Scatter reduction for high resolution image detectors with a region of interest attenuator.

Amit Jain1, Daniel R Bednarek1, Stephen Rudin1.   

Abstract

Compton scatter is the main interaction of x-rays with objects undergoing radiographic and fluoroscopic imaging procedures. Such scatter is responsible for reducing image signal to noise ratio which can negatively impact object detection especially for low contrast objects. To reduce scatter, possible methods are smaller fields-of-view, larger air gaps and the use of an anti-scatter grid. Smaller fields of view may not be acceptable and scanned-beam radiography is not practical for real-time imaging. Air gaps can increase geometric unsharpness and thus degrade image resolution. Deployment of an anti-scatter grid is not well suited for high resolution imagers due to the unavailability of high line density grids needed to prevent grid-line artifacts. However, region of interest (ROI) imaging can be used not only for dose reduction but also for scatter reduction in the ROI. The ROI region receives unattenuated x-rays while the peripheral region receives x-rays reduced in intensity by an ROI attenuator. The scatter within the ROI part of the image originates from both the unattenuated ROI and the attenuated peripheral region. The scatter contribution from the periphery is reduced in intensity because of the reduced primary x-rays in that region and the scatter fraction in the ROI is thus reduced. In this study, the scatter fraction for various kVp's, air-gaps and field sizes was measured for a uniform head equivalent phantom. The scatter fraction in the ROI was calculated using a derived scatter fraction formula, which was validated with experimental measurements. It is shown that use of a ROI attenuator can be an effective way to reduce both scatter and patient dose while maintaining the superior image quality of high resolution detectors.

Entities:  

Keywords:  Digital Imaging; ROI Attenuator; Region of Interest; Scatter

Year:  2014        PMID: 25302000      PMCID: PMC4188397          DOI: 10.1117/12.2043061

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  2 in total

1.  Region of interest fluoroscopy.

Authors:  S Rudin; D R Bednarek
Journal:  Med Phys       Date:  1992 Sep-Oct       Impact factor: 4.071

2.  Limitations of anti-scatter grids when used with high resolution image detectors.

Authors:  V Singh; A Jain; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19
  2 in total
  1 in total

1.  Incorporating Corrections for the Head-Holder and Compensation Filter when Calculating Skin Dose during Fluoroscopically-Guided Interventions.

Authors:  Sarath Vijayan; Vijay K Rana; Stephen Rudin; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-03-18
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.