| Literature DB >> 25329463 |
Janaina Oliveira Bentivi Pulcherio1, Aline Gomes Bittencourt2, Patrick Rademaker Burke2, Rafael da Costa Monsanto3, Rubens de Brito2, Robinson Koji Tsuji2, Ricardo Ferreira Bento2.
Abstract
OBJECTIVE: To review the outcomes of the fully implantable middle ear devices Carina and Esteem regarding the treatment of hearing loss. DATA SOURCES: PubMed, Embase, Scielo, and Cochrane Library databases were searched. STUDY SELECTION: Abstracts of 77 citations were screened, and 43 articles were selected for full review. From those, 22 studies and two literature reviews in English directly demonstrating the results of Carina and Esteem were included. DATA EXTRACTION: There were a total of 244 patients ranging from 18 to 88 years. One hundred and 10 patients were implanted with Carina and with 134 Esteem. There were registered 92 males and 67 females. Five studies provided no information about patients' age or gender. From the data available, the follow-up ranged from 2 to 29.4 months. DATA SYNTHESIS: The comparison of the results about word recognition is difficult as there was no standardization of measurement. The results were obtained from various sound intensities and different frequencies. The outcomes comparing to conventional HAs were conflicting. Nevertheless, all results comparing to unaided condition showed improvement and showed a subjective improvement of quality of life.Entities:
Mesh:
Year: 2014 PMID: 25329463 PMCID: PMC4201540 DOI: 10.1371/journal.pone.0110636
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Figure 1Decision process of the selection of the studies included in this review.
Adapted from: Moher D, Liberati A, Tetzlaff J, Altman DG, The PRISMA Group (2009). Preferred Reporting Items for Systematic Reviews and Meta-Analyses: The PRISMA Statement. PLoS Med 6(6): e1000098. Doi:10.1371/journal.pmed1000097.
Demographic characteristics of the studies.
| Study | Year | Patients(n) | MED | Gender | MeanAge | Side | MeanFollow-up(mths) |
|
| 2009 | 6 | Esteem | NR | NR | NR | NR |
|
| 2011 | 18 | Esteem | NR | NR | NR | NR |
|
| 2009 and 2010 | 8 | Carina | 7 M, 1F | 46.6 | 6R, 2L | 16.9 |
|
| 2004 | 7 | Esteem | 2F, 5 M | 64.4 | 4R, 3L | 10 |
|
| 2010 | 1 | Carina | 1F | 63 | 1R | 6 |
|
| 2012 | 13 | Esteem | 9F, 4 M | NR | 9R, 4L | 4 |
|
| 2007 and 2008 | 20 | Carina | 10F, 10 M | 62.8 | 9R, 11L | 2 |
|
| 2012 | 6 | Carina | 3F, 3 M | 48.5 | 4R, 2L | 12 |
|
| 2011 | 57 | Esteem | 19F, 38 M | 52.9 | 35R, 22L | 12 |
|
| 2012 | 1 | Carina | 1F | 18 | 1R, 1L | 9 |
|
| 2009 | 6 | Carina | NR | NR | NR | 12 |
|
| 2009 | 11 | Carina | 7F, 4 M | 50.8 | 5R, 6L | 24 |
|
| 2011 | 10 | Esteem | 7 M, 3F | 32.7 | 6L, 4R | 29.4 |
|
| 2013 | 15 | Esteem | 8F, 7 M | NR | NR | NR |
|
| 2009 | 3 | Esteem | 2 M, 1F | 28,6 | 2R, 1L | NR |
|
| 2011 | 5 | Esteem | NR | NR | NR | 12 |
|
| 2007 | 5 | Carina | 4 M, 1F | 31,4 | NR | 3 |
|
| 2008 | 1 | Carina | 1 M | 14 | 1R | 2 |
|
| 2009 | 1 | Carina | 1F | 48 | 1R | 15 |
|
| 2010 | 50 | Carina | NR | NR | NR | NR |
n – number; MED – middle ear device; mths – months; M – male; F – female; NR – not reported.
Complications related to the fully implantable middle ear devices.
| Study | Year | MED | Patients(n) | AdverseEffects (n) | SurgicalComplications(n) | DeviceAdverseEffects (n) | Revisions(n) |
|
| 2009 | Carina | 5 | NR | None | Feedback (5) | 1 |
|
| 2010 | Carina | 8 | Extrusion of the cable of microphone (1),Psychologicalproblems andexplantation (1) | NR | Feedback (1),Device failure (1) | 2 |
|
| 2004 | Esteem | 7 | Woundcomplication (1) | Temporary swelling ofthe lower eyelid, sorejaw, nausea, diarrhea,elbow pain, arm andhand pain, andnumbness (2) | No benefit (3),Low gain (3) | 3 |
|
| 2012 | Esteem | 13 | Secondaryhealing difficulty(1), recurrenttissue edema (1),S aureus woundinfection (2) | Temporary partialfacial palsy (1),rupture of chordatympani nerve (8), | Poor anddeterioratingfunctionalresults (4) | 4 |
|
| 2007 and 208 | Carina | 20 | fullness orpressure sensation(2), conductivehearing loss (4),lightheadedness(1), tinnitus (1),partial deviceextrusion (3), andmiddle eareffusion (3) | partial deviceextrusion (3),inability tocharge orestablishcommunication(2),increasedcharging timesbeyond 1.5 hours | ||
|
| 2012 | 6 | None | None | feedback | 1 | |
|
| 2011 | Esteem | 57 | Taste disturbance(24), middle eareffusion (18),pain (8), dizziness(9), tinnitus (7),headache (3),infection (2) | Facialparesis/paralysis (4) | Limitedbenefit (4) | 3 |
|
| 2009 | Carina | 11 | Infection (2),vertigo (1),tinnitus (1) | NR | Hearing loss(1), poor soundtransmission (1) | 1 |
|
| 2011 | Esteem | 10 | NR | Temporary facialweakness (1) | Nobenefit (2) | NR |
|
| 2009 | Esteem | 3 | None | transient facialparesis (1) | NR | 0 |
n – number; MED – middle ear device; NR – not reported.