| Literature DB >> 24995686 |
Ilona Piotrowiak1, Bartłomiej Jan Kaluzny1, Beata Danek1, Adam Chwiędacz2, Bartosz Lukasz Sikorski1, Grażyna Malukiewicz1.
Abstract
BACKGROUND: This study aimed to evaluate anterior segment spectral optical coherence tomography (AS SOCT) for assessing the lens-to-cornea fit of rigid gas-permeable (RGP) lenses. The results were verified with the fluorescein pattern method, considered the criterion standard for RGP lens alignment evaluations.Entities:
Mesh:
Substances:
Year: 2014 PMID: 24995686 PMCID: PMC4095701 DOI: 10.12659/MSM.890269
Source DB: PubMed Journal: Med Sci Monit ISSN: 1234-1010
Figure 1Schematic of apical clearance evaluated with the fluorescein pattern method. Grey region indicates the area of no contact between a rigid gas-permeable contact lens and the cornea. BCR: base curve radius of the trial contact lens.
Figure 2Box plot of apical clearance evaluated with anterior segment spectral optical coherence tomography (AS SOCT) measurements. Values indicate the distance (μm) of no contact between a rigid gas-permeable contact lens and the cornea. BCR: base curve radius of the trial contact lens.
Figure 3Examples of alignment fitting of trial rigid gas-permeable contact lenses evaluated with fluorescein pattern images and spectral optical coherence tomography (SOCT) measurements. Four trial lenses, each with a different base curve radius (BCR), were evaluated in the same eye with either (A): the fluorescein pattern or (B): SOCT measurements (microns ±SD).
Number of cases with apical clearance detected by fluorescein pattern (FP) and anterior segment spectral coherence optical tomography (AS SOCT).
| BCR –0.1 mm | BCR –0.2 mm | BCR –0.3 mm | |
|---|---|---|---|
| Total number of measurements | 26 | 26 | 20 |
| Apical clearance in FP (sensitivity) | 24 (92.31%) | 26 (100%) | 20 (100%) |
| Apical clearance in SOCT (sensitivity) | 20 (76.92%) | 25 (96.15%) | 20 (100%) |
Sensitivity of these methods is given in brackets. BCR – base curve radius.