Real time implementation of anti-scatter grid artifact elimination method for high resolution x-ray imaging CMOS detectors using Graphics Processing Units (GPUs).

R Rana1, S V Setlur Nagesh1, D R Bednarek1, S Rudin1.   

Abstract

Scatter is one of the most important factors effecting image quality in radiography. One of the best scatter reduction methods in dynamic imaging is an anti-scatter grid. However, when used with high resolution imaging detectors these grids may leave grid-line artifacts with increasing severity as detector resolution improves. The presence of such artifacts can mask important details in the image and degrade image quality. We have previously demonstrated that, in order to remove these artifacts, one must first subtract the residual scatter that penetrates through the grid followed by dividing out a reference grid image; however, this correction must be done fast so that corrected images can be provided in real-time to clinicians. In this study, a standard stationary Smit-Rontgen x-ray grid (line density - 70 lines/cm, grid ratio - 13:1) was used with a high-resolution CMOS detector, the Dexela 1207 (pixel size - 75 micron) to image anthropomorphic head phantoms. For a 15 × 15 cm field-of-view (FOV), scatter profiles of the anthropomorphic head phantoms were estimated then iteratively modified to minimize the structured noise due to the varying grid-line artifacts across the FOV. Images of the head phantoms taken with the grid, before and after the corrections, were compared, demonstrating almost total elimination of the artifact over the full FOV. This correction is done fast using Graphics Processing Units (GPUs), with 7-8 iterations and total time taken to obtain the corrected image of only 87 ms, hence, demonstrating the virtually real-time implementation of the grid-artifact correction technique.

Entities:  

Keywords:  CMOS detector; GPUs; anthropomorphic head phantoms; anti-scatter grid; grid artifacts; high resolution detector; iterative method; real time correction; scatter estimation; x-ray imaging

Year:  2017        PMID: 28659654      PMCID: PMC5484410          DOI: 10.1117/12.2254120

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  7 in total

Review 1.  Contrast and scatter in x-ray imaging.

Authors:  G T Barnes
Journal:  Radiographics       Date:  1991-03       Impact factor: 5.333

2.  Grids or air gaps for scatter reduction in digital radiography: a model calculation.

Authors:  U Neitzel
Journal:  Med Phys       Date:  1992 Mar-Apr       Impact factor: 4.071

3.  Grid line artifact formation: A comprehensive theory.

Authors:  David M Gauntt; Gary T Barnes
Journal:  Med Phys       Date:  2006-06       Impact factor: 4.071

4.  Artifacts produced by moving grids.

Authors:  D R Bednarek; S Rudin; R Wong
Journal:  Radiology       Date:  1983-04       Impact factor: 11.105

5.  Anti-scatter grid artifact elimination for high resolution x-ray imaging CMOS detectors.

Authors:  R Rana; V Singh; A Jain; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-03-18

6.  Endovascular image-guided interventions (EIGIs).

Authors:  Stephen Rudin; Daniel R Bednarek; Kenneth R Hoffmann
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

7.  Limitations of anti-scatter grids when used with high resolution image detectors.

Authors:  V Singh; A Jain; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19
  7 in total
  1 in total

1.  Anti-scatter grid artifact elimination for high-resolution x-ray imaging detectors without a prior scatter distribution profile.

Authors:  S V Setlur Nagesh; R Rana; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2018-03-09
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.