| Literature DB >> 24423187 |
Florian André1, Dirk Müller, Grigorios Korosoglou, Waldemar Hosch, Hans-Ulrich Kauczor, Hugo A Katus, Henning Steen.
Abstract
BACKGROUND: The important detection of in-stent restenosis in cardiovascular computed tomography (CT) is still challenging. The first study assessing the in-vitro stent lumen visualization of the state of the art 256-multislice CT (256-MSCT), which was performed by our research group, yielded promising results. As the applied technical approach is not suitable for daily routine, we assessed the capability of the 256-MSCT and its different reconstruction kernels for the coronary stent lumen visualization employing a clinically applicable technique in a phantom study.Entities:
Mesh:
Year: 2014 PMID: 24423187 PMCID: PMC3904014 DOI: 10.1186/1756-0500-7-38
Source DB: PubMed Journal: BMC Res Notes ISSN: 1756-0500
List of coronary artery stents
| Medtronic | Endeavor Resolute | 2.25 | 12 | CoCr | 0.091 | Zotarolimus | 45 |
| Medtronic | Endeavor Resolute | 2.25 | 8 | CoCr | 0.091 | Zotarolimus | 44 |
| Medtronic | Micro-Driver | 2.25 | 24 | Cobalt alloy | 0.091 | | 41 |
| Abbott Vascular | Multi-Link Mini Vision | 2.50 | 28 | Cobalt alloy | 0.081 | | 41 |
| Abbott Vascular | Multi-Link Mini Vision | 2.50 | 18 | Cobalt alloy | 0.081 | | 40 |
| Biotronik | PRO-Kinetic | 2.50 | 22 | CoCr + SiC coating | 0.060 | | 53 |
| Cordis | Cypher Select Plus | 2.50 | 18 | 316 L | 0.140 | Sirolimus | 43 |
| Medtronic | Endeavor Resolute | 2.50 | 12 | CoCr | 0.091 | Zotarolimus | 36 |
| Medtronic | Endeavor Resolute | 2.50 | 24 | CoCr | 0.091 | Zotarolimus | 32 |
| Medtronic | Micro-Driver | 2.50 | 24 | Cobalt alloy | 0.091 | | 36 |
| Terumo | Tsunami Gold | 2.50 | 20 | 316 L | 0.080 | | 43 |
| Abbott Vascular | Multi-Link Vision | 2.75 | 8 | CoCr | 0.081 | | 36 |
| Medtronic | Endeavor Resolute | 2.75 | 8 | CoCr | 0.091 | Zotarolimus | 33 |
| Medtronic | Endeavor Resolute | 2.75 | 30 | CoCr | 0.091 | Zotarolimus | 35 |
| Medtronic | Endeavor Sprint | 2.75 | 24 | CoCr | 0.091 | Zotarolimus | 33 |
| Medtronic | Micro-Driver | 2.75 | 18 | Cobalt alloy | 0.091 | | 33 |
| Medtronic | Micro-Driver | 2.75 | 24 | Cobalt alloy | 0.091 | | 36 |
| B. Braun | Coroflex Blue | 3.00 | 19 | CoCr | 0.065 | | 41 |
| B. Braun | Coroflex Please | 3.00 | 19 | 316 L | 0.120 | Paclitaxel | 40 |
| Biotronik | PRO-Kinetic | 3.00 | 30 | CoCr + SiC coating | 0.060 | | 50 |
| Boston Scientific | Liberté | 3.00 | 20 | 316 L | 0.097 | | 38 |
| Boston Scientific | Liberté | 3.00 | 16 | 316 L | 0.097 | | 44 |
| Boston Scientific | Taxus Liberté | 3.00 | 16 | 316 L | 0.097 | Paclitaxel | 39 |
| Boston Scientific | Taxus Liberté | 3.00 | 20 | 316 L | 0.097 | Paclitaxel | 40 |
| Cordis | Presillion | 3.00 | 17 | CoCr | 0.073 | | 38 |
| Cordis | Presillion | 3.00 | 12 | CoCr | 0.073 | | 39 |
| Medtronic | Driver | 3.00 | 12 | Cobalt alloy | 0.091 | | 40 |
| Medtronic | Endeavor Resolute | 3.00 | 12 | CoCr | 0.091 | Zotarolimus | 44 |
| MSM | Experimental* | 3.00 | 16 | 316 L + Tantal coating | 0.080 | | 38 |
| Terumo | Tsunami Gold | 3.00 | 18 | 316 L | 0.080 | | 42 |
| Terumo | Tsunami Gold | 3.00 | 18 | 316 L | 0.080 | | 39 |
| Translumina | Yukon* | 3.00 | 18 | 316 L | 0.115 | | 41 |
| Translumina | Yukon Choice* | 3.00 | 18 | 316 L | 0.097 | | 40 |
| Translumina | Yukon Choice CC* | 3.00 | 18 | 316 L | 0.100 | | 41 |
| Abbott Vascular | Multi-Link Vision | 3.50 | 28 | CoCr | 0.081 | | 25 |
