Anna K Nowinska1, Sławomir J Teper2, Dominika A Janiszewska1, Anita Lyssek-Boron3, Dariusz Dobrowolski1, Robert Koprowski4, Edward Wylegala2.
Abstract
PURPOSE: To compare anterior eye segment measurements and morphology obtained with two optical coherence tomography systems (TD OCT, SS OCT) in eyes with corneal dystrophies (CDs).Entities:
Mesh:
Year: 2015 PMID: 26457303 PMCID: PMC4589615 DOI: 10.1155/2015/805367
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Results of automated and manual central corneal thickness (aCCT, mCCT), anterior chamber depth (ACD), and nasal and temporal trabecular iris angle (nTIA, tTIA) measurements by swept-source optical coherence tomography SS OCT and time-domain optical coherence tomography TD OCT. Values were calculated as mean ± standard deviation (SD) or median and range depending on the sample size (<30 or ≥30). BCVA results were presented as range for all groups. EBMD = epithelial basement membrane dystrophy, TBCD = Thiel-Behnke corneal dystrophy, LCD1 = lattice corneal dystrophy TGFBI type, GCD1 = granular corneal dystrophy type 1, GCD2 = granular corneal dystrophy type 2, MCD = macular corneal dystrophy, and FECD = Fuchs endothelial corneal dystrophy.
| Parameter | OCT device | Control group | Study group | ||||||
|
| |||||||||
| CD type | EBMD | TBCD | LCD | GCD1 | GCD2 | MCD | FECD | ||
| Number of eyes | 50 eyes | 12 eyes | 6 eyes | 15 eyes | 7 eyes | 2 eyes | 23 eyes | 31 eyes | |
|
| |||||||||
| BCVA | 1.0 | 1.0 | 0.3–0.9 | 0.1–0.5 | 0.05–0.8 | 0.5–0.6 | 0.05–0.2 | 0.05–0.4 | |
|
| |||||||||
| aCCT | TD OCT | 548.96 | 545 | 600 | 583 | 550 | 538.5 | 459 | 675.54 |
| ±37.34 | 524–586 | 587–615 | 550–620 | 498–567 | 529–548 | 414–492 | ±42.19 | ||
| SS OCT | 553.96 | 554.5 | 598.5 | 587 | 546 | 528 | 453 | 682.03 | |
| ±31.91 | 518–600 | 578–620 | 546–619 | 480–583 | 518–538 | 419–502 | ±41.38 | ||
|
| |||||||||
| mCCT | TD OCT | 546.94 | 541.5 | 602 | 588 | 555 | 530 | 456 | 676.03 |
| ±31.13 | 517–582 | 587–618 | 528–610 | 490–578 | 520–540 | 417–503 | ±43.13 | ||
| SS OCT | 550.34 | 553.5 | 599.5 | 593 | 550 | 513 | 447 | 675.45 | |
| ±31.13 | 507–580 | 580–625 | 550–621 | 498–576 | 507–519 | 418–508 | ±52.98 | ||
|
| |||||||||
| ACD | TD OCT | 3.0514 | 3.085 | 2.9 | 3.13 | 3.11 | 3.22 | 3.05 | 2.30 |
| ±0.24 | 2.71–3.45 | 2.71–3.15 | 2.71–3.59 | 2.72–3.24 | 3.06–3.39 | 2.61–3.59 | ±0.35 | ||
| SS OCT | 2.9718 | 2.955 | 2.91 | 3.015 | 2.98 | 3.18 | 2.98 | 2.36 | |
| ±0.25 | 2.61–3.38 | 2.61–3.05 | 2.61–3.59 | 2.68–3.05 | 3.01–3.36 | 2.69–3.59 | ±0.46 | ||
|
| |||||||||
| nTIA | TD OCT | 31.44 | 31 | 27 | 30 | 33 | 33 | 31 | 21.38 |
| ±4.98 | 24–39 | 24–32 | 24–47 | 29–38 | 30–36 | 24–41 | ±4.40 | ||
| SS OCT | 33.02 | 33.5 | 26.5 | 33 | 34 | 30.5 | 32 | 19.61 | |
| ±5.67 | 22–42 | 22–39 | 26–47 | 32–42 | 27–34 | 20–42 | ±4.49 | ||
|
| |||||||||
| tTIA | TD OCT | 29.9 | 27.5 | 27 | 29 | 32 | 30 | 30 | 19.38 |
| ±5.81 | 22–42 | 23–34 | 22–39 | 29–41 | 26–34 | 24–37 | ±5.33 | ||
| SS OCT | 32.12 | 31.5 | 30 | 33 | 34 | 32.5 | 31 | 20.29 | |
| ±5.57 | 24–44 | 24–36 | 23–40 | 30–40 | 29–36 | 24–41 | ±6.01 | ||
Bland-Altman plot comparing automated and manual central corneal thickness (aCCT, mCCT), anterior chamber depth (ACD), and nasal and temporal trabecular iris angle (nTIA, tTIA) measurements by swept-source optical coherence tomography SS OCT and time-domain optical coherence tomography TD OCT in control and corneal dystrophies group (CD group). 95% CI:95% confidence interval of the mean difference; SD: standard deviation.
