| Literature DB >> 26369914 |
T Ten Cate1, M van Wely2, H Gehlmann2, M Mauti3, C Camaro2, N Reifart4, H Suryapranata2, M J de Boer2.
Abstract
AIMS: The consequences of high radiation dose for patient and staff demand constant improvements in X-ray dose reduction technology. This study assessed non-inferiority of image quality and quantified patient dose reduction in interventional cardiology for an anatomy-specific optimised cine acquisition chain combined with advanced real-time image noise reduction algorithms referred to as 'study cine', compared with conventional angiography.Entities:
Keywords: DAP; Dose reduction; Image processing; Interventional cardiology; Safety
Year: 2015 PMID: 26369914 PMCID: PMC4608924 DOI: 10.1007/s12471-015-0742-1
Source DB: PubMed Journal: Neth Heart J ISSN: 1568-5888 Impact factor: 2.380
Patient and procedure characteristics
| Rated ( | ||
|---|---|---|
| Age (years) | Mean (SD) | 63.4 (10.7) |
| Gender | Female ( | 18 (46.2 %) |
| Male ( | 21 (53.8 %) | |
| BMI (kg/m2) | Mean (SD) | 26.4 (3.3) |
| Median | 26.1 | |
| Procedure indication | Diagnostic | 30 (76.9 %) |
| Intervention | 9 (23.1 %) | |
| Vessels treated | Left main stem | 1 (2.6 %) |
| LAD | 5 (12.8 %) | |
| LCX | 4 (10.3 %) |
SD standard deviation, BMI body mass index.
Fig. 1Summary of image assessment
Fig. 2Summary of image assessment of patients with a BMI > 28 kg/m2
Fig. 3Example of a 55-year-old female (BMI 30.2 kg/m2), current smoker with a medical history of hypertension, hyperlipidaemia and a family history of premature coronary artery disease (CAD). The patient already underwent PCI (stent in LAD and RCA) and was admitted for diagnostic procedure for complaints of progressive angina without symptoms at rest. The top panel, a represents the reference image and b the study image. Cine acquisitions obtained at 15 fps, field of view 20 cm, LAO 45 CRA 20 with (a) reference cine (DAP: 60.40 mGy.cm2/frame), (b) study cine (DAP: 27.01 mGy.cm2/frame)
Radiation characteristics for the complete study population
| Study cine | Reference cine |
| ||
|---|---|---|---|---|
| AirKerma [mGy/frame] |
| 39 | 39 | |
| Mean (SD) | 0.44 (0.1) | 0.95 (0.2) | < 0.001 | |
| Median | 0.44 | 0.94 | < 0.001 | |
| DAP [mGy.cm2/frame] |
| 39 | 39 | |
| Mean (SD) | 26 (5.7) | 55 (6.9) | < 0.001 | |
| Median | 26 | 55 | < 0.001 |
SD standard deviation, DAP dose area product.
Radiation characteristics for patients with BMI > 28 kg/m2
| Study cine | Reference cine |
| ||
|---|---|---|---|---|
| AirKerma [mGy/frame] |
| 10 | 10 | |
| Mean (SD) | 0.5 (0.1) | 0.95 (0.1) | < 0.0001 | |
| Median | 0.48 | 0.97 | ||
| DAP [mGy.cm2/frame] |
| 10 | 10 | |
| Mean (SD) | 31 (5.7) | 57 (10.4) | < 0.0001 | |
| Median | 31 | 59 |