Prospective ECG-triggered, axial 4-D imaging of the aortic root, valvular, and left ventricular structures: a lower radiation dose option for preprocedural TAVR imaging.

Michael A Bolen1, Zoran B Popovic, Arun Dahiya, Samir R Kapadia, E Murat Tuzcu, Scott D Flamm, Sandra S Halliburton, Paul Schoenhagen.   

Abstract

BACKGROUND: Transcatheter valve interventions rely on imaging for patient selection, preprocedural planning, and intraprocedural guidance.
OBJECTIVE: We explored the use of prospective electrocardiogram (ECG)-triggered 4-dimensional (4-D) CT imaging in patients evaluated for transcatheter aortic valve replacement (TAVR).
METHODS: A total of 47 consecutive patients underwent 128-slice dual-source CT with wide-window dose-modulated prospective ECG-triggered, axial acquisition of the aortic root, reconstructed during diastolic and systolic cardiac phases. Image quality was evaluated, aortic root and left ventricular (LV) geometry and function were analyzed, and radiation exposure was estimated.
RESULTS: Image quality was generally good, with 41 of 47 (87%) patients scored as good or excellent. The mean aortic valve area was 0.93 ± 0.24 cm(2). Mean LV ejection fraction was 56.8% ± 16.4%, and mean LV mass was 130.4 ± 43.8 g. The minor diameter of the annulus was larger in systole (systole, 2.29 ± 0.24 cm; diastole, 2.14 ± 0.25 cm; P = 0.006), but the mean and major diameters did not vary significantly between systole and diastole. The mean estimated effective dose was 5.9 ± 2.4 mSv.
CONCLUSION: Multiphase, prospective ECG-triggered axial image acquisition is a lower dose acquisition technique for 4-D aortic root imaging in patients being considered for TAVR.
Copyright © 2012 Society of Cardiovascular Computed Tomography. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 23217462     DOI: 10.1016/j.jcct.2012.10.006

Source DB:  PubMed          Journal:  J Cardiovasc Comput Tomogr        ISSN: 1876-861X


  3 in total

1.  Novel single-source high-pitch protocol for CT angiography of the aorta: comparison to high-pitch dual-source protocol in the context of TAVI planning.

Authors:  Bernhard Bischoff; Felix G Meinel; Maximilian Reiser; Hans-Christoph Becker
Journal:  Int J Cardiovasc Imaging       Date:  2013-01-19       Impact factor: 2.357

2.  Contrast media extravasations in patients undergoing computerized tomography scanning: a systematic review and meta-analysis of risk factors and interventions.

Authors:  Sandrine Ding; Nicole Richli Meystre; Cosmin Campeanu; Giuseppe Gullo
Journal:  JBI Database System Rev Implement Rep       Date:  2018-01

3.  CT-angiography-based evaluation of the aortic annulus for prosthesis sizing in transcatheter aortic valve implantation (TAVI)-predictive value and optimal thresholds for major anatomic parameters.

Authors:  Florian Schwarz; Philipp Lange; Dominik Zinsser; Martin Greif; Peter Boekstegers; Christoph Schmitz; Maximilian F Reiser; Christian Kupatt; Hans C Becker
Journal:  PLoS One       Date:  2014-08-01       Impact factor: 3.240