In vitro evaluation of 56 coronary artery stents by 256-slice multi-detector coronary CT.

Henning Steen1, Florian André, Grigorios Korosoglou, Dirk Mueller, Waldemar Hosch, Hans-Ulrich Kauczor, Evangelos Giannitsis, Hugo A Katus.   

Abstract

OBJECTIVE: We sought to investigate stent lumen visibility of 56 coronary stents with the newest 256-multi-slice-CT (256-MDCT) technology for different reconstruction algorithms in an in vitro model.
BACKGROUND: Early identification of in-stent restenosis (ISR) is important to avoid recurrent ischemia and prevent acute myocardial infarction (AMI). Since angiography has the disadvantage of high costs and its invasiveness, MDCT could be a convenient and safe non-invasive alternative for detection of ISR.
MATERIAL AND METHODS: Percentages of in-stent lumen diameter and in-stent signal attenuation (measured as contrast-to-noise ratio (CNR)) of 56 coronary stents (group A ≤2.5mm; group B=2.75-3.0mm; group C=3.5-4.0mm) were evaluated in a coronary vessel in vitro phantom (iodine-filled plastic tubes) employing four different reconstruction algorithms (XCD, CC, CD, XCB) on a novel 256-MDCT (Philips-iCT, collimation=128 mm × 0.625 mm; rotation time=270 ms; tube current=800 mAs with 120 kV). Analysis was conducted with the semi-automatical full-width-at-half-maximum (FWHM) method. P-values <0.05 were regarded statistically significant.
RESULTS: In-stent lumen diameter >60% for group C stents was significantly larger and CNR was significantly lower (both p<0.05) for sharp kernels (CD; XCD) when compared to groups A/B. The FWHM-method showed significantly smaller in-stent lumen diameter (p<0.05) when compared to the manual method.
CONCLUSION: 256-MDCT could potentially be employed for clinical assessment of stent patency in stents >3.0mm when analysed with cardio-dedicated sharp kernels, although clinical studies corroborating this claim should be performed. However, stents ≤3.0mm reconstructed by soft kernels revealed insufficient in-stent lumen visualisation and should not be used in clinical practice. Further improvements in spatial and temporal image resolution as well as reductions of radiation exposure and image noise have to be accomplished for the ambitious goal of characterising both CT coronary artery anatomy and in-stent lumen.
Copyright © 2010 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20817435     DOI: 10.1016/j.ejrad.2010.08.007

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  4 in total

1.  Off-resonance magnetic resonance angiography improves visualization of in-stent lumen in peripheral nitinol stents compared to conventional T1-weighted acquisitions: an in vitro comparison study.

Authors:  Gitsios Gitsioudis; Philipp Fortner; Matthias Stuber; Anna Missiou; Florian Andre; Oliver J Müller; Hugo A Katus; Grigorios Korosoglou
Journal:  Int J Cardiovasc Imaging       Date:  2016-08-17       Impact factor: 2.357

2.  The superiority of 256-slice spiral computed tomography angiography for preoperative evaluation of surrounding arteries in patients with gastric cancer.

Authors:  Deqing Wu; Linyong Zhao; Ying Liu; Junjiang Wang; Weixian Hu; Xingyu Feng; Zejian Lv; Yong Li; Xueqing Yao
Journal:  Onco Targets Ther       Date:  2017-02-16       Impact factor: 4.147

3.  In-vitro assessment of coronary artery stents in 256-multislice computed tomography angiography.

Authors:  Florian André; Dirk Müller; Grigorios Korosoglou; Waldemar Hosch; Hans-Ulrich Kauczor; Hugo A Katus; Henning Steen
Journal:  BMC Res Notes       Date:  2014-01-14

4.  Coronary Stent Artifact Reduction with an Edge-Enhancing Reconstruction Kernel - A Prospective Cross-Sectional Study with 256-Slice CT.

Authors:  Stéphanie Tan; Gilles Soulez; Patricia Diez Martinez; Sandra Larrivée; Louis-Mathieu Stevens; Yves Goussard; Samer Mansour; Carl Chartrand-Lefebvre
Journal:  PLoS One       Date:  2016-04-29       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.