Literature DB >> 20391948

Dental burs and endodontic files: are routine sterilization procedures effective?

Archie Morrison1, Susan Conrod.   

Abstract

PURPOSE: The complex miniature architecture of dental burs and endodontic files makes precleaning and sterilization difficult. Devising a sterilization protocol for endodontic files and dental burs requires care, and some have suggested that these instruments be considered single-use devices. One purpose of this study was to determine the effective-ness of various sterilization techniques currently used in dentistry for the resterilization of dental burs and endodontic files. The second aim was to determine whether new dental burs and endodontic files, as supplied in packages from the manufacturer, are sterile.
MATERIALS AND METHODS: The sterility of new (unused) and used dental burs and endodontic files before and after various sterilization procedures was analyzed. New burs and files were tested immediately after removal from manufacturers' packaging, with or without prior sterilization. Burs and files that had been used in various dental offices were precleaned, packaged, resterilized and then tested for various pathogens. Each item was individually removed from the sterilization packaging, transferred by sterile technique into Todd-Hewitt broth, incubated at 37degreesC for 72 hours and observed for bacterial growth.
RESULTS: Sterilization procedures were 100 percent effective for unused burs and unused files but were less than 100 percent effective for all other test groups. Contamination rates following sterilization ranged from 15 percent for one group of used burs (p = 0.01) to 58 percent for one group of used files (p <0.001).
CONCLUSIONS: Dental burs and endodontic files, as packaged by the manufacturer, are not sterile and should therefore be sterilized before first use. The resterilization procedures tested here were not adequate, and more rigorous sterilization procedures are needed. If such procedures cannot be devised, these instruments should perhaps be considered single-use devices.

Entities:  

Year:  2010        PMID: 20391948

Source DB:  PubMed          Journal:  Tex Dent J        ISSN: 0040-4284


  3 in total

1.  Efficacy of Different Techniques for Removing Debris from Endodontic Files Prior to Sterilization.

Authors:  Saeid Nosouhian; Farshad Bajoghli; Mahmoud Sabouhi; Masoud Barati; Amin Davoudi; Maryam Sharifipour
Journal:  J Int Oral Health       Date:  2015-08

2.  Effect of Autoclave Cycles on Surface Characteristics of S-File Evaluated by Scanning Electron Microscopy.

Authors:  Hamid Razavian; Pedram Iranmanesh; Hamid Mojtahedi; Rahman Nazeri
Journal:  Iran Endod J       Date:  2015-12-24

3.  Should Surgical Burs Be Used as Single-Use Devices to Avoid Cross Infection? A Case-Control Study.

Authors:  Badr A Al-Jandan; Mohamed Gafar Ahmed; Khalifa Sulaiman Al-Khalifa; Imran Farooq
Journal:  Med Princ Pract       Date:  2015-11-04       Impact factor: 1.927

  3 in total