Grid artifact reduction for direct digital radiography detectors based on rotated stationary grids with homomorphic filtering.

Dong Sik Kim1, Sanggyun Lee.   

Abstract

PURPOSE: Grid artifacts are caused when using the antiscatter grid in obtaining digital x-ray images. In this paper, research on grid artifact reduction techniques is conducted especially for the direct detectors, which are based on amorphous selenium.
METHODS: In order to analyze and reduce the grid artifacts, the authors consider a multiplicative grid image model and propose a homomorphic filtering technique. For minimal damage due to filters, which are used to suppress the grid artifacts, rotated grids with respect to the sampling direction are employed, and min-max optimization problems for searching optimal grid frequencies and angles for given sampling frequencies are established. The authors then propose algorithms for the grid artifact reduction based on the band-stop filters as well as low-pass filters.
RESULTS: The proposed algorithms are experimentally tested for digital x-ray images, which are obtained from direct detectors with the rotated grids, and are compared with other algorithms. It is shown that the proposed algorithms can successfully reduce the grid artifacts for direct detectors.
CONCLUSIONS: By employing the homomorphic filtering technique, the authors can considerably suppress the strong grid artifacts with relatively narrow-bandwidth filters compared to the normal filtering case. Using rotated grids also significantly reduces the ringing artifact. Furthermore, for specific grid frequencies and angles, the authors can use simple homomorphic low-pass filters in the spatial domain, and thus alleviate the grid artifacts with very low implementation complexity.

Entities:  

Mesh:

Year:  2013        PMID: 23718596     DOI: 10.1118/1.4807085

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Design and evaluation of a grid reciprocation scheme for use in digital breast tomosynthesis.

Authors:  Tushita Patel; Helen Sporkin; Heather Peppard; Mark B Williams
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-29

2.  Limitations of anti-scatter grids when used with high resolution image detectors.

Authors:  V Singh; A Jain; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19

3.  Patch Based Grid Artifact Suppressing in Digital Mammography.

Authors:  Qingqing Ling; Shuyu Wu; Xiaoman Duan; Genggeng Qin; Jianhui Ma; Chaomin Chen; Hongliang Qi; Linghong Zhou; Yuan Xu
Journal:  Biomed Res Int       Date:  2018-08-12       Impact factor: 3.411

4.  Air gap technique is recommended in axiolateral hip radiographs.

Authors:  Susanne Kivistö; Antti Kotiaho; Anja Henner; Terhi Nevala; Jaakko Niinimäki; Miika T Nieminen; Matti Hanni
Journal:  J Appl Clin Med Phys       Date:  2020-09-21       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.