Literature DB >> 12664124

Imaging of coronary artery stents using multislice computed tomography: in vitro evaluation.

David Maintz1, Kai-Uwe Juergens, Thomas Wichter, Matthias Grude, Walter Heindel, Roman Fischbach.   

Abstract

The aim of this study was to evaluate imaging features of different coronary artery stents during multislice CT Angiography (MSCTA). Nineteen stents made of varying material (steel, nitinol, tantalum) and of varying stent design were implanted in plastic tubes with an inner diameter of 3 mm to simulate a coronary artery. The tubes were filled with iodinated contrast material diluted to 200 Hounsfield units (HU), closed at both ends and positioned in a plastic container filled with oil (-70 HU). The MSCT scans were obtained perpendicular to the stent axes (detector collimation 4x1 mm, table feed 2 mm/rotation, 300 mAs, 120 kV). Axial images and multiplanar reformations were evaluated regarding artifact size, lumen visibility, and intraluminal attenuation values. Artifacts characterized by artifactual thickening of the stent struts leading to apparent reduction in the lumen diameter and increased intraluminal attenuation values were observed in all cases. The stent lumen was totally obscured in the Wiktor stent, the Wallgraft stent, and the Nir Royal stent. Partial residual of the stent lumen could be visualized in all other utilized stent products (artificial lumen reductions ranged from 62% in the V-Flex stent to 94% in the Bx Velocity stent). Parts of the stent lumen can be visualized in most coronary artery stents; however, detectability of in-stent stenoses remains to be evaluated for each stent type.

Entities:  

Mesh:

Year:  2002        PMID: 12664124     DOI: 10.1007/s00330-002-1651-4

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  25 in total

1.  Small vessel stents for intracranial angioplasty: in vitro evaluation of in-stent stenoses using CT angiography.

Authors:  M Trossbach; M Hartmann; C Braun; K Sartor; S Hähnel
Journal:  Neuroradiology       Date:  2004-05-01       Impact factor: 2.804

2.  Assessment of coronary artery stent patency by electron-beam CT.

Authors:  Friedrich D Knollmann; Jörg Möller; Axel Gebert; Christian Bethge; Roland Felix
Journal:  Eur Radiol       Date:  2004-06-03       Impact factor: 5.315

3.  MDCT of the proximal anastomoses created by nitinol implants in coronary artery bypass grafting: a retrospective two-observer evaluation.

Authors:  M Fawad Khan; Christopher Herzog; Kai Landenberger; Sven Martens; Adel Maataoui; Hanns Ackermann; Markus Dietrich; Anton Moritz; Thomas J Vogl
Journal:  Eur Radiol       Date:  2004-11-24       Impact factor: 5.315

4.  Assessment of coronary artery stents using 16-slice MDCT angiography: evaluation of a dedicated reconstruction kernel and a noise reduction filter.

Authors:  Harald Seifarth; Rainer Raupach; Stefan Schaller; Eva Maria Fallenberg; Thomas Flohr; Walter Heindel; Roman Fischbach; David Maintz
Journal:  Eur Radiol       Date:  2005-02-12       Impact factor: 5.315

Review 5.  Coronary artery stents: identification and evaluation.

Authors:  J Butany; K Carmichael; S W Leong; M J Collins
Journal:  J Clin Pathol       Date:  2005-08       Impact factor: 3.411

6.  Evaluation of in-stent restenosis in proximal coronary arteries with multidetector computed tomography (MDCT).

Authors:  V Chabbert; D Carrie; M Bennaceur; E Maupas; V Lauwers; M Mhem; T Lhermusier; M Elbaz; F Joffre; H Rousseau; J Puel
Journal:  Eur Radiol       Date:  2006-11-18       Impact factor: 5.315

7.  Coronary artery stent geometry and in-stent contrast attenuation with 64-slice computed tomography.

Authors:  Tiziano Schepis; Pascal Koepfli; Sebastian Leschka; Lotus Desbiolles; Lars Husmann; Oliver Gaemperli; Franz R Eberli; Simon Wildermuth; Borut Marincek; Thomas F Lüscher; Hatem Alkadhi; Philipp A Kaufmann
Journal:  Eur Radiol       Date:  2007-01-06       Impact factor: 5.315

8.  Preoperative assessment and follow-up of congenital abnormalities of the pulmonary arteries using CT and MRI.

Authors:  Sebastian Ley; Julia Zaporozhan; Raoul Arnold; Joachim Eichhorn; Jens-Peter Schenk; Herbert Ulmer; Karl-Friedrich Kreitner; Hans-Ulrich Kauczor
Journal:  Eur Radiol       Date:  2006-06-24       Impact factor: 5.315

9.  CT angiographic appearance of in-stent restenosis of intracranial arteries treated with the Wingspan stent.

Authors:  A S Turk; H A Rowley; D B Niemann; D Fiorella; B Aagaard-Kienitz; K Pulfer; C M Strother
Journal:  AJNR Am J Neuroradiol       Date:  2007-09-24       Impact factor: 3.825

10.  High-resolution ex vivo imaging of coronary artery stents using 64-slice computed tomography--initial experience.

Authors:  Carsten Rist; Konstantin Nikolaou; Thomas Flohr; Bernd J Wintersperger; Maximilian F Reiser; Christoph R Becker
Journal:  Eur Radiol       Date:  2006-03-02       Impact factor: 5.315