Size-specific dose estimates: Localizer or transverse abdominal computed tomography images?

Sarvenaz Pourjabbar1, Sarabjeet Singh1, Atul Padole1, Akshay Saini1, Michael A Blake1, Mannudeep K Kalra1.   

Abstract

AIM: To investigate effect of body dimensions obtained from localizer radiograph and transverse abdominal computed tomography (CT) images on Size Specific Dose Estimate.
METHODS: This study was approved by Institutional Review Board and was compliant with Health Insurance Portability and Accountability Act. Fifty patients with abdominal CT examinations (58 ± 13 years, Male:Female 28:22) were included in this study. Anterior-posterior (AP) and lateral (Lat) diameters were measured at 5 cm intervals from the CT exam localizer radiograph (simple X-ray image acquired for planning the CT exam before starting the scan) and transverse CT images. Average of measured AP and Lat diameters, as well as maximum, minimum and mid location AP and Lat were measured on both image sets. In addition, off centering of patients from the gantry iso-center was calculated from the localizers. Conversion factors from American Association of Physicists in Medicine (AAPM) report 204 were obtained for AP, Lat, AP + Lat, and effective diameter (√ AP * Lat) to determine size specific dose estimate (SSDE) from the CT dose index volume (CTDIvol) recorded from the dose reports. Data were analyzed using SPSS v19.
RESULTS: Total number of 5376 measurements was done. In some patients entire body circumference was not covered on either projection radiograph or transverse CT images; hence accurate measurement of AP and Lat diameters was not possible in 11% (278/2488) of locations. Forty one patients were off-centered with mean of 1.9 ± 1.8 cm (range: 0.4-7 cm). Conversion factors for attained diameters were not listed on AAPM look-up tables in 3% (80/2488) of measurements. SSDE values were significantly different compared to CTDIvol, ranging from 32% lower to 74% greater than CTDIvol.
CONCLUSION: There is underestimation and overestimation of dose comparing SSDE values to CTDIvol. Localizer radiographs are associated with overestimation of patient size and therefore underestimation of SSDE.

Entities:  

Keywords:  Computed tomography dose index; Radiation dose optimization; Radiation dose reduction; Size specific dose estimate

Year:  2014        PMID: 24876925      PMCID: PMC4037547          DOI: 10.4329/wjr.v6.i5.210

Source DB:  PubMed          Journal:  World J Radiol        ISSN: 1949-8470


  21 in total

1.  The feasibility of a scanner-independent technique to estimate organ dose from MDCT scans: using CTDIvol to account for differences between scanners.

Authors:  Adam C Turner; Maria Zankl; John J DeMarco; Chris H Cagnon; Di Zhang; Erin Angel; Dianna D Cody; Donna M Stevens; Cynthia H McCollough; Michael F McNitt-Gray
Journal:  Med Phys       Date:  2010-04       Impact factor: 4.071

2.  Misoperation of CT automatic tube current modulation systems with inappropriate patient centering: phantom studies.

Authors:  Kosuke Matsubara; Kichiro Koshida; Katsuhiro Ichikawa; Masayuki Suzuki; Tadanori Takata; Tomoyuki Yamamoto; Osamu Matsui
Journal:  AJR Am J Roentgenol       Date:  2009-04       Impact factor: 3.959

3.  Efficient use of automatic exposure control systems in computed tomography requires correct patient positioning.

Authors:  J Gudjonsdottir; J R Svensson; S Campling; P C Brennan; B Jonsdottir
Journal:  Acta Radiol       Date:  2009-11       Impact factor: 1.990

4.  CT dose index and patient dose: they are not the same thing.

Authors:  Cynthia H McCollough; Shuai Leng; Lifeng Yu; Dianna D Cody; John M Boone; Michael F McNitt-Gray
Journal:  Radiology       Date:  2011-05       Impact factor: 11.105

5.  Ablation margin assessment of liver tumors with intravenous contrast-enhanced C-arm computed tomography.

Authors:  Mi Sung Kim; Sarabjeet Singh; Elkan Halpern; Sanjay Saini; Mannudeep K Kalra
Journal:  World J Radiol       Date:  2012-03-28

6.  Size-specific dose estimates for adult patients at CT of the torso.

Authors:  Jodie A Christner; Natalie N Braun; Megan C Jacobsen; Rickey E Carter; James M Kofler; Cynthia H McCollough
Journal:  Radiology       Date:  2012-10-22       Impact factor: 11.105

7.  Investigation of American Association of Physicists in Medicine Report 204 size-specific dose estimates for pediatric CT implementation.

Authors:  Samuel L Brady; Robert A Kaufman
Journal:  Radiology       Date:  2012-10-23       Impact factor: 11.105

8.  Entrance skin dosimetry and size-specific dose estimate from pediatric chest CTA.

Authors:  Sjirk J Westra; Xinhua Li; Kavita Gulati; Sarabjeet Singh; Bob Liu; Mannudeep K Kalra; Suhny Abbara
Journal:  J Cardiovasc Comput Tomogr       Date:  2013-08-23

9.  Effects of protocol and obesity on dose conversion factors in adult body CT.

Authors:  Xiang Li; Ehsan Samei; Cameron H Williams; W Paul Segars; Daniel J Tward; Michael I Miller; J Tilak Ratnanather; Erik K Paulson; Donald P Frush
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

10.  The feasibility of a regional CTDIvol to estimate organ dose from tube current modulated CT exams.

Authors:  Maryam Khatonabadi; Hyun J Kim; Peiyun Lu; Kyle L McMillan; Chris H Cagnon; John J DeMarco; Michael F McNitt-Gray
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

View more
  6 in total

1.  How accurate is size-specific dose estimate in pediatric body CT examinations?

Authors:  Boaz Karmazyn; Huisi Ai; Paul Klahr; Fangqian Ouyang; S Gregory Jennings
Journal:  Pediatr Radiol       Date:  2016-04-06

2.  A comparative study of radiation doses between phantom and patients via CT angiography of the intra-/extra-cranial, pulmonary, and abdominal/pelvic arteries.

Authors:  M K A Karim; A Sabarudin; N A Muhammad; K H Ng
Journal:  Radiol Phys Technol       Date:  2019-08-29

3.  Does body mass index outperform body weight as a surrogate parameter in the calculation of size-specific dose estimates in adult body CT?

Authors:  Johannes Boos; Rotem S Lanzman; Philipp Heusch; Joel Aissa; Christoph Schleich; Christoph Thomas; Lino M Sawicki; Gerald Antoch; Patric Kröpil
Journal:  Br J Radiol       Date:  2015-12-23       Impact factor: 3.039

4.  Multi-slice CT examinations of adult patients at Sudanese hospitals: radiation exposure based on size-specific dose estimates (SSDE).

Authors:  Einas H Bashier; I I Suliman
Journal:  Radiol Med       Date:  2018-01-30       Impact factor: 3.469

5.  Appraisal of radiation dose with 64-slice computed tomography perfusion in lung cancer patients with special reference to SSDE: An initial experience in a tertiary care hospital.

Authors:  Shuchi Bhatt; Ajai K Srivastava; Neha Meena; Shweta Thakur
Journal:  Indian J Radiol Imaging       Date:  2017 Oct-Dec

6.  Automated Calculation of Water-equivalent Diameter (DW) Based on AAPM Task Group 220.

Authors:  Choirul Anam; Freddy Haryanto; Rena Widita; Idam Arif; Geoff Dougherty
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.