Literature DB >> 1624507

Corrosion between the components of modular femoral hip prostheses.

J P Collier1, V A Surprenant, R E Jensen, M B Mayor, H P Surprenant.   

Abstract

We studied the tapered interface between the head and the neck of 139 modular femoral components of hip prostheses which had been removed for a variety of reasons. In 91 the same alloy had been used for the head and the stem; none of them showed evidence of corrosion. In contrast, there was definite corrosion in 25 of the 48 prostheses in which the stem was of titanium alloy and the head of cobalt-chrome. This corrosion was time-dependent: no specimens were corroded after less than nine months in the body, but all which had been in place for more than 40 months were damaged. We discuss the factors which may influence the rate of these changes and present evidence that they were due to galvanically-accelerated crevice corrosion, which was undetected in previous laboratory testing of this type of prosthesis.

Mesh:

Substances:

Year:  1992        PMID: 1624507

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  35 in total

1.  One hundred and fifty years of history of the Morse taper: from Stephen A. Morse in 1864 to complications related to modularity in hip arthroplasty.

Authors:  Philippe Hernigou; Steffen Queinnec; Charles Henri Flouzat Lachaniette
Journal:  Int Orthop       Date:  2013-05-30       Impact factor: 3.075

Review 2.  What is the trouble with trunnions?

Authors:  Christina I Esposito; Timothy M Wright; Stuart B Goodman; Daniel J Berry
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

Review 3.  Metallic debris from metal-on-metal total hip arthroplasty regulates periprosthetic tissues.

Authors:  Christoph H Lohmann; Gurpal Singh; Hans-Georg Willert; Gottfried H Buchhorn
Journal:  World J Orthop       Date:  2014-11-18

4.  Modern trunnions are more flexible: a mechanical analysis of THA taper designs.

Authors:  David A Porter; Robert M Urban; Joshua J Jacobs; Jeremy L Gilbert; José A Rodriguez; H John Cooper
Journal:  Clin Orthop Relat Res       Date:  2014-09-30       Impact factor: 4.176

Review 5.  What Surgeons Need to Know About Adverse Local Tissue Reaction in Total Hip Arthroplasty.

Authors:  Deborah J Hall; Robin Pourzal; Joshua J Jacobs
Journal:  J Arthroplasty       Date:  2020-01-15       Impact factor: 4.757

6.  Stem-Sleeve Junction Failure of a Modular Femoral Hip System: a Retrieval Analysis.

Authors:  Feras Waly; Fahad H Abduljabbar; Trevor Gascoyne; Thomas R Turgeon; Olga Huk
Journal:  HSS J       Date:  2015-07-09

7.  Does Taper Angle Clearance Influence Fretting and Corrosion Damage at the Head-Stem Interface? A Matched Cohort Retrieval Study.

Authors:  Sevi B Kocagöz; Richard J Underwood; Shiril Sivan; Jeremy L Gilbert; Daniel W Macdonald; Judd S Day; Steven M Kurtz
Journal:  Semin Arthroplasty       Date:  2013-12-01

8.  Adverse local tissue reaction arising from corrosion at the femoral neck-body junction in a dual-taper stem with a cobalt-chromium modular neck.

Authors:  H John Cooper; Robert M Urban; Richard L Wixson; R Michael Meneghini; Joshua J Jacobs
Journal:  J Bone Joint Surg Am       Date:  2013-05-15       Impact factor: 5.284

9.  Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty.

Authors:  H John Cooper; Craig J Della Valle; Richard A Berger; Matthew Tetreault; Wayne G Paprosky; Scott M Sporer; Joshua J Jacobs
Journal:  J Bone Joint Surg Am       Date:  2012-09-19       Impact factor: 5.284

10.  Modular titanium alloy neck adapter failures in hip replacement--failure mode analysis and influence of implant material.

Authors:  Thomas M Grupp; Thomas Weik; Wilhelm Bloemer; Hanns-Peter Knaebel
Journal:  BMC Musculoskelet Disord       Date:  2010-01-04       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.