Wear patterns of taper connections in retrieved large diameter metal-on-metal bearings.

Nicholas Bishop1, Florian Witt, Robin Pourzal, Alfons Fischer, Marcel Rütschi, Markus Michel, Michael Morlock.   

Abstract

Wear of the modular taper between head and shaft has been related to clinical failure resulting from adverse reactions to metallic debris. The problem has become pronounced in large metal-on-metal bearings, but the mechanism has not yet been fully understood. We analyzed retrieved components from five patients revised with various diagnoses. Two distinct wear patterns were observed for the head tapers. Three samples demonstrated "asymmetric" wear towards the inner end of the head taper. The other two showed "axisymmetric" radial wear (up to 65 µm) presenting the largest wear volumes (up to 20 mm(3)). Stem tapers demonstrated relatively little wear, and the fine thread on the stem taper surface was observed to be imprinted on the taper inside of the head. Our findings demonstrate that the cobalt-chrome head wears preferentially to the titanium stem taper. "asymmetric" wear suggests toggling due to the offset of the joint force vector from the taper. In contrast, samples with "axisymmetric" radial wear and a threaded imprint suggested that corrosion led to head subsidence onto the stem taper with gradual rotation.
Copyright © 2013 Orthopaedic Research Society.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23440943     DOI: 10.1002/jor.22326

Source DB:  PubMed          Journal:  J Orthop Res        ISSN: 0736-0266            Impact factor:   3.494


  26 in total

Review 1.  What Factors Drive Taper Corrosion?

Authors:  Robin Pourzal; Hannah J Lundberg; Deborah J Hall; Joshua J Jacobs
Journal:  J Arthroplasty       Date:  2018-03-30       Impact factor: 4.757

Review 2.  Metallic debris from metal-on-metal total hip arthroplasty regulates periprosthetic tissues.

Authors:  Christoph H Lohmann; Gurpal Singh; Hans-Georg Willert; Gottfried H Buchhorn
Journal:  World J Orthop       Date:  2014-11-18

3.  [Failure analysis as basis for quality assurance strategies in implant technology].

Authors:  D Behrend; M Warkentin; D Klüß; R Bader; S Kopp; M Frank; W Mittelmeier
Journal:  Orthopade       Date:  2014-06       Impact factor: 1.087

4.  Is there material loss at the backside taper in modular CoCr acetabular liners?

Authors:  Matthias T Agne; Richard J Underwood; Sevi B Kocagoz; Daniel W MacDonald; Judd S Day; Javad Parvizi; Matthew J Kraay; Michael A Mont; Gregg R Klein; Harold E Cates; Steven M Kurtz
Journal:  Clin Orthop Relat Res       Date:  2014-10-16       Impact factor: 4.176

5.  Mechanical, chemical and biological damage modes within head-neck tapers of CoCrMo and Ti6Al4V contemporary hip replacements.

Authors:  Deborah J Hall; Robin Pourzal; Hannah J Lundberg; Mathew T Mathew; Joshua J Jacobs; Robert M Urban
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2017-08-26       Impact factor: 3.368

6.  Does Taper Angle Clearance Influence Fretting and Corrosion Damage at the Head-Stem Interface? A Matched Cohort Retrieval Study.

Authors:  Sevi B Kocagöz; Richard J Underwood; Shiril Sivan; Jeremy L Gilbert; Daniel W Macdonald; Judd S Day; Steven M Kurtz
Journal:  Semin Arthroplasty       Date:  2013-12-01

7.  Five Hundred Fifty-five Retrieved Metal-on-metal Hip Replacements of a Single Design Show a Wide Range of Wear, Surface Features, and Histopathologic Reactions.

Authors:  Sang-Hyun Park; Zhen Lu; Robert S Hastings; Patricia A Campbell; Edward Ebramzadeh
Journal:  Clin Orthop Relat Res       Date:  2018-02       Impact factor: 4.176

Review 8.  Trunnion Corrosion in Total Hip Arthroplasty-Basic Concepts.

Authors:  Kenneth L Urish; Nicholas John Giori; Jack E Lemons; William M Mihalko; Nadim Hallab
Journal:  Orthop Clin North Am       Date:  2019-04-16       Impact factor: 2.472

9.  Diagnosis and Management of Adverse Reactions to Metal Debris.

Authors:  Richard A Wawrose; Kenneth L Urish
Journal:  Oper Tech Orthop       Date:  2019-10-01

Review 10.  Do retrieval analysis and blood metal measurements contribute to our understanding of adverse local tissue reactions?

Authors:  Patricia A Campbell; Michael S Kung; Andrew R Hsu; Joshua J Jacobs
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176