Build-up modification of commercial diodes for entrance dose measurements in 'higher energy' photon beams.

D Georg1, B De Ost, M T Hoornaert, P Pilette, J Van Dam, M Van Dycke, D Huyskens.   

Abstract

BACKGROUND AND
PURPOSE: Several commercially available p-type diodes do not provide sufficient build-up for in-vivo dosimetry in 'higher' energy photon beams, and only limited information could be found in the literature describing the correction factor variation and/or the achievable accuracy for in-vivo dosimetry methods in this energy range. The first aim of this study is to assess and analyze the variation of diode correction factors for entrance dose measurements at higher photon energies. In a second step the total build up thickness of the diode has been modified in order to minimize the correction factor variation.
MATERIALS AND METHODS: Diode correction factors accounting for non-reference conditions (field size, source surface distance, tray, wedge, and block) are determined in 18-25 MV photon beams provided by different treatment units for Scanditronix p-type diodes recommended for higher energy photon beams: old type and new type EDP-20, and EDP-30 diodes. Hemispherical build-up caps of different materials (copper, iron, lead) are used to increase the total build-up thickness. Perturbation effects with and without additional build-up caps are assessed for the three diode types.
RESULTS: For unmodified diodes field size correction factors (C(FS)) vary between 1.7% and 6%, dependent on diode type and treatment unit. For example, for an old type EDP-20 the C(FS) variation at 18 MV is much higher on a GE linac (5%) as compared to the Philips machine (1.7%). Depending on diode type, this variation can be reduced to 1-2% when adding additional build-up. The variation of source to surface distance correction factors is almost independent of build-up thickness. By adding additional build-up the influence of trays and blocks can be almost eliminated.
CONCLUSIONS: The correction factor variation of unmodified diodes reflects the variation of the electron contamination with treatment geometry. A total build-up thickness of 30 mm is found to be the 'best compromise' for the three types of diodes investigated when measuring entrance doses in the energy range between 18 and 25 MV.

Entities:  

Mesh:

Year:  1999        PMID: 10435820     DOI: 10.1016/s0167-8140(99)00058-4

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  4 in total

1.  In vivo dosimetry using a single diode for megavoltage photon beam radiotherapy: implementation and response characterization.

Authors:  V C Colussi; A S Beddar; T J Kinsella; C H Sibata
Journal:  J Appl Clin Med Phys       Date:  2001       Impact factor: 2.102

2.  Patient skin dose measurements using a cable free system MOSFETs based in fluoroscopically guided percutaneous vertebroplasty, percutaneous disc decompression, radiofrequency medial branch neurolysis, and endovascular critical limb ischemia.

Authors:  Maria D Falco; Salvatore Masala; Matteo Stefanini; Roberto Fiori; Roberto Gandini; Paolo Bagalà; Daniele Morosetti; Eros Calabria; Alessia Tonnetti; Gianluca Verona-Rinati; Riccardo Santoni; Giovanni Simonetti
Journal:  J Appl Clin Med Phys       Date:  2015-01-08       Impact factor: 2.102

3.  Percent depth-dose distribution discrepancies from very small volume ion chambers.

Authors:  Vikren Sarkar; Brian Wang; Hui Zhao; Bart Lynch; Joshua A James; Kiernan T McCullough; Bill J Salter
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

4.  Characterization of an in vivo diode dosimetry system for clinical use.

Authors:  Kai Huang; William S Bice; Oscar Hidalgo-Salvatierra
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.