Ablation margin assessment of liver tumors with intravenous contrast-enhanced C-arm computed tomography.

Mi Sung Kim1, Sarabjeet Singh, Elkan Halpern, Sanjay Saini, Mannudeep K Kalra.   

Abstract

AIM: To determine the influence of anthropomorphic parameters on the relationship between patient centering, mean computed tomography (CT) numbers and quantitative image noise in abdominal CT.
METHODS: Our Institutional Review Board approved study included 395 patients (age range 21-108, years; male:female = 195:200) who underwent contrast-enhanced abdominal CT on a 16-section multi-detector row scanner (GE LightSpeed 16). Patient centering in the gantry isocenter was measured from the lateral localizer radiograph (off center S = patient off centered superior to isocenter; off center I = patient off centered inferior to isocenter). Mean CT numbers (Hounsfield Units: HU) and noise (standard deviation of CT numbers: SD) were measured in the anterior (aHU, aSD) and posterior (pHU, pSD) abdominal wall subcutaneous fat and liver parenchyma (LivHU, LivSD) at the level of the porta hepatis. Patients' age, gender, weight, body mass index and maximal anteroposterior diameter were recorded. The data were analyzed using linear regression analysis.
RESULTS: Most patients (81%; 320/395) were not correctly centered in the gantry isocenter for abdominal CT scanning. Mean CT numbers in the abdominal wall increased significantly with an increase in the off-centering distance, regardless of the direction of the off-center (P < 0.05). There was a substantial increase in pSD (P = 0.01) and LivSD (P = 0.017) with off-centering. Change in mean CT numbers and image noise along the off-center distance was influenced by the patient size (P < 0.01).
CONCLUSION: Inappropriate patient centering for CT scanning adversely affects the reliability of mean CT numbers and image noise.

Entities:  

Keywords:  Automatic exposure control; Computed tomography radiation dose reduction; Patient offcentering; Tube current modulation

Year:  2012        PMID: 22468191      PMCID: PMC3314926          DOI: 10.4329/wjr.v4.i3.102

Source DB:  PubMed          Journal:  World J Radiol        ISSN: 1949-8470


  11 in total

1.  Analysis of surface dose variation in CT procedures.

Authors:  P Avilés Lucas; I A Castellano; D R Dance; E Vañó Carruana
Journal:  Br J Radiol       Date:  2001-12       Impact factor: 3.039

Review 2.  Application of rational practice and technical advances for optimizing radiation dose for chest CT.

Authors:  Michael M Maher; Mannudeep K Kalra; Thomas L Toth; Conrad Wittram; Sanjay Saini; Joanne Shepard
Journal:  J Thorac Imaging       Date:  2004-01       Impact factor: 3.000

3.  Misoperation of CT automatic tube current modulation systems with inappropriate patient centering: phantom studies.

Authors:  Kosuke Matsubara; Kichiro Koshida; Katsuhiro Ichikawa; Masayuki Suzuki; Tadanori Takata; Tomoyuki Yamamoto; Osamu Matsui
Journal:  AJR Am J Roentgenol       Date:  2009-04       Impact factor: 3.959

4.  Minimizing radiation dose for pediatric body applications of single-detector helical CT: strategies at a large Children's Hospital.

Authors:  L F Donnelly; K H Emery; A S Brody; T Laor; V M Gylys-Morin; C G Anton; S R Thomas; D P Frush
Journal:  AJR Am J Roentgenol       Date:  2001-02       Impact factor: 3.959

5.  Radiation dose and cancer risk from pediatric CT examinations on 64-slice CT: a phantom study.

Authors:  Shi-Ting Feng; Martin Wai-Ming Law; Bingsheng Huang; Sherry Ng; Zi-Ping Li; Quan-Fei Meng; Pek-Lan Khong
Journal:  Eur J Radiol       Date:  2010-04-03       Impact factor: 3.528

6.  Automatic patient centering for MDCT: effect on radiation dose.

Authors:  Jianhai Li; Unni K Udayasankar; Thomas L Toth; John Seamans; William C Small; Mannudeep K Kalra
Journal:  AJR Am J Roentgenol       Date:  2007-02       Impact factor: 3.959

Review 7.  Strategies for CT radiation dose optimization.

Authors:  Mannudeep K Kalra; Michael M Maher; Thomas L Toth; Leena M Hamberg; Michael A Blake; Jo-Anne Shepard; Sanjay Saini
Journal:  Radiology       Date:  2004-01-22       Impact factor: 11.105

8.  Application of automatic vertical positioning software to reduce radiation exposure in multidetector row computed tomography of the chest.

Authors:  Jianhai Li; Unni K Udayasankar; Thomas L Toth; William C Small; Mannudeep K Kalra
Journal:  Invest Radiol       Date:  2008-06       Impact factor: 6.016

9.  Recurrent CT, cumulative radiation exposure, and associated radiation-induced cancer risks from CT of adults.

Authors:  Aaron Sodickson; Pieter F Baeyens; Katherine P Andriole; Luciano M Prevedello; Richard D Nawfel; Richard Hanson; Ramin Khorasani
Journal:  Radiology       Date:  2009-04       Impact factor: 11.105

10.  Cumulative radiation exposure and cancer risk estimates in emergency department patients undergoing repeat or multiple CT.

Authors:  Richard T Griffey; Aaron Sodickson
Journal:  AJR Am J Roentgenol       Date:  2009-04       Impact factor: 3.959

View more
  3 in total

1.  Size-specific dose estimates: Localizer or transverse abdominal computed tomography images?

Authors:  Sarvenaz Pourjabbar; Sarabjeet Singh; Atul Padole; Akshay Saini; Michael A Blake; Mannudeep K Kalra
Journal:  World J Radiol       Date:  2014-05-28

2.  The Improvement of Patient Centering in Computed Tomography Through a Technologist-Focused Education Initiative.

Authors:  Lindsay DeWeese; Thomas Griglock; Alexander Moody; Aaron Mehlberg; Celeste Winters
Journal:  J Digit Imaging       Date:  2022-01-12       Impact factor: 4.056

3.  Impact of patient centering in CT on organ dose and the effect of using a positioning compensation system: Evidence from OSLD measurements in postmortem subjects.

Authors:  Izabella Barreto; Rebecca Lamoureux; Catherine Olguin; Nathan Quails; Nathalie Correa; Lynn Rill; Manuel Arreola
Journal:  J Appl Clin Med Phys       Date:  2019-05-02       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.