Literature DB >> 16461493

Radiation dose levels during interventional cardiology procedures in a tertiary care hospital.

O Dragusin1, W Desmet, H Heidbuchel, R Padovani, H Bosmans.   

Abstract

The aim of our investigation was to prospectively measure the patient radiation exposure from different cardiological procedures performed in the Catheterisation laboratory of the University Hospital Gasthuisberg in Leuven. The following local reference values were proposed: 40, 47 and 80 Gycm2 for coronary angiography (CA) or angioplasty (PTCA and stent implantation for elective patients), radio frequency ablation with angiographic images and CA plus ad hoc PTCA, respectively.

Entities:  

Mesh:

Year:  2006        PMID: 16461493     DOI: 10.1093/rpd/nci754

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  3 in total

1.  How to set up and apply reference levels in fluoroscopy at a national level.

Authors:  A Aroua; H Rickli; J-C Stauffer; P Schnyder; P R Trueb; J-F Valley; P Vock; F R Verdun
Journal:  Eur Radiol       Date:  2006-10-27       Impact factor: 5.315

2.  Patient radiation doses in interventional cardiology procedures.

Authors:  Ioannis Pantos; Georgios Patatoukas; Demosthenes G Katritsis; Efstathios Efstathopoulos
Journal:  Curr Cardiol Rev       Date:  2009-01

3.  Novel X-ray image noise reduction technology reduces patient radiation dose while maintaining image quality in coronary angiography.

Authors:  T Ten Cate; M van Wely; H Gehlmann; M Mauti; C Camaro; N Reifart; H Suryapranata; M J de Boer
Journal:  Neth Heart J       Date:  2015-11       Impact factor: 2.380

  3 in total