Misoperation of CT automatic tube current modulation systems with inappropriate patient centering: phantom studies.

Kosuke Matsubara1, Kichiro Koshida, Katsuhiro Ichikawa, Masayuki Suzuki, Tadanori Takata, Tomoyuki Yamamoto, Osamu Matsui.   

Abstract

OBJECTIVE: Inappropriate patient centering on the gantry changes the size of the localizer radiographs used for CT examinations, influencing the operation of CT automatic tube current modulation because tube current is controlled with information from localizer radiographs. The purpose of this study was to examine the influence of inappropriate patient centering on the gantry isocenter on automatic tube current modulation.
MATERIALS AND METHODS: An elliptical phantom was scanned with four automatic tube current modulation techniques after acquisition of localizer radiographs in the horizontal and vertical directions with the phantom center shifted from the gantry isocenter in the vertical direction. After scanning, the magnification rate of the frontal localizer radiographs, tube current-time product, and image noise were examined.
RESULTS: On phantom studies, the magnification rate of localizer radiographs showed a linear relation to the vertical deviation of the phantom from the gantry isocenter. From 50 mm above to 50 mm below the gantry isocenter, tube current-time products ranged from 75% to 141% compared with those at the gantry isocenter. In addition, increases and decreases in the amount of image noise related to changes in tube current-time product were confirmed.
CONCLUSION: Inappropriate patient centering causes misoperation of automatic tube current modulation systems, in which tube current is controlled with information from localizer radiographs, and thus causes increases in tube current or image noise.

Entities:  

Mesh:

Year:  2009        PMID: 19304687     DOI: 10.2214/AJR.08.1472

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  18 in total

1.  Local noise estimation in low-dose chest CT images.

Authors:  J Padgett; A M Biancardi; C I Henschke; D Yankelevitz; A P Reeves
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-07-23       Impact factor: 2.924

2.  An education and training programme for radiological institutes: impact on the reduction of the CT radiation dose.

Authors:  Sebastian T Schindera; Reto Treier; Gabriel von Allmen; Claude Nauer; Philipp R Trueb; Peter Vock; Zsolt Szucs-Farkas
Journal:  Eur Radiol       Date:  2011-05-31       Impact factor: 5.315

3.  Investigating the CT localizer radiograph: acquisition parameters, patient centring and their combined influence on radiation dose.

Authors:  J W Lambert; S Kumar; J S Chen; Z J Wang; R G Gould; B M Yeh
Journal:  Br J Radiol       Date:  2015-01-22       Impact factor: 3.039

Review 4.  Radiation dose management for pediatric cardiac computed tomography: a report from the Image Gently 'Have-A-Heart' campaign.

Authors:  Cynthia K Rigsby; Sarah E McKenney; Kevin D Hill; Anjali Chelliah; Andrew J Einstein; B Kelly Han; Joshua D Robinson; Christina L Sammet; Timothy C Slesnick; Donald P Frush
Journal:  Pediatr Radiol       Date:  2018-01-01

5.  Abdominal CT during pregnancy: a phantom study on the effect of patient centring on conceptus radiation dose and image quality.

Authors:  G Solomou; A E Papadakis; J Damilakis
Journal:  Eur Radiol       Date:  2014-11-28       Impact factor: 5.315

6.  Ablation margin assessment of liver tumors with intravenous contrast-enhanced C-arm computed tomography.

Authors:  Mi Sung Kim; Sarabjeet Singh; Elkan Halpern; Sanjay Saini; Mannudeep K Kalra
Journal:  World J Radiol       Date:  2012-03-28

7.  Effect of vertical positioning on organ dose, image noise and contrast in pediatric chest CT--phantom study.

Authors:  Touko Kaasalainen; Kirsi Palmu; Anniina Lampinen; Mika Kortesniemi
Journal:  Pediatr Radiol       Date:  2013-01-23

8.  Size-specific dose estimates: Localizer or transverse abdominal computed tomography images?

Authors:  Sarvenaz Pourjabbar; Sarabjeet Singh; Atul Padole; Akshay Saini; Michael A Blake; Mannudeep K Kalra
Journal:  World J Radiol       Date:  2014-05-28

9.  Patient Vertical Centering and Correlation with Radiation Output in Adult Abdominopelvic CT.

Authors:  Phillip M Cheng
Journal:  J Digit Imaging       Date:  2016-08       Impact factor: 4.056

10.  A questionnaire survey reviewing radiologists' and clinical specialist radiographers' knowledge of CT exposure parameters.

Authors:  S J Foley; M G Evanoff; L A Rainford
Journal:  Insights Imaging       Date:  2013-09-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.