G A Bahia, V T L Buono.   

Abstract

AIM: The purpose of this preliminary study was to analyze surface irregularities on ProFile nickel-titanium (NiTi) rotary instruments before use, after sterilization by two different methods and after instrumentation of molar root canals.
METHODOLOGY: Five new ProFile NiTi rotary instruments with 0.04 taper, size 20, five instruments with 0.04 taper, size 25 and five with 0.06 taper, size 20 were analyzed by scanning electron microscopy (SEM) and X-ray energy-dispersive spectroscopy (EDS). The instruments were then divided into two groups: three of each taper and size were sterilized in a dry heat sterilizer, whilst two of each taper and size were sterilized in an autoclave. After inspection by the same techniques, the files were used to instrument six molar root canals, and again analyzed by SEM/EDS.
RESULTS: Together with the usual machining defects, a variable amount of material containing mainly carbon and sulphur was found on the surface of new instruments. This material was not removed by standard cleaning and sterilization procedures. The use of the files for instrumentation of root canals eliminated a considerable amount of the adhered substance, but deposits rich in calcium and phosphorus accumulated on the remaining impurities.
CONCLUSIONS: The adherence of dentine on surface deposits found on new ProFile instruments requires special cleaning, prior to sterilization.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12406379     DOI: 10.1046/j.1365-2591.2002.00583.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  7 in total

1.  The effect of 5% sodium hypochlorite, 17% EDTA and triphala on two different rotary Ni-Ti instruments: An AFM and EDS analysis.

Authors:  Pramod Siva Prasad; Jonathan Emi Sam; Arvind Kumar
Journal:  J Conserv Dent       Date:  2014-09

2.  Analysis of Surface Characteristics of ProTaper Universal and ProTaper Next Instruments by Scanning Electron Microscopy.

Authors:  Jeffery Bennett; Kwok-Hung Chung; Hanson Fong; James Johnson; Avina Paranjpe
Journal:  J Clin Exp Dent       Date:  2017-07-01

3.  Geometric and dimensional characteristics of simulated curved canals prepared with proTaper instruments.

Authors:  Renata de Castro Martins; Maria Guiomar de Azevedo Bahia; Vicente Tadeu Lopes Buono
Journal:  J Appl Oral Sci       Date:  2010 Jan-Feb       Impact factor: 2.698

4.  New heat-treated vs electropolished nickel-titanium instruments used in root canal treatment: Influence of autoclave sterilization on surface roughness.

Authors:  Rahaf A Almohareb; Reem Barakat; Fatimah Albohairy
Journal:  PLoS One       Date:  2022-03-18       Impact factor: 3.240

5.  Comparative Evaluation of Physical Surface Changes and Incidence of Separation in Rotary Nickel-Titanium Instruments: An in Vitro SEM Study.

Authors:  Rudra Kaul; Riyaz Farooq; Vibhuti Kaul; Shafayat Ullah Khateeb; Aamir Rashid Purra; Roopali Mahajan
Journal:  Iran Endod J       Date:  2014-07-05

6.  Effect of Autoclave Cycles on Surface Characteristics of S-File Evaluated by Scanning Electron Microscopy.

Authors:  Hamid Razavian; Pedram Iranmanesh; Hamid Mojtahedi; Rahman Nazeri
Journal:  Iran Endod J       Date:  2015-12-24

Review 7.  Metallurgical Tests in Endodontics: A Narrative Review.

Authors:  Alessio Zanza; Marco Seracchiani; Rodolfo Reda; Gabriele Miccoli; Luca Testarelli; Dario Di Nardo
Journal:  Bioengineering (Basel)       Date:  2022-01-12
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.