Image quality, radiation dose, and diagnostic accuracy of prospectively ECG-triggered high-pitch coronary CT angiography at 70 kVp in a clinical setting: comparison with invasive coronary angiography.

Long Jiang Zhang1, Yining Wang2, U Joseph Schoepf3,4, Felix G Meinel4,5, Richard R Bayer4, Li Qi3, Jian Cao2, Chang Sheng Zhou3, Yan E Zhao3, Xie Li3, Jian Bin Gong6, Zhengyu Jin7, Guang Ming Lu8.   

Abstract

PURPOSE: To investigate image quality, radiation dose, and diagnostic performance of prospectively ECG-triggered high-pitch coronary CT angiography (CCTA) at 70 kVp compared to invasive coronary angiography (ICA) as reference standard.
MATERIALS AND METHODS: Forty-three patients underwent prospectively ECG-triggered high-pitch CCTA at 70 kVp using 30 cc (11 g iodine) contrast medium and ICA. Subjective and objective image quality was evaluated for each CCTA study. CCTA performance for diagnosing ≥50% stenosis was assessed. Results were stratified according to heart rate (HR), body mass index (BMI), Agatston score, and image quality.
RESULTS: At CCTA, 94.3% (500/530) of coronary segments were of diagnostic quality. Using ICA as reference standard, sensitivity and accuracy were 100% and 93.0% on a per-patient basis. Per-vessel and per-segment performances were 92.2% and 89.5%; 79.5% and 88.3%, respectively. No differences were found in diagnostic accuracy between different HR, BMI, and calcification subgroups (all P > 0.05) on a per-patient basis. However, low image quality reduced diagnostic accuracy on a per-patient, per-vessel and per-segment basis (all P < 0.05). The mean effective radiation dose was 0.2 ± 0.0 mSv.
CONCLUSION: Our presented protocol results in an effective radiation dose of 0.2 mSv and high diagnostic accuracy for stenosis detection in a selected, non-obese population. KEY POINTS: Prospectively ECG-triggered high-pitch CCTA at 70 kVp is feasible. This protocol has a high diagnostic accuracy for stenosis detection. The mean effective radiation dose was 0.2 ± 0.0 mSv. Only 30 cc of contrast material is used in this protocol. Low image quality reduced diagnostic accuracy of CCTA.

Entities:  

Keywords:  Coronary CT angiography; High-pitch acquisition; Low contrast medium; Low kilovoltage; Prospective ECG gating

Mesh:

Substances:

Year:  2015        PMID: 26382844     DOI: 10.1007/s00330-015-3868-z

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  33 in total

1.  A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association.

Authors:  W G Austen; J E Edwards; R L Frye; G G Gensini; V L Gott; L S Griffith; D C McGoon; M L Murphy; B B Roe
Journal:  Circulation       Date:  1975-04       Impact factor: 29.690

2.  Combined use of automatic tube potential selection with tube current modulation and iterative reconstruction technique in coronary CT angiography.

Authors:  Young Joo Suh; Young Jin Kim; Sae Rom Hong; Yoo Jin Hong; Hye-Jeong Lee; Jin Hur; Byoung Wook Choi
Journal:  Radiology       Date:  2013-10-28       Impact factor: 11.105

3.  CT coronary angiography at an ultra-low radiation dose (<0.1 mSv): feasible and viable in times of constraint on healthcare costs.

Authors:  Filippo Cademartiri; Erica Maffei; Teresa Arcadi; Onofrio Catalano; Massimo Midiri
Journal:  Eur Radiol       Date:  2013-01-24       Impact factor: 5.315

Review 4.  Radiation risks from cardiovascular imaging tests.

Authors:  Felix G Meinel; John W Nance; Brett S Harris; Carlo N De Cecco; Philip Costello; U Joseph Schoepf
Journal:  Circulation       Date:  2014-07-29       Impact factor: 29.690

Review 5.  Coronary artery computed tomography scanning.

Authors:  Carlo Nicola De Cecco; Felix G Meinel; Salvatore A Chiaramida; Philip Costello; Fabian Bamberg; U Joseph Schoepf
Journal:  Circulation       Date:  2014-03-25       Impact factor: 29.690

6.  Diagnostic accuracy of dual-source CT coronary angiography: The effect of average heart rate, heart rate variability, and calcium score in a clinical perspective.

Authors:  Long-Jiang Zhang; Sheng-Yong Wu; Jing Wang; Ying Lu; Zhuo-Li Zhang; Shi-Sen Jiang; Chang-Sheng Zhou; Guang-Ming Lu
Journal:  Acta Radiol       Date:  2010-09       Impact factor: 1.990

7.  Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom study.

Authors:  Felix G Meinel; Christian Canstein; U Joseph Schoepf; Martin Sedlmaier; Bernhard Schmidt; Brett S Harris; Thomas G Flohr; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2014-05-10       Impact factor: 5.315

8.  Closing in on the K edge: coronary CT angiography at 100, 80, and 70 kV-initial comparison of a second- versus a third-generation dual-source CT system.

Authors:  Mathias Meyer; Holger Haubenreisser; U Joseph Schoepf; Rozemarijn Vliegenthart; Christianne Leidecker; Thomas Allmendinger; Ralf Lehmann; Sonja Sudarski; Martin Borggrefe; Stefan O Schoenberg; Thomas Henzler
Journal:  Radiology       Date:  2014-05-31       Impact factor: 11.105

9.  Contrast material-induced nephrotoxicity and intravenous low-osmolality iodinated contrast material: risk stratification by using estimated glomerular filtration rate.

