How accurate is size-specific dose estimate in pediatric body CT examinations?

Boaz Karmazyn1,2, Huisi Ai3, Paul Klahr4, Fangqian Ouyang5, S Gregory Jennings6.   

Abstract

BACKGROUND: Size-specific dose estimate is gaining increased acceptance as the preferred index of CT dose in children. However it was developed based on non-clinical data.
OBJECTIVE: To compare the accuracy of size-specific dose estimate (SSDE) based on geometric and body weight measures in pediatric chest and abdomen CT scans, versus the more accurate [Formula: see text] (mean SSDE based on water-equivalent diameter).
MATERIALS AND METHODS: We retrospectively identified 50 consecutive children (age <18 years) who underwent chest CT examination and 50 children who underwent abdomen CT. We measured anteroposterior diameter (DAP) and lateral diameter (DLAT) at the central slice (of scan length) of each patient and calculated DAP+LAT (anteroposterior diameter plus lateral diameter) and DED (effective diameter) for each patient. We calculated the following in each child: (1) SSDEs based on DAP, DLAT, DAP+LAT, DED, and body weight, and (2) SSDE based on software calculation of mean water-equivalent diameter ([Formula: see text] adopted standard within our study). We used intraclass correlation coefficient (ICC) and Bland-Altman analysis to compare agreement between the SSDEs and [Formula: see text].
RESULTS: Gender and age distribution were similar between chest and abdomen CT groups; mean body weight was 37 kg for both groups, with ranges of 6-130 kg (chest) and 8-107 kg (abdomen). SSDEs had very strong agreement (ICC>0.9) with [Formula: see text]. SSDEs based on DLAT had 95% limits of agreement of up to 43% with [Formula: see text]. SSDEs based on other parameters (body weight, DAP, DAP+LAT, DED) had 95% limits of agreement of up to 25%.
CONCLUSION: Differences between SSDEs calculated using various indications of patient size (geometric indices and patient weight) and the more accurate [Formula: see text] calculated using proprietary software were generally small, with the possible exception for lateral diameter, and provide acceptable dose estimates for body CT in children.

Entities:  

Keywords:  Abdomen; Chest; Children; Computed tomography dose index volume; Size-specific dose estimate

Mesh:

Year:  2016        PMID: 27053280     DOI: 10.1007/s00247-016-3604-0

Source DB:  PubMed          Journal:  Pediatr Radiol        ISSN: 0301-0449


  15 in total

1.  The ALARA concept in pediatric CT: myth or reality?

Authors:  Thomas L Slovis
Journal:  Radiology       Date:  2002-04       Impact factor: 11.105

2.  Where we are since ALARA and the series of articles on CT dose in children and risk of long-term cancers: what has changed?

Authors:  Walter E Berdon; Thomas L Slovis
Journal:  Pediatr Radiol       Date:  2002-09-03

3.  CT dose index and patient dose: they are not the same thing.

Authors:  Cynthia H McCollough; Shuai Leng; Lifeng Yu; Dianna D Cody; John M Boone; Michael F McNitt-Gray
Journal:  Radiology       Date:  2011-05       Impact factor: 11.105

4.  How to Appropriately Calculate Effective Dose for CT Using Either Size-Specific Dose Estimates or Dose-Length Product.

Authors:  Samuel L Brady; Amy E Mirro; Bria M Moore; Robert A Kaufman
Journal:  AJR Am J Roentgenol       Date:  2015-03-02       Impact factor: 3.959

5.  Automated size-specific CT dose monitoring program: assessing variability in CT dose.

Authors:  Olav Christianson; Xiang Li; Donald Frush; Ehsan Samei
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

6.  Size-specific Dose Estimates for Chest, Abdominal, and Pelvic CT: Effect of Intrapatient Variability in Water-equivalent Diameter.

Authors:  Shuai Leng; Maria Shiung; Xinhui Duan; Lifeng Yu; Yi Zhang; Cynthia H McCollough
Journal:  Radiology       Date:  2015-02-25       Impact factor: 11.105

Review 7.  Estimated pediatric radiation dose during CT.

Authors:  Keith J Strauss; Marilyn J Goske
Journal:  Pediatr Radiol       Date:  2011-08-17

8.  Size-specific dose estimates for adult patients at CT of the torso.

Authors:  Jodie A Christner; Natalie N Braun; Megan C Jacobsen; Rickey E Carter; James M Kofler; Cynthia H McCollough
Journal:  Radiology       Date:  2012-10-22       Impact factor: 11.105

9.  Use of Water Equivalent Diameter for Calculating Patient Size and Size-Specific Dose Estimates (SSDE) in CT: The Report of AAPM Task Group 220.

Authors:  Cynthia McCollough; Donovan M Bakalyar; Maryam Bostani; Samuel Brady; Kristen Boedeker; John M Boone; H Heather Chen-Mayer; Olav I Christianson; Shuai Leng; Baojun Li; Michael F McNitt-Gray; Roy A Nilsen; Mark P Supanich; Jia Wang
Journal:  AAPM Rep       Date:  2014-09

10.  Size-specific dose estimates: Localizer or transverse abdominal computed tomography images?

Authors:  Sarvenaz Pourjabbar; Sarabjeet Singh; Atul Padole; Akshay Saini; Michael A Blake; Mannudeep K Kalra
Journal:  World J Radiol       Date:  2014-05-28
View more
  4 in total

1.  A comparison study of size-specific dose estimate calculation methods.

Authors:  Roshni A Parikh; Michael A Wien; Ronald D Novak; David W Jordan; Paul Klahr; Stephanie Soriano; Leslie Ciancibello; Sheila C Berlin
Journal:  Pediatr Radiol       Date:  2017-09-27

2.  Multiphase acquisitions in pediatric abdominal-pelvic CT are a common practice and contribute to unnecessary radiation dose.

Authors:  Bradley S Rostad; Kimberly E Applegate; Tammy Kim; Renee M Mansour; Sarah S Milla
Journal:  Pediatr Radiol       Date:  2018-07-07

3.  Can Patient's Body Weight Represent Body Diameter for Pediatric Size-Specific Dose Estimate in Thoracic and Abdominal Computed Tomography?

Authors:  Supika Kritsaneepaiboon; Suwadee Eng-Chuan; Saowapark Yoykaew
Journal:  J Clin Imaging Sci       Date:  2019-05-24

4.  Size-Specific Dose Estimates of Radiation Based on Body Weight and Body Mass Index for Chest and Abdomen-Pelvic CTs.

Authors:  Jian Xu; Xiangquan Wang; Panfeng Yang; Kuangnan Luo; Xiaolong He
Journal:  Biomed Res Int       Date:  2020-07-10       Impact factor: 3.411

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.