A Retrospective Multicentre Cohort Review of Patient Characteristics and Surgical Aspects versus the Long-Term Outcomes for Recipients of a Fully Implantable Active Middle Ear Implant.

Philippe P Lefebvre1, Javier Gisbert, Domenico Cuda, Stéphane Tringali, Arnaud Deveze.   

Abstract

OBJECTIVE: To summarise treatment outcomes compared to surgical and patient variables for a multicentre recipient cohort using a fully implantable active middle ear implant for hearing impairment. To describe the authors' preferred surgical technique to determine microphone placement. STUDY
DESIGN: Multicentre retrospective, observational survey.
SETTING: Five tertiary referral centres. PATIENTS: Carina recipients (66 ears, 62 subjects) using the current Cochlear® Carina® System or the legacy device, the Otologics® Fully Implantable Middle Ear, with a T2 transducer.
METHODS: Patient file review and routine clinical review. Patient outcomes assessed were satisfaction, daily use and feedback reports at the first fitting and ≥12 months after implantation. Descriptive and statistical analysis of correlations of variables and their influence on outcomes was performed. Independently reported preferred methods for microphone placement are collectively summarised.
RESULTS: The average implant experience was 3.5 years. Satisfaction increased significantly over time (p < 0.05). No correlation with covariates examined was observed. Feedback significantly decreased over time, showing a significant correlation with microphone location, primary motivation, gender, age at implantation, and contralateral hearing aid use (p < 0.05). Patient satisfaction was inversely correlated with reports of system feedback (p < 0.05). The implantable microphone was most commonly on the posterior inferior mastoid line, in 42/66 (65%) cases, correlating with less likelihood for feedback and consistent with author surgical preference.
CONCLUSION: Carina recipients in this study present as satisfied consistent daily users with very few reports of persistent feedback. As microphone location is an influencing factor, a careful surgical consideration of microphone placement is required. The authors prefer a posterior inferior mastoid line position whenever possible.
© 2017 The Author(s) Published by S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2017        PMID: 28052264      PMCID: PMC5348730          DOI: 10.1159/000454666

Source DB:  PubMed          Journal:  Audiol Neurootol        ISSN: 1420-3030            Impact factor:   1.854


  17 in total

1.  A pilot study of the safety and performance of the Otologics fully implantable hearing device: transducing sounds via the round window membrane to the inner ear.

Authors:  P P Lefebvre; C Martin; C Dubreuil; M Decat; A Yazbeck; J Kasic; S Tringali
Journal:  Audiol Neurootol       Date:  2008-11-13       Impact factor: 1.854

2.  Fully implantable Otologics MET Carina(™) device for the treatment of sensorineural and mixed hearing loss: Audio-otological results.

Authors:  Luca Bruschini; Francesca Forli; Susanna Passetti; Paolo Bruschini; Stefano Berrettini
Journal:  Acta Otolaryngol       Date:  2010-10       Impact factor: 1.494

3.  Simulating the open-loop transfer function as a means for understanding acoustic feedback in hearing aids.

Authors:  D P Egolf; B T Haley; H C Howell; S Legowski; V D Larson
Journal:  J Acoust Soc Am       Date:  1989-01       Impact factor: 1.840

4.  Experience with the active middle ear implant in patients with moderate-to-severe mixed hearing loss: indications and results.

Authors:  Jorge de Abajo; Ignacio Sanhueza; Laura Giron; Manuel Manrique
Journal:  Otol Neurotol       Date:  2013-10       Impact factor: 2.311

5.  Evolution of the reliability of the fully implantable middle ear transducer over successive generations.

Authors:  Mathilde Debeaupte; Evelyne Decullier; Stephane Tringali; Arnaud Devèze; Thierry Mom; Vincent Darrouzet; Eric Truy
Journal:  Otol Neurotol       Date:  2015-04       Impact factor: 2.311

Review 6.  Middle ear implantable hearing devices: an overview.

Authors:  David S Haynes; Jadrien A Young; George B Wanna; Michael E Glasscock
Journal:  Trends Amplif       Date:  2009-09

7.  Comparison of Carina active middle-ear implant with conventional hearing aids for mixed hearing loss.

Authors:  V A Savaş; B Gündüz; R Karamert; R Cevizci; M Düzlü; H Tutar; Y A Bayazit
Journal:  J Laryngol Otol       Date:  2016-04       Impact factor: 1.469

8.  U.S. Phase I preliminary results of use of the Otologics MET Fully-Implantable Ossicular Stimulator.

Authors:  Herman A Jenkins; James S Atkins; Drew Horlbeck; Michael E Hoffer; Ben Balough; Joseph V Arigo; George Alexiades; William Garvis
Journal:  Otolaryngol Head Neck Surg       Date:  2007-08       Impact factor: 3.497

9.  Active middle ear implantation for patients with sensorineural hearing loss and external otitis: long-term outcome in patient satisfaction.

Authors:  Joost W Zwartenkot; Javad Hashemi; Cor W R J Cremers; Jef J S Mulder; Ad F M Snik
Journal:  Otol Neurotol       Date:  2013-07       Impact factor: 2.311

Review 10.  Implantable and semi-implantable hearing AIDS: a review of history, indications, and surgery.

Authors:  Aline Gomes Bittencourt; Patrick Rademaker Burke; Isabela de Souza Jardim; Rubens de Brito; Robinson Koji Tsuji; Anna Carolina de Oliveira Fonseca; Ricardo Ferreira Bento
Journal:  Int Arch Otorhinolaryngol       Date:  2014-04-09
View more
  5 in total

1.  Air- and Bone-Conducted Sources of Feedback With an Active Middle Ear Implant.

Authors:  Renee M Banakis Hartl; James R Easter; Mohamed A Alhussaini; Daniel J Tollin; Herman A Jenkins
Journal:  Ear Hear       Date:  2019 May/Jun       Impact factor: 3.570

Review 2.  [Active hearing implants in chronic otitis media].

Authors:  S Lailach; C Müller; N Lasurashvili; H Seidler; T Zahnert
Journal:  HNO       Date:  2021-06       Impact factor: 1.284

3.  The first results of a totally implanted active middle ear device.

Authors:  Maria da Conceição Peixoto; Cristina Miranda; Mafalda Bento; Susana Oliveira; Rui Pratas; Victor Correia da Silva
Journal:  Eur Arch Otorhinolaryngol       Date:  2019-07-24       Impact factor: 2.503

4.  Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI.

Authors:  Guy Fierens; Nicolas Verhaert; Farida Benoudiba; Marie-France Bellin; Dennis Ducreux; Jean-François Papon; Jérôme Nevoux
Journal:  PLoS One       Date:  2020-04-09       Impact factor: 3.240

5.  Battery Lifespan of an Implantable Middle Ear Device.

Authors:  Luca Bruschini; Francesca Forli; Giacomo Fiacchini; Rachele Canelli; Stefano Berrettini; Francesco Lazzerini
Journal:  Audiol Res       Date:  2022-09-07
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.