Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques.

Sarabjeet Singh1, Mannudeep K Kalra, Jiang Hsieh, Paul E Licato, Synho Do, Homer H Pien, Michael A Blake.   

Abstract

PURPOSE: To compare image quality and lesion conspicuity on abdominal computed tomographic (CT) images acquired with different x-ray tube current-time products (50-200 mAs) and reconstructed with adaptive statistical iterative reconstruction (ASIR) and filtered back projection (FBP) techniques.
MATERIALS AND METHODS: Twenty-two patients (mean age, 60.1 years ± 7.3 [standard deviation]; age range, 52.8-67.4 years; mean weight, 78.9 kg ± 18.3; 12 men, 10 women) gave informed consent for this prospective institutional review board-approved and HIPAA-compliant study, which involved the acquisition of four additional image series at multidetector CT. Images were acquired at different tube current-time products (200, 150, 100, and 50 mAs) and encompassed an abdominal lesion over a 10-cm scan length. Images were reconstructed separately with FBP and with three levels of ASIR-FBP blending. Two radiologists reviewed FBP and ASIR images for image quality in a blinded and randomized manner. Volume CT dose index (CTDI(vol)), dose-length product, patient weight, objective noise, and CT numbers were recorded. Data were analyzed by using analysis of variance and the Wilcoxon signed rank test.
RESULTS: CTDI(vol) values were 16.8, 12.6, 8.4, and 4.2 mGy for 200, 150, 100, and 50 mAs, respectively (P < .001). Subjective noise was graded as below average at 150 mAs and average at 100 and 50 mAs for ASIR images, as compared with FBP images, on which noise was graded as average at 150 mAs, above average at 100 mAs, and unacceptable at 50 mAs. A substantial blotchy image appearance was noted in four of 22 image series acquired at 4.2 mGy with 70% ASIR. Lesion conspicuity was significantly better at 4.2 mGy on ASIR than on FBP images (observed P < .044), and overall diagnostic confidence changed from unacceptable on FBP to acceptable on ASIR images.
CONCLUSION: ASIR lowers noise and improves diagnostic confidence in and conspicuity of subtle abdominal lesions at 8.4 mGy when images are reconstructed with 30% ASIR blending and at 4.2 mGy in patients weighing 90 kg or less when images are reconstructed with 50% or 70% ASIR blending. © RSNA, 2010.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20829535     DOI: 10.1148/radiol.10092212

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  130 in total

1.  Iterative reconstruction in head CT: image quality of routine and low-dose protocols in comparison with standard filtered back-projection.

Authors:  A Korn; M Fenchel; B Bender; S Danz; T K Hauser; D Ketelsen; T Flohr; C D Claussen; M Heuschmid; U Ernemann; H Brodoefel
Journal:  AJNR Am J Neuroradiol       Date:  2011-10-27       Impact factor: 3.825

2.  Current status of low dose multi-detector CT in the urinary tract.

Authors:  Mi Kim Sung; Sarabjeet Singh; Mannudeep K Kalra
Journal:  World J Radiol       Date:  2011-11-28

3.  Impact of iterative reconstruction on image quality and radiation dose in multidetector CT of large body size adults.

Authors:  Gaurav S Desai; Raul N Uppot; Elaine W Yu; Avinash R Kambadakone; Dushyant V Sahani
Journal:  Eur Radiol       Date:  2012-04-19       Impact factor: 5.315

4.  Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT.

Authors:  Cynthia H McCollough; Guang Hong Chen; Willi Kalender; Shuai Leng; Ehsan Samei; Katsuyuki Taguchi; Ge Wang; Lifeng Yu; Roderic I Pettigrew
Journal:  Radiology       Date:  2012-06-12       Impact factor: 11.105

5.  Knowledge-based iterative model reconstruction: comparative image quality and radiation dose with a pediatric computed tomography phantom.

Authors:  Young Jin Ryu; Young Hun Choi; Jung-Eun Cheon; Seongmin Ha; Woo Sun Kim; In-One Kim
Journal:  Pediatr Radiol       Date:  2015-11-06

6.  A Feasibility Study of Extracting Tissue Textures From a Previous Full-Dose CT Database as Prior Knowledge for Bayesian Reconstruction of Current Low-Dose CT Images.

Authors:  Yongfeng Gao; Zhengrong Liang; William Moore; Hao Zhang; Marc J Pomeroy; John A Ferretti; Thomas V Bilfinger; Jianhua Ma; Hongbing Lu
Journal:  IEEE Trans Med Imaging       Date:  2019-01-03       Impact factor: 10.048

7.  Optimization-based image reconstruction with artifact reduction in C-arm CBCT.

Authors:  Dan Xia; David A Langan; Stephen B Solomon; Zheng Zhang; Buxin Chen; Hao Lai; Emil Y Sidky; Xiaochuan Pan
Journal:  Phys Med Biol       Date:  2016-10-03       Impact factor: 3.609

8.  Repeated head CT in the neurosurgical intensive care unit: feasibility of sinogram-affirmed iterative reconstruction-based ultra-low-dose CT for surveillance.

Authors:  I Corcuera-Solano; A H Doshi; A Noor; L N Tanenbaum
Journal:  AJNR Am J Neuroradiol       Date:  2014-02-20       Impact factor: 3.825

9.  Model-based iterative reconstruction: effect on patient radiation dose and image quality in pediatric body CT.

Authors:  Ethan A Smith; Jonathan R Dillman; Mitchell M Goodsitt; Emmanuel G Christodoulou; Nahid Keshavarzi; Peter J Strouse
Journal:  Radiology       Date:  2013-10-29       Impact factor: 11.105

10.  Detection of Colorectal Hepatic Metastases Is Superior at Standard Radiation Dose CT versus Reduced Dose CT.

Authors:  Corey T Jensen; Nicolaus A Wagner-Bartak; Lan N Vu; Xinming Liu; Bharat Raval; David Martinez; Wei Wei; Yuan Cheng; Ehsan Samei; Shiva Gupta
Journal:  Radiology       Date:  2018-11-27       Impact factor: 11.105

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.