Tobias A Fuchs1, Julia Stehli, Sacha Bull, Svetlana Dougoud, Olivier F Clerc, Bernhard A Herzog, Ronny R Buechel, Oliver Gaemperli, Philipp A Kaufmann.
Abstract
AIMS: To evaluate the feasibility and image quality of coronary computed tomography angiography (CCTA) acquisition with a submillisievert fraction of effective radiation dose using model-based iterative reconstruction (MBIR) for noise reduction. METHODS ANDEntities:
Keywords: Feasibility; Model-based iterative reconstruction; Ultra-low-dose coronary computed tomography angiography
Mesh:
Year: 2014 PMID: 24553723 PMCID: PMC4006092 DOI: 10.1093/eurheartj/ehu053
Source DB: PubMed Journal: Eur Heart J ISSN: 0195-668X Impact factor: 29.983
Body mass index adapted tube current and voltage
| BMI (kg/m2) | Ultra-low-dose CCTA | Standard CCTA | ||
|---|---|---|---|---|
| Tube current (mA) | Tube voltage (kV) | Tube current (mA) | Tube voltage (kV) | |
| <22.5 | 150 | 80 | 450 | 100 |
| 22.5–24.9 | 165 | 80 | 500 | 100 |
| 25–27.4 | 180 | 80 | 550 | 100 |
| 27.5–29.9 | 195 | 80 | 600 | 100 |
| 30–39.9 | 210 | 100 | 650 | 120 |
| ≥40 | 210 | 100 | 700 | 120 |
BMI, body mass index; CCTA, coronary computed tomography angiography.
Patient baseline characteristics
| Number of patients ( | 42 |
| Age (years) | 55 ± 10 (34–71) |
| Male/female | 28/14 |
| BMI (kg/m2) | 25.2 ± 3.8 (18.4–40.2) |
| Drug administration, | |
| Beta-blocker | 33 (79) |
| Nytroglycerin | 42 (100) |
| Heart rate (b.p.m) | |
| Standard CCTA | 56.9 ± 5.7 (49–70) |
| Ultra-low-dose CCTA | 56.7 ± 5.7 (49–75) |
| Clinical symptoms, | |
| Dyspnoea | 9 (21) |
| Typical AP | 10 (24) |
| Atypical chest pain | 22 (52) |
| None | 10 (24) |
| Cardiovascular risk factors, | |
| Smoking | 21 (50) |
| Diabetes | 1 (2) |
| Arterial hypertension | 13 (31) |
| Dyslipidaemia | 16 (38) |
| Positive family history | 13 (31) |
Values are given as mean ± SD and ranges (in brackets) or absolute numbers and percentages (in brackets).
BMI, body mass index; CCTA, coronary computed tomography angiography; AP, angina pectoris.
Image quality
| Ultra-low-dose CCTA | Standard CCTA | ||
|---|---|---|---|
| Per segment ( | 3.5 (2.5–4.0) | 3.5 (3.0–4.0) | 0.29 |
| Per artery ( | 3.5 (3.1–4.0) | 3.5 (3.0–4.0) | 0.84 |
| Per patient ( | 3.4 (2.9–3.6) | 3.3 (3.1–3.6) | 0.60 |
Values of image quality are given as median and IQR (in brackets).
P-values are reported from Wilcoxon signed-rank test.
CCTA, coronary computed tomography angiography.
Image quality parameters
| Ultra-low-dose CCTA | Standard CCTA | ||
|---|---|---|---|
| Noise | 20 ± 3 | 32 ± 8 | <0.001 |
| Attenuation LMA | 558 ± 121 | 456 ± 82 | <0.001 |
| Attenuation RCA | 511 ± 100 | 444 ± 72 | <0.001 |
| SNR LMA | 29 ± 7 | 15 ± 5 | <0.001 |
| SNR RCA | 27 ± 6 | 14 ± 4 | <0.001 |
Noise and attenuation are given in HU as mean ± SD.
CCTA, coronary computed tomography angiography; SNR, signal-to-noise ratio; LMA, left main artery; RCA, right coronary artery.