A SEM evaluation of debris removal from endodontic files after cleaning and steam sterilization procedures.

D A Van Eldik1, P S Zilm, A H Rogers, P D Marin.   

Abstract

BACKGROUND: In recent times, it has been proposed to classify endodontic files as single-use items due to a perceived inability to adequately clean the instruments. The purpose of the present study was to quantify the surface debris on files removed from the manufacturer's packaging, and after cleaning using an ultrasonic bath or a thermal disinfector.
METHODS: Stainless steel and rotary nickel-titanium files were examined after removal from the manufacturer's packaging, after instrumentation in broth-contaminated human teeth, and after various cleaning procedures. The cleaning procedures consisted of either a thermal disinfector cycle, ultrasonication with the files placed in a perforated container or ultrasonication with the files loosely placed in a beaker. The presence of manufacturing debris and biological debris was evaluated using scanning electron microscopy and quantified using image analysis software.
RESULTS: The effectiveness of cleaning was not affected by variation in the size or taper of the files when an effective cleaning procedure was used. Cleaning the files in a thermal disinfector or by ultrasonication within a container did not consistently achieve complete removal of biological debris. Placing the files loosely in the ultrasonic bath achieved the most effective cleaning, an average of 98.33 per cent of the file surface area was freed of any biological debris.
CONCLUSIONS: A conventional cleaning method is capable of effectively removing biological debris from endodontic files. The efficacy of ultrasonic cleaning was impaired when the files were placed within a perforated container.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15497356     DOI: 10.1111/j.1834-7819.2004.tb00061.x

Source DB:  PubMed          Journal:  Aust Dent J        ISSN: 0045-0421            Impact factor:   2.291


  6 in total

1.  Representative survey on the reprocessing of endodontic instruments in Germany.

Authors:  D Sonntag; E Martin; W H M Raab
Journal:  Br Dent J       Date:  2016-05-13       Impact factor: 1.626

2.  Regenerating titanium ventricular assist device surfaces after gold/palladium coating for scanning electron microscopy.

Authors:  Hardean E Achneck; Michael J Serpe; Ryan M Jamiolkowski; Leslie M Eibest; Stephen L Craig; Jeffrey H Lawson
Journal:  Microsc Res Tech       Date:  2010-01       Impact factor: 2.769

3.  Debris, roughness and friction of stainless steel archwires following clinical use.

Authors:  Isabella Silva Vieira Marques; Adriana M Araújo; Júlio A Gurgel; David Normando
Journal:  Angle Orthod       Date:  2010-05       Impact factor: 2.079

4.  Efficacy of Different Techniques for Removing Debris from Endodontic Files Prior to Sterilization.

Authors:  Saeid Nosouhian; Farshad Bajoghli; Mahmoud Sabouhi; Masoud Barati; Amin Davoudi; Maryam Sharifipour
Journal:  J Int Oral Health       Date:  2015-08

5.  Effect of Autoclave Cycles on Surface Characteristics of S-File Evaluated by Scanning Electron Microscopy.

Authors:  Hamid Razavian; Pedram Iranmanesh; Hamid Mojtahedi; Rahman Nazeri
Journal:  Iran Endod J       Date:  2015-12-24

Review 6.  Management of Instrument Sterilization Workflow in Endodontics: A Systematic Review and Meta-Analysis.

Authors:  Mario Dioguardi; Diego Sovereto; Gaetano Illuzzi; Enrica Laneve; Bruna Raddato; Claudia Arena; Vito Carlo Alberto Caponio; Giorgia Apollonia Caloro; Khrystyna Zhurakivska; Giuseppe Troiano; Lorenzo Lo Muzio
Journal:  Int J Dent       Date:  2020-02-08
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.