Fully implantable hearing device with transducer on the round window as a treatment of mixed hearing loss.

Stéphane Tringali1, Nick Pergola, Paul Berger, Christian Dubreuil.   

Abstract

OBJECTIVE: To demonstrate the safety and efficacy of the Otologics Carina Middle Ear Transducer for treatment of mixed hearing loss through a case report.
METHODS: A Carina fully implantable device with the MET transducer for conductive applications was implanted in a 48-year-old woman suffering from right mixed hearing loss (mean PTA loss: 80 dB). A facial recess approach was used to access the middle ear. Sclerotic tissue obliterated the stapes footplate so the approach selected was to place the transducer directly on the round window. The mounting bracket was placed on the mastoid and the prosthesis was advanced toward the round window until the Otologics surgical software indicated contact. Effective stimulation of the cochlea was confirmed intraoperatively by ABR monitoring.
RESULTS: Postoperative unaided PTA thresholds were unchanged after surgery. When the implant was activated, the mean PTA functional gain was 39 dB. DISCUSSION-
CONCLUSION: The capability of the Carina MET Ossicular stimulator to provide appropriate gain relative to the degree of hearing loss indicates that the device offers a viable treatment option for mixed hearing loss. However, these promising initial results establish the need for future work on two fronts: (1) further studies are needed including a greater number of patients to confirm these preliminary results; (2) a long term follow-up must be carried out to detect any possible cochlear adverse effects on the cochlea, in particular on the basilar membrane.

Entities:  

Mesh:

Year:  2008        PMID: 19013036     DOI: 10.1016/j.anl.2008.09.003

Source DB:  PubMed          Journal:  Auris Nasus Larynx        ISSN: 0385-8146            Impact factor:   1.863


  13 in total

1.  Controlling the position and the dislocation of the middle ear transducer with high-resolution computed tomography and digital volume tomography: implications for the transducers' design.

Authors:  Georgios Kontorinis; Anja M Giesemann; Thomas Witt; Friedrich Goetz; Burkard Schwab
Journal:  Eur Arch Otorhinolaryngol       Date:  2011-09-22       Impact factor: 2.503

Review 2.  [Implantable hearing aids].

Authors:  J C Luers; D Beutner; K-B Hüttenbrink
Journal:  HNO       Date:  2011-10       Impact factor: 1.284

3.  Round window membrane implantation with an active middle ear implant: a study of the effects on the performance of round window exposure and transducer tip diameter in human cadaveric temporal bones.

Authors:  Stéphane Tringali; Kanthaiah Koka; Arnaud Deveze; N Julian Holland; Herman A Jenkins; Daniel J Tollin
Journal:  Audiol Neurootol       Date:  2010-02-11       Impact factor: 1.854

Review 4.  [Active hearing implants in chronic otitis media].

Authors:  S Lailach; C Müller; N Lasurashvili; H Seidler; T Zahnert
Journal:  HNO       Date:  2021-06       Impact factor: 1.284

5.  The location of the mastoid portion of the facial nerve in patients with congenital aural atresia.

Authors:  Yaoyao Fu; Peidong Dai; Tianyu Zhang
Journal:  Eur Arch Otorhinolaryngol       Date:  2013-06-21       Impact factor: 2.503

6.  A tri-coil bellows-type round window transducer with improved frequency characteristics for middle-ear implants.

Authors:  Dong Ho Shin; Ki Woong Seong; Sunil Puria; Kyu-Yup Lee; Jin-Ho Cho
Journal:  Hear Res       Date:  2016-09-02       Impact factor: 3.208

7.  Dynamic properties of human round window membrane in auditory frequencies running head: dynamic properties of round window membrane.

Authors:  Xiangming Zhang; Rong Z Gan
Journal:  Med Eng Phys       Date:  2012-06-04       Impact factor: 2.242

Review 8.  Limits on normal cochlear 'third' windows provided by previous investigations of additional sound paths into and out of the cat inner ear.

Authors:  John J Rosowski; Peter Bowers; Hideko H Nakajima
Journal:  Hear Res       Date:  2017-11-10       Impact factor: 3.208

9.  The Carina© middle ear implant: surgical and functional outcomes.

Authors:  Luca Bruschini; Stefano Berrettini; Francesca Forli; Alessandra Murri; Domenico Cuda
Journal:  Eur Arch Otorhinolaryngol       Date:  2016-03-23       Impact factor: 2.503

10.  Carina® and Esteem®: a systematic review of fully implantable hearing devices.

Authors:  Janaina Oliveira Bentivi Pulcherio; Aline Gomes Bittencourt; Patrick Rademaker Burke; Rafael da Costa Monsanto; Rubens de Brito; Robinson Koji Tsuji; Ricardo Ferreira Bento
Journal:  PLoS One       Date:  2014-10-17       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.