Literature DB >> 15497355

Microbiological evaluation of endodontic files after cleaning and steam sterilization procedures.

D A Van Eldik1, P S Zilm, A H Rogers, P D Marin.   

Abstract

BACKGROUND: Infection control procedures are essential for modern dental practice and they are continually evolving to meet the dental profession's high standards. The present study evaluated the efficacy of two cleaning procedures to reduce bacterial numbers on endodontic files, and evaluated the effect of biological debris on the subsequent sterilization of files.
METHODS: Stainless steel and nickel-titanium (NiTi) files were examined upon removal from the manufacturer's packaging, after instrumentation in root canals of human teeth inoculated with a broth containing two anaerobic species and one facultative anaerobic species of bacteria, and after instrumentation and cleaning with either an ultrasonic bath or a thermal disinfector. For each file, the bacterial numbers were quantified using routine microbiological techniques in an anaerobic chamber.
RESULTS: No bacteria were detected from files direct from their packets. The size, taper and type of file did not affect the ability of either of the cleaning procedures to reduce bacterial numbers. However, an absence of bacteria was more likely when files were cleaned in the thermal disinfector. No bacteria were detected from files that were-subjected to steam sterilization irrespective of the type of prior cleaning procedure.
CONCLUSIONS: Steam sterilization eliminated all bacteria from the endodontic files irrespective of the presence of biological debris. The majority of bacteria were eliminated from endodontic files after either ultrasonic cleaning or using a thermal disinfector.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15497355     DOI: 10.1111/j.1834-7819.2004.tb00060.x

Source DB:  PubMed          Journal:  Aust Dent J        ISSN: 0045-0421            Impact factor:   2.291


  7 in total

1.  Representative survey on the reprocessing of endodontic instruments in Germany.

Authors:  D Sonntag; E Martin; W H M Raab
Journal:  Br Dent J       Date:  2016-05-13       Impact factor: 1.626

2.  Microbial Culture and Scanning Electron Microscopic Evaluation of Endodontic Hand Files: An In Vitro Study.

Authors:  Ahmad H Almehmadi; Faisal T Alghamdi
Journal:  Cureus       Date:  2022-06-05

3.  Efficacy of Different Techniques for Removing Debris from Endodontic Files Prior to Sterilization.

Authors:  Saeid Nosouhian; Farshad Bajoghli; Mahmoud Sabouhi; Masoud Barati; Amin Davoudi; Maryam Sharifipour
Journal:  J Int Oral Health       Date:  2015-08

4.  Assessment of the Sterility of New Endodontic Files Received From the Manufacturer Using Microbial Culture and Scanning Electron Microscopic Analysis: An In Vitro Study.

Authors:  Khalid A Merdad; Faisal T Alghamdi
Journal:  Cureus       Date:  2022-08-17

5.  Effect of Autoclave Cycles on Surface Characteristics of S-File Evaluated by Scanning Electron Microscopy.

Authors:  Hamid Razavian; Pedram Iranmanesh; Hamid Mojtahedi; Rahman Nazeri
Journal:  Iran Endod J       Date:  2015-12-24

6.  Evaluation of new technique of sterilization using biological indicator.

Authors:  Nomal Chintan Sheth; Yogesh V Rathod; Pratima R Shenoi; Deepa D Shori; Rajiv T Khode; Amruta P Khadse
Journal:  J Conserv Dent       Date:  2017 Sep-Oct

Review 7.  Management of Instrument Sterilization Workflow in Endodontics: A Systematic Review and Meta-Analysis.

Authors:  Mario Dioguardi; Diego Sovereto; Gaetano Illuzzi; Enrica Laneve; Bruna Raddato; Claudia Arena; Vito Carlo Alberto Caponio; Giorgia Apollonia Caloro; Khrystyna Zhurakivska; Giuseppe Troiano; Lorenzo Lo Muzio
Journal:  Int J Dent       Date:  2020-02-08
  7 in total