| Literature DB >> 26529592 |
Ji-Hyun Chin1, Wook-Jong Kim1, Jeong-Hyun Choi2, Yun A Han1, Seon-Ok Kim3, Woo-Jong Choi1.
Abstract
BACKGROUND: The FloTrac/Vigileo™ system does not thoroughly reflect variable arterial tones, due to a lack of external calibration. The ability of this system to measure stroke volume and track its changes after fluid administration has not been fully evaluated in patients with the high systemic vascular resistance that can develop during laparoscopic surgery.Entities:
Mesh:
Year: 2015 PMID: 26529592 PMCID: PMC4631474 DOI: 10.1371/journal.pone.0142125
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Demographics and preoperative characteristics.
| Age (years) | 62.8 ± 7.1 (range, 49–63) |
| Body mass index (kg/m2) | 25.2 ± 2.7 (range, 19.7–25.5) |
| Comorbidities | |
| Diabetes mellitus | 4 (9.5%) |
| Hypertension | 17 (40.5%) |
| Congestive heart failure | 0 (0%) |
| Peripheral vascular disease | 0 (0%) |
| Mean blood pressure (mmHg) | 88.4 ± 9.7 |
| Hematocrit (%) | 43.7 ± 2.9 |
| Serum creatinine (mg/dl) | 0.9 ± 0.2 |
| Left ventricular ejection fraction (%) | 61.7 ± 4.0 |
Data are presented as mean ± SD or number (percentage).
Hemodynamic data at different time points.
| T0 | T1 | T2 | |
|---|---|---|---|
| SV-Vigileo (ml) | 66.5 ± 13.7 | 68.8 ± 15.9 | 73.0 ± 14.8 |
| SV-TEE (ml) | 61.2 ± 15.6 | 57.0 ± 11.0 | 64.9 ± 12.2 |
| MABP (mmHg) | 75.8 ± 9.7 | 87.5 ± 11.2 | 86.6 ± 10.9 |
| HR (beats/min) | 67.4 ± 15.1 | 60.1 ± 11.7 | 62.9 ± 11.6 |
| TSVR (dyn·s/cm5) | 1604.5 ± 482.0 | 2159.4 ± 523.5 | 1801.9 ± 475.5 |
| AC (ml/mmHg) | 1.5 ± 0.4 | 1.4 ± 0.4 | 1.6 ± 0.4 |
Data are presented as mean ± SD. T0, after anesthetic induction in supine position; T1, 3 minutes after the steep Trendelenburg position (35°) was added to pneumoperitoneum during which time insufflation pressure was set to 15 mmHg; T2, 3 minutes after 500 ml of colloid infusion in T1; SV-Vigileo, stroke volume derived by the FloTrac/Vigileo™ system; SV-TEE, stroke volume measured using transesophageal echocardiography; MABP, mean arterial blood pressure; HR, heart rate; TSVR, total systemic vascular resistance; AC, arterial compliance.
*P < 0.05 vs T0;
† P < 0.05 vs T1.
Fig 1Bland-Altman plot for the difference between the SV-TEE and the SV-Vigileo during pneumoperitoneum combined with the steep Trendelenburg position.
SV-TEE, stroke volume measured using transesophageal echocardiography; SV-Vigileo, stroke volume derived by the third-generation FloTrac/Vigileo™ system.
Bland-Altman analyses of the SV-Vigileo compared with the SV-TEE.
| T0 | T1 | T2 | |
|---|---|---|---|
| Bias | |||
| Mean (ml) | -5.4 | -11.7 | -8.1 |
| SD (ml) | 13.7 | 16.3 | 16.3 |
| 95% LOA (ml) | -32.3 to 21.5 | -43.6 to 20.2 | -40.0 to 23.9 |
| Lower LOA (ml) | -39.7 to -24.9 | -52.4 to -34.8 | -48.8 to -31.2 |
| Upper LOA (ml) | 14.1 to 28.9 | 11.4 to 29.0 | 15.1 to 32.7 |
| Percentage error (%) | 44.9 | 57.2 | 50.3 |
T0, after anesthetic induction in supine position; T1, 3 minutes after the steep Trendelenburg position (35°) was added to pneumoperitoneum during which time insufflation pressure was set to 15 mmHg; T2, 3 minutes after 500 ml of colloid infusion in T1; SV-Vigileo, stroke volume derived by the FloTrac/Vigileo™ system; SV-TEE, stroke volume measured using transesophageal echocardiography; LOA, limit of agreement.
Fig 2The four-quadrant plot and polar plot analysis.
These figures identify an unreliable ability of the third-generation FloTrac/Vigileo™ system to track changes in the stroke volume after fluid administration in the high systemic vascular resistance state observed during laparoscopic prostatectomy. (A) The four-quadrant plot analysis shows a concordance rate of 69.2%, when applying an exclusion zone of 15%. (B) The polar plot analysis shows a mean polar angle of 20.6° and a radial sector containing 95% of the data points of ±51.5%, when applying an exclusion zone of 10%.
Fig 3Correlation analyses
(A) between total systemic vascular resistance (TSVR) and the differences between the stroke volumes measured using transesophageal echocardiography (SV-TEE) and the third-generation FloTrac/Vigileo™ system (SV-Vigileo) and (B) between arterial compliance the differences between the SV-TEE and the SV-Vigileo.