| Literature DB >> 24610994 |
Sevi B Kocagöz1, Richard J Underwood2, Shiril Sivan3, Jeremy L Gilbert3, Daniel W Macdonald1, Judd S Day2, Steven M Kurtz2.
Abstract
Previous studies have speculated that modular taper design may have an effect on the corrosion and material loss at the taper surfaces. We present a novel method to measure taper angle for retrieved head taper and stem trunnions using a roundness machine (Talyrond 585, Taylor Hobson, UK). We also investigated the relationship between taper angle clearance and visual fretting-corrosion score at the taper-trunnion junction using a matched cohort study of 50 ceramic and 50 metal head-stem pairs. In this study, no correlation was observed between the taper angle clearance and the visual fretting-corrosion scores in either the ceramic or the metal cohorts.Entities:
Year: 2013 PMID: 24610994 PMCID: PMC3941749 DOI: 10.1053/j.sart.2014.01.002
Source DB: PubMed Journal: Semin Arthroplasty ISSN: 1045-4527