Ilona Piotrowiak1, Beata Soldanska1, Mateusz Burduk1, Bartlomiej J Kaluzny1, Jozef Kaluzny1.
Abstract
Background and Objective. Evaluation of agreement, repeatability, and reproducibility of central and minimal corneal thickness (CCT and MCT) measurements obtained by SOCT, the Scheimpflug system, and ultrasound pachymetry. Materials and Methods. 28 eyes of healthy patients were enrolled. Pachymetry measurements were performed with SOCT, the Scheimpflug system, and ultrasound instrument. Each measurement was taken by 3 operators on 3 devices providing a total of 2100 measurements. Results. The mean CCT for SOCT, Scheimpflug system, and ultrasound instrument was 537.92, 545.94, and 555.74 μm, respectively, (P < .001). The respective mean coefficients of repeatability for CCT were 0.61, 0.82 and 0.80, whereas mean coefficients of interoperator reproducibility for CCT were 0.91, 1.11, and 1.25. Conclusions. CCT and MCT measurements show moderate agreement between instruments. The repeatability and interoperator reproducibility of the results obtained by SOCT are somewhat higher. The operator's impact on CCT and MCT measurements is insignificant in all devices.Entities:
Year: 2012 PMID: 24558594 PMCID: PMC3914255 DOI: 10.5402/2012/869319
Source DB: PubMed Journal: ISRN Ophthalmol ISSN: 2090-5688
Figure 1Mean CCT and MCT values obtained with 3 different devices.
Figure 2Mean CCT and MCT obtained with different instruments and operators.
Mean standard deviation for CCT and MCT measurements obtained with different instruments.
| Instrument | Mean standard deviation | |
|---|---|---|
| CCT | MCT | |
| RTVue 100 FD | ±3.29 | ±2.17 |
| Pentacam HR | ±4.47 | ±4.41 |
| Pachette 3 | ±4.44 | — |
Coefficient of repeatability of CCT and MCT measurements for different instruments and operators.
| Coefficient of repeatability | Operator | System | ||
|---|---|---|---|---|
| RTVue 100 FD | Pentacam HR | Pachette 3 | ||
| CCT | 1 | 0.63% | 0.81% | 0.66% |
| 2 | 0.60% | 0.85% | 0.88% | |
| 3 | 0.61% | 0.80% | 0.85% | |
|
| ||||
| MCT | 1 | 0.42% | 0.81% | — |
| 2 | 0.41% | 0.89% | — | |
| 3 | 0.40% | 0.73% | — | |
Coefficient of interoperator reproducibility of CCT and MCT measurements for different instruments and operators.
| Coefficient of reproducibility | Operator | System | ||
|---|---|---|---|---|
| RTVue 100 FD | Pentacam HR | Pachette 3 | ||
| CCT | 1 and 2 | 0.92% | 1.00% | 1.25% |
| 1 and 3 | 1.08% | 1.27% | 1.05% | |
| 2 and 3 | 0.72% | 1.06% | 1.44% | |
|
| ||||
| MCT | 1 and 2 | 1.00% | 0.95% | — |
| 1 and 3 | 1.16% | 1.55% | — | |
| 2 and 3 | 0.58% | 1.31% | — | |