Authors:  Matthew S Davenport; Shokoufeh Khalatbari; Richard H Cohan; Jonathan R Dillman; James D Myles; James H Ellis
Journal:  Radiology       Date:  2013-04-11       Impact factor: 11.105

10.  Prospectively electrocardiogram-gated high-pitch spiral acquisition mode dual-source CT coronary angiography in patients with high heart rates: comparison with retrospective electrocardiogram-gated spiral acquisition mode.

Authors:  Kai Sun; Rui-Juan Han; Li-Jun Ma; Li-Jun Wang; Li-Gang Li; Jiu-Hong Chen
Journal:  Korean J Radiol       Date:  2012-10-12       Impact factor: 3.500

View more
  21 in total

1.  Optimal boundary detection method and window settings for coronary atherosclerotic plaque volume analysis in coronary computed tomography angiography: comparison with intravascular ultrasound.

Authors:  Ran Heo; Hyung-Bok Park; Byoung Kwon Lee; Sanghoon Shin; Reza Arsanjani; James K Min; Hyuk-Jae Chang
Journal:  Eur Radiol       Date:  2015-12-02       Impact factor: 5.315

2.  Image quality, radiation dose and diagnostic accuracy of 70 kVp whole brain volumetric CT perfusion imaging: a preliminary study.

Authors:  Xiao Kun Fang; Qian Qian Ni; U Joseph Schoepf; Chang Sheng Zhou; Guo Zhong Chen; Song Luo; Stephen R Fuller; Carlo N De Cecco; Long Jiang Zhang; Guang Ming Lu
Journal:  Eur Radiol       Date:  2016-02-06       Impact factor: 5.315

3.  Diagnostic accuracy of low and high tube voltage coronary CT angiography using an X-ray tube potential-tailored contrast medium injection protocol.

Authors:  Moritz H Albrecht; John W Nance; U Joseph Schoepf; Brian E Jacobs; Richard R Bayer; Sheldon E Litwin; Michael A Reynolds; Katharina Otani; Stefanie Mangold; Akos Varga-Szemes; Domenico De Santis; Marwen Eid; Georg Apfaltrer; Christian Tesche; Markus Goeller; Thomas J Vogl; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2017-11-27       Impact factor: 5.315

4.  Feasibility of 320-row area detector CT coronary angiography using 40 mL of contrast material: assessment of image quality and diagnostic accuracy.

Authors:  Rihyeon Kim; Eun-Ah Park; Whal Lee; Jin Wook Chung
Journal:  Eur Radiol       Date:  2016-02-23       Impact factor: 5.315

5.  Image quality and diagnostic value of ultra low-voltage, ultra low-contrast coronary CT angiography.

Authors:  Chong-Fu Jia; Jie Zhong; Xin-Yi Meng; Xi-Xia Sun; Zhi-Qiang Yang; Yu-Jie Zou; Xiang-Yue Wang; Shuang Pan; Da Yin; Zhao-Qian Wang
Journal:  Eur Radiol       Date:  2019-03-19       Impact factor: 5.315

6.  Long-term prognostic performance of low-dose coronary computed tomography angiography with prospective electrocardiogram triggering.

Authors:  Olivier F Clerc; Basil P Kaufmann; Mathias Possner; Riccardo Liga; Jan Vontobel; Fran Mikulicic; Christoph Gräni; Dominik C Benz; Tobias A Fuchs; Julia Stehli; Aju P Pazhenkottil; Oliver Gaemperli; Philipp A Kaufmann; Ronny R Buechel
Journal:  Eur Radiol       Date:  2017-05-12       Impact factor: 5.315

7.  Ultra-low-dose CT coronary angiography using 128-slice dual source CT with low concentration contrast agent: initial experience.

Authors:  Lei Zhao; Aishi Liu; Youmin Guo
Journal:  Jpn J Radiol       Date:  2017-10-19       Impact factor: 2.374

Review 8.  Recent advances in cardiac computed tomography dose reduction strategies: a review of scientific evidence and technical developments.

Authors:  Sandeep S Hedgire; Vinit Baliyan; Brian B Ghoshhajra; Mannudeep K Kalra
Journal:  J Med Imaging (Bellingham)       Date:  2017-08-24

9.  Submillisievert imaging protocol using full reconstruction and advanced patient motion correction in 320-row area detector coronary CT angiography.

Authors:  Yuko Kawaguchi; Shinichiro Fujimoto; Kazuhisa Takamura; Etsuro Kato; Shoko Suda; Rie Matsumori; Makoto Hiki; Kanako K Kumamaru; Hiroyuki Daida
Journal:  Int J Cardiovasc Imaging       Date:  2017-09-12       Impact factor: 2.357

10.  Image quality and radiation dose of coronary CT angiography performed with whole-heart coverage CT scanner with intra-cycle motion correction algorithm in patients with atrial fibrillation.

Authors:  Daniele Andreini; Gianluca Pontone; Saima Mushtaq; Maria Elisabetta Mancini; Edoardo Conte; Marco Guglielmo; Valentina Volpato; Andrea Annoni; Andrea Baggiano; Alberto Formenti; Valentina Ditali; Marco Perchinunno; Cesare Fiorentini; Antonio L Bartorelli; Mauro Pepi
Journal:  Eur Radiol       Date:  2017-11-21       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.