| Control group | CD group | |||||||
|---|---|---|---|---|---|---|---|---|
| Mean difference | 95% CI | SD |
| Mean difference | 95% CI | SD |
| |
| aCCT | −4.4 | −10.47–1.63 | 21.29 | .14 | −4.32 | −8.51–0.13 | 20.35 |
|
| mCCT | −3.4 | −7.7–0.97 | 15.37 | .12 | −3.75 | −8.81–1.31 | 24.58 | .14 |
| ACD | 0.07 | 0.06–0.09 | 0.06 | < | 0.06 | −0.01–0.14 | 0.38 |
|
| nTIA | −1.58 | −2.48–−0.67 | 3.2 |
| 1.8 | 0.14–3.47 | 6.62 |
|
| tTIA | −2.22 | −3.04–−0.39 | 2.91 | < | −1.97 | −2.59–−1.36 | 2.97 | < |
Figure 1The graphic presentation of Bland-Altman plot comparing measurements of TD OCT and SS OCT in control and corneal dystrophies group (CD group). Dash-dot line: 95% CI: 95% confidence interval of the mean difference; dashed line: ±1.95 SD (standard deviation); aCCT: automated central corneal thickness; mCCT: manual central corneal thickness; ACD: anterior chamber depth; nTIA: nasal trabecular iris angle; tTIA: temporal trabecular iris angle.
Figure 2Anterior segment single 0–180° scan of TD OCT with measured results of following parameters: mCCT, ACD, nTIA, tTIA, ATA, angle to angle distance, and CLR, crystalline lens rise. (a) Control group. (b) FECD. Note the difference between mCCT, ACD, nTIA, and tTIA, which is statistically significant (p < .0001).
Statistical difference of anterior eye segment measurements: automated and manual central corneal thickness (aCCT, mCCT), anterior chamber depth (ACD), and nasal and temporal trabecular iris angle (nTIA, tTIA) measurements by swept-source optical coherence tomography SS OCT and time-domain optical coherence tomography TD OCT comparing different corneal dystrophies and normal eyes. EBMD = epithelial basement membrane dystrophy, TBCD = Thiel-Behnke corneal dystrophy, LCD1 = lattice corneal dystrophy TGFBI type, GCD1 = granular corneal dystrophy type 1, GCD2 = granular corneal dystrophy type 2, MCD = macular corneal dystrophy, and FECD = Fuchs endothelial corneal dystrophy.
| Statistical difference | Study group versus normal controls | |||||
|---|---|---|---|---|---|---|
| EBMD | TBCD | LCD | GCD1 | MCD | FECD | |
| aCCT | ||||||
| TD OCT |
|
|
|
|
|
|
|
|
|
|
|
| ||
| SS OCT |
|
|
|
|
|
|
|
|
|
|
|
| ||
| mCCT | ||||||
| TD OCT |
|
|
|
|
|
|
|
|
|
|
|
| ||
| SS OCT |
|
|
|
|
|
|
|
|
|
|
|
| ||
| ACD | ||||||
| TD OCT |
|
|
|
|
|
|
|
|
|
|
|
| ||
| SS OCT |
|
|
|
|
|
|
|
|
|
|
|
| ||
| nTIA | ||||||
| TD OCT |
|
|
|
|
|
|
|
|
|
|
|
| ||
| SS OCT |
|
|
|
|
|
|
|
|
|
|
|
| ||
| tTIA | ||||||
| TD OCT |
|
|
|
|
|
|
|
|
|
|
|
| ||
| SS OCT |
|
|
|
|
|
|
|
|
|
|
|
| ||
Figure 3Comparison of representative TD and SS OCT corneal scans in CD group. There are no differences in corneal deposits visualization between both OCT systems. (a) TBCD: SS OCT pachymetry map showing the increase of CCT. aCCT of 612 μm. TD OCT high-resolution corneal scan and SS OCT cornea scan showing increased reflectivity in the area of Bowman layer and anterior corneal stroma. The deposits are causing the irregularity of the anterior stromal border from the epithelium side. (b) LCD: SS OCT pachymetry map showing the increase of CCT. aCCT of 588 μm. TD OCT high-resolution corneal scan and SS OCT cornea scan presenting diffuse areas of increased reflectivity in the area of Bowman layer and anterior to mid stroma. (c) MCD: TD OCT pachymetry map indicating corneal thinning with aCCT of 464 μm. TD OCT high-resolution corneal scan and SS OCT cornea scan showing general increased reflectivity throughout the corneal stroma. Note the irregularity of the anterior stromal border from the epithelium side and the diffuse areas of hyperreflectivity in Bowman's layer. There is a noticeable flat layer of increased reflectivity in the posterior, peripheral corneal part.