Literature DB >> 17077725

Patient and staff dosimetry in vertebroplasty.

Niki T Fitousi1, Efstathios P Efstathopoulos, Harry B Delis, Sofia Kottou, Alexis D Kelekis, George S Panayiotakis.   

Abstract

STUDY
DESIGN: Eleven vertebroplasty operations were studied in terms of radiation dose.
OBJECTIVE: Doses to patients and staff associated with vertebroplasty were measured. Occupational doses were compared with the annual dose limits, and the effectiveness of the used radiation protection means was estimated. Patient dose was estimated by means of both surface and effective dose, and the radiation-induced risk was evaluated. SUMMARY OF BACKGROUND DATA: Vertebroplasty is a recent minimally invasive technique for the restoration of vertebral body fractures. It involves fluoroscopic exposure, and so, it demands dose measurements for both patient and staff exposed to radiation.
METHODS: Thermoluminescent dosimeters (TLDs) were placed on the medical personnel and the effective dose was derived. Slow films were placed to patients' skin to measure entrance surface dose. Furthermore, a Rando phantom loaded with TLDs was irradiated under conditions simulating vertebroplasty, in order to estimate effective dose to the patient.
RESULTS: Mean fluoroscopy time was 27.7 minutes. Patient's mean skin dose was 688 mGy, while effective dose was calculated to be 34.45 mGy. It was estimated that the primary operator can perform about 150 vertebroplasty operations annually without exceeding the annual dose constraints, whereas occupational dose can be reduced by 76% using mobile shielding.
CONCLUSIONS: Measures have to be taken to reduce patient's skin dose, which, in extreme cases, may be close to deterministic effects threshold. The highest dose rates, recorded during the procedure, were found for primary operator's hands and chest when no shielding was used.

Entities:  

Mesh:

Year:  2006        PMID: 17077725     DOI: 10.1097/01.brs.0000244586.02151.18

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  15 in total

Review 1.  Occupational radiation doses to operators performing fluoroscopically-guided procedures.

Authors:  Kwang Pyo Kim; Donald L Miller; Amy Berrington de Gonzalez; Stephen Balter; Ruth A Kleinerman; Evgenia Ostroumova; Steven L Simon; Martha S Linet
Journal:  Health Phys       Date:  2012-07       Impact factor: 1.316

2.  Is lead shielding of patients necessary during fluoroscopic procedures? A study based on kyphoplasty.

Authors:  Joshua R Smith; Rebecca M Marsh; Michael S Silosky
Journal:  Skeletal Radiol       Date:  2017-08-18       Impact factor: 2.199

3.  Effective-dose estimation in interventional radiological procedures.

Authors:  Maria D Falco; Salvatore Masala; Matteo Stefanini; Paolo Bagalà; Daniele Morosetti; Eros Calabria; Alessia Tonnetti; Gianluca Verona-Rinati
Journal:  Radiol Phys Technol       Date:  2018-03-08

4.  Evaluation of operator radioprotection using a new injection device during vertebroplasty.

Authors:  L Nguyen-Kim; C Fargeot; H Beaussier; S Payen; J Chiras
Journal:  Interv Neuroradiol       Date:  2013-05-21       Impact factor: 1.610

5.  A novel "three-dimensional-printed individual guide template-assisted percutaneous vertebroplasty" for osteoporotic vertebral compression fracture: a prospective, controlled study.

Authors:  Pei Lun Hu; Ji Sheng Lin; Hai Meng; Nan Su; Yong Yang; Qi Fei
Journal:  J Orthop Surg Res       Date:  2021-05-20       Impact factor: 2.359

6.  The effects of low dose X-irradiation on osteoblastic MC3T3-E1 cells in vitro.

Authors:  Wei Xu; Lan Xu; Ming Chen; Yong Tao Mao; Zong Gang Xie; Shi Liang Wu; Qi Rong Dong
Journal:  BMC Musculoskelet Disord       Date:  2012-06-08       Impact factor: 2.362

7.  A retrospective comparison of intraoperative CT and fluoroscopy evaluating radiation exposure in posterior spinal fusions for scoliosis.

Authors:  Jacob Riis; Rebecca R Lehman; Robert A Perera; John Ryan Quinn; Patricia Rinehart; Hans Robert Tuten; Victoria Kuester
Journal:  Patient Saf Surg       Date:  2017-12-21

8.  Dosimetry during intramedullary nailing of the tibia.

Authors:  George Kirousis; Harry Delis; Panagiotis Megas; Elias Lambiris; George Panayiotakis
Journal:  Acta Orthop       Date:  2009-10       Impact factor: 3.717

9.  Comparing radiation exposure during percutaneous vertebroplasty using one- vs. two-fluoroscopic technique.

Authors:  Yen-Yao Li; Tsung-Jen Huang; Chin-Chang Cheng; Meng-Huang Wu; Ching-Yu Lee
Journal:  BMC Musculoskelet Disord       Date:  2013-01-22       Impact factor: 2.362

10.  Significant Improvement of Puncture Accuracy and Fluoroscopy Reduction in Percutaneous Transforaminal Endoscopic Discectomy With Novel Lumbar Location System: Preliminary Report of Prospective Hello Study.

Authors:  Guoxin Fan; Xiaofei Guan; Hailong Zhang; Xinbo Wu; Xin Gu; Guangfei Gu; Yunshan Fan; Shisheng He
Journal:  Medicine (Baltimore)       Date:  2015-12       Impact factor: 1.817

View more