Florian Schwarz1, Philipp Lange2, Dominik Zinsser1, Martin Greif2, Peter Boekstegers3, Christoph Schmitz4, Maximilian F Reiser5, Christian Kupatt6, Hans C Becker5.
Abstract
BACKGROUND/Entities:
Mesh:
Year: 2014 PMID: 25084451 PMCID: PMC4118882 DOI: 10.1371/journal.pone.0103481
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Figure 1Inclusion chart for our analysis of 351 patients who underwent a successful TAVI procedure.
Figure 2Examples of anatomic measurements at the aortic annulus performed for this study:
isotropic small field-of-view CTA dataset of a 78 year old female patient with severe aortic stenosis, A) sagittal reformation and B) coronal reformation showing the orientation of the aortic annulus plane, C) double-oblique reformation of the aortic valve annulus, demonstrating the diameter of the long axis (black line), the diameter of the short axis (blue line), the annulus circumference (red polygonal), and annulus area (yellow shading). D) angiographic image after implantation of a 26 mm model of the Edward Sapien XT valve prosthesis demonstrates no paravalvular leakage.
Patient parameters, N(%), mean ± SD or median [interquartile range] and summary of implanted valve sizes.
| All Patients | CoreValveSubgroup | Edward SapienSubgroup | p | |
|
| 351 (100%) | 235 (67%) | 116 (33%) | |
|
| 211 (58%) | 132 (54%) | 74 (62%) | 0.18 |
|
| 80.9±10.8 | 81.6±6.5 | 76.3±21.5 | 0.06 |
|
| 165.7±8.9 | 165.5±13.9 | 163.2±18.2 | 0.125 |
|
| 72.9±14.7 | 72.6±15.0 | 72.1±17.0 | 0.83 |
|
| 26.5±4.8 | 26.3±4.5 | 26.8±5.4 | 0.62 |
|
| 0.70±0.15 | 0.69±0.16 | 0.72±0.13 | 0.21 |
|
| 66.9±22.4 | 67.3±23.8 | 66.0±19.6 | 0.65 |
|
| 58.0±15.3 | 57.8±16.0 | 58.3±13.6 | 0.80 |
|
| 18.9 [13.4; 25.7] | 19.6 [14.5; 26.7] | 16.5 [11.0; 23.4] | <0.01 |
|
| ||||
|
| 131 (37%) | 93 (40%) | 38 (33%) | 0.25 |
|
| 207 (59%) | 135 (57%) | 72 (62%) | 0.44 |
|
| 13 (4%) | 7 (3%) | 6 (5%) | 0.52 |
Interobserver Agreement for anatomical measurements of the aortic root reported as Intraclass Correlation Coefficient (ICC) and parameters of Bland-Altman-Analysis.
| Diameter long axis | Diameter short axis | Diameter | Annulus Circum-ference | Annulus Area | |
|
| 0.87 [0.84; 0.89] | 0.86 [0.83; 0.88] | 0.90 [0.88; 0.92] | 0.92 [0.90; 0.93] | 0.93 [0.92; 0.94] |
|
| 0.00 cm | 0.00 cm | 0.00 cm | 0.06 cm | 0.03 cm2 |
|
| 0.14 cm | 0.12 cm | 0.20 cm | 0.54 cm | 0.6 cm2 |
Primary anatomic parameters of the aortic annulus, N(%), mean ± SD or median [interquartile range] as well as “virtual diameters” derived from the mean of long axis and short axis, from annulus circumference and annulus area.
| All Patients | CoreValveSubgroup | Edward SapienSubgroup | p | |
|
| 2.70 [2.52; 2.90] | 2.75 [2.54; 2.90] | 2.65 [2.50; 2.85] | 0.049 |
|
| 2.05 [1.95; 2.25] | 2.10 [1.95; 2.28] | 2.05 [1.90; 2.20] | <0.01 |
|
| 7.65 [7.28; 8.15] | 7.73 [7.30; 8.25] | 7.55 [7.05; 8.00] | <0.01 |
|
| 4.34 [3.90; 4.95] | 4.44 [3.98; 5.05] | 4.19 [3.70; 4.75] | <0.01 |
|
| 2.40 [2.25; 2.55] | 2.42 [2.27; 2.58] | 2.35 [2.22; 2.52] | <0.01 |
|
| 2.44 [2.32; 2.60] | 2.46 [2.32; 2.63] | 2.40 [2.24; 2.55] | <0.01 |
|
| 2.35 [2.23; 2.51] | 2.38 [2.25; 2.54] | 2.31 [2.17; 2.46] | <0.01 |
Figure 3Summary chart displaying major anatomic parameters (length of long and short axis; circumference-derived and area-derived virtual diameter) in patients in whom the small, medium or large size model was considered optimal for A) the CoreValve valve and the B) Edward Sapien XT valve.
Except for the large vs. medium CoreValve size, all differences were statistically significant (p<0.05).
Multi-Class AUC’s as measures for the predictive value of the respective anatomic parameters for the valve size considered optimal by the implantation team.
| PARAMETER | OPTIMAL CUTOFFSMALL vs. MEDIUM | OPTIMAL CUTOFFMEDIUM vs. LARGE | MULTI-CLASS AUC | |
|
| Diameter Long Axis | 2.70 cm | 3.04 cm | 0.8133 |
| Diameter Short Axis | 2.05 cm | 2.29 cm | 0.7478 | |
| Diameter | 2.35 cm | 2.70 cm | 0.7974 | |
| Annulus Area | 4.30 cm2 | 5.72 cm2 | 0.8812 | |
| Annulus Circumference | 7.65 cm | 8.60 cm | 0.9146 | |
|
| Diameter Long Axis | 2.55 cm | 3.05 cm | 0.8347 |
| Diameter Short axis | 1.95 cm | 2.30 cm | 0.7995 | |
| Diameter | 2.25 cm | 2.55 cm | 0.8793 | |
| Annulus Area | 3.85 cm2 | 5.35 cm2 | 0.8823 | |
| Annulus Circumference | 7.15 cm | 8.20 cm | 0.8884 |
Optimal cutoff values for the selection of the small, medium or large prosthesis size, defined as cutoff values that result in highest predictive accuracy.