Literature DB >> 12151892

A multicenter retrieval study of the taper interfaces of modular hip prostheses.

Jay R Goldberg1, Jeremy L Gilbert, Joshua J Jacobs, Thomas W Bauer, Wayne Paprosky, Sue Leurgans.   

Abstract

A multicenter retrieval analysis of 231 modular hip implants was done to investigate the effects of material combination, metallurgic condition, flexural rigidity, head and neck moment arm, neck length, and implantation time on corrosion and fretting of modular taper surfaces. Scores for corrosion and fretting were assigned to medial, lateral, anterior, and posterior quadrants of the necks, and proximal and distal regions of the heads. Neck and head corrosion and fretting scores were found to be significantly higher for mixed alloy versus similar alloy couples. Moderate to severe corrosion was observed in 28% of the heads of similar alloy couples and 42% of the heads of mixed alloy couples. Differences in corrosion scores were observed between components made from the same base alloy, but of different metallurgic conditions. Corrosion and fretting scores tended to be higher for heads than necks. Implantation time and flexural rigidity of the neck were predictors of head and neck corrosion and head fretting. The results of this study suggest that in vivo corrosion of modular hip taper interfaces is attributable to a mechanically-assisted crevice corrosion process. Larger diameter necks will increase neck stiffness and may reduce fretting and subsequent corrosion of the taper interface regardless of the alloy used. Increasing neck diameter must be balanced, however, with the resulting loss of range of motion and joint stability.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12151892     DOI: 10.1097/00003086-200208000-00018

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  105 in total

1.  Revision for taper corrosion at the neck-body junction following total hip arthroplasty: pearls and pitfalls.

Authors:  Mitchell C Weiser; Darwin D Chen
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

2.  Is increased modularity associated with increased fretting and corrosion damage in metal-on-metal total hip arthroplasty devices?: a retrieval study.

Authors:  Genymphas B Higgs; Josa A Hanzlik; Daniel W MacDonald; Jeremy L Gilbert; Clare M Rimnac; Steven M Kurtz
Journal:  J Arthroplasty       Date:  2013-08-01       Impact factor: 4.757

3.  Taper design affects failure of large-head metal-on-metal total hip replacements.

Authors:  Nader A Nassif; Danyal H Nawabi; Kirsten Stoner; Marcella Elpers; Timothy Wright; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

4.  Stability and trunnion wear potential in large-diameter metal-on-metal total hips: a finite element analysis.

Authors:  Jacob M Elkins; John J Callaghan; Thomas D Brown
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

5.  Modelling changes in modular taper micromechanics due to surgeon assembly technique in total hip arthroplasty.

Authors:  Jonathan A Gustafson; Robin Pourzal; Brett R Levine; Joshua J Jacobs; Hannah J Lundberg
Journal:  Bone Joint J       Date:  2020-07       Impact factor: 5.082

Review 6.  Catastrophic Femoral Neck Failure after THA with the Accolade(®) I Stem in Three Patients.

Authors:  Jonathon Spanyer; Jennifer Hines; Christopher Maxwell Beaumont; Jonathan Yerasimides
Journal:  Clin Orthop Relat Res       Date:  2015-07-08       Impact factor: 4.176

Review 7.  What is the trouble with trunnions?

Authors:  Christina I Esposito; Timothy M Wright; Stuart B Goodman; Daniel J Berry
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

8.  Diagnosis and Management of Adverse Reactions to Metal Debris.

Authors:  Richard A Wawrose; Kenneth L Urish
Journal:  Oper Tech Orthop       Date:  2019-10-01

Review 9.  [Endoprostheses in the elderly : Biomaterials, implant selection and fixation technique].

Authors:  M M Morlock; M Jäger
Journal:  Orthopade       Date:  2017-01       Impact factor: 1.087

10.  Case reports: fractures of threaded cups: rare complications of a well-established implant.

Authors:  Arnd T Hoburg; Manav Mehta; Stephan Tohtz; Carsten Perka
Journal:  Clin Orthop Relat Res       Date:  2008-09-27       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.