| Abbott Vascular | Multi-Link Vision | 3.50 | 12 | CoCr | 0.081 | | 33 |
| Biotronik | PRO-Kinetic | 3.50 | 20 | CoCr + SiC coating | 0.080 | | 27 |
| Cordis | Cypher Select Plus | 3.50 | 13 | 316 L | 0.140 | Sirolimus | 32 |
| Cordis | Cypher Select Plus | 3.50 | 18 | 316 L | 0.140 | Sirolimus | 30 |
| Medtronic | Driver | 3.50 | 9 | Cobalt alloy | 0.091 | | 30 |
| Medtronic | Endeavor | 3.50 | 9 | CoCr | 0.091 | | 30 |
| Medtronic | Endeavor Resolute | 3.50 | 24 | CoCr | 0.091 | Zotarolimus | 22 |
| Medtronic | Endeavor Resolute | 3.50 | 9 | CoCr | 0.091 | Zotarolimus | 25 |
| Medtronic | Endeavor Sprint | 3.50 | 24 | CoCr | 0.091 | Zotarolimus | 21 |
| Medtronic | Endeavor Sprint | 3.50 | 18 | CoCr | 0.091 | Zotarolimus | 25 |
| Terumo | Tsunami Gold | 3.50 | 10 | 316 L | 0.080 | | 33 |
| Terumo | Tsunami Gold | 3.50 | 18 | 316 L | 0.080 | | 30 |
| Biotronik | PRO-Kinetic | 4.00 | 20 | CoCr + SiC coating | 0.080 | | 29 |
| Medtronic | Driver | 4.00 | 9 | Cobalt alloy | 0.091 | | 29 |
| Medtronic | Driver | 4.00 | 24 | Cobalt alloy | 0.091 | | 33 |
| Medtronic | Driver | 4.00 | 12 | Cobalt alloy | 0.091 | 31 |
*stent without a delivery system which was expanded using a balloon catheter.
List of different manufacturers, stent types, dimensions, materials and drugs of the 51 applied coronary artery stents. In addition, the ALN values of every stent using the XCD kernel are provided.
∅: nominal stent diameter.
L: stent length.
CoCr: cobalt chrome.
SiC: silicon carbide.
ST: strut thickness.
ALN: artificial lumen narrowing (given as percentage of the nominal lumen diameter).
Artificial lumen narrowing
| All | 36(21/53) ± 7 | 45(26/79) ± 11 | 45(29/71) ± 10 | 42(26/73) ± 10 |
| A | 39(32/53) ± 6 | 52(37/79) ± 12 | 51(39/71) ± 9 | 48(37/73) ± 10 |
| B | 41(38/58) ± 3 | 50(44/54) ± 3 | 49(36/57) ± 6 | 46(40/53) ± 4 |
| C | 29(21/33) ± 4 | 35(26/44) ± 5 | 34(29/46) ± 5 | 32(26/42) ± 5 |
Artificial lumen narrowing values are given as mean (minimum/maximum) ± standard deviation in percent of the nominal lumen diameter.
Figure 1Differences of reconstruction kernels. Reconstructions of the same stent applying the four different kernels. (B. Braun Coroflex Please, nominal diameter 3.0 mm, length 19 mm).
Figure 2Effect of different stent diameter. Reconstructions of three different sizes of the same stent type (top: Biotronik PRO-Kinetic, bottom: Medtronic Endeavor Resolute). Diameter is given in mm. Zoom factors differ between the stent types.
Figure 3Artificial lumen narrowing. Comparison of the ALN values of different stent groups applying the four reconstruction kernels.
Figure 4Example of the halo effect. The ALN for this stent was manually measured as 45%. However, the lumen visualization was considerably impaired by the halo effect (Medtronic Endeavor Resolute, nominal diameter 2.25 mm, length 12 mm).
Attenuation
| All | 43(22/174)IQR34 | 18(-177/392)IQR183 | 40(-78/254)IQR96 | 128(6/396)IQR122 |
| A | 55(22/174)IQR55 | 127(14/392)IQR186 | 78(14/254)IQR34 | 189(116/396)IQR158 |
| B | 41(27/125)IQR48 | 36(-2/151)IQR65 | 48(9/153)IQR56 | 131(87/265)IQR59 |
| C | 42(22/116)IQR17 | -113(-177/-18)IQR35 | -29(-78/33)IQR29 | 49(6/163)IQR33 |
Attenuation values are given as median (minimum/maximum) interquartile range in HU.
Figure 5Attenuation. Attenuation values as provided by the different reconstruction kernels. Note that with increasing stent size the values of the CC and CD kernels become negative.