High-pitch 128-slice dual-source CT for the assessment of coronary stents in a phantom model.

Olivio F Donati1, Matthias C Burg, Lotus Desbiolles, Christoph Karlo, Paul Stolzmann, Alexander Bunck, Stephan Baumueller, Borut Marincek, Harald Seifarth, Hatem Alkadhi, Sebastian Leschka, David Maintz.   

Abstract

RATIONALE AND
OBJECTIVES: To evaluate the quality of stent lumen delineation using dual-source computed tomography (DSCT) in the standard-pitch mode (SP) as compared to the high-pitch mode (HP) in a phantom study.
MATERIALS AND METHODS: Forty different coronary stents placed in plastic tubes filled with contrast agent were imaged with a second generation DSCT system in a SP (pitch 0.23) and HP (pitch 3.4) mode in orientations of 0°, 45°, and 90° relative to the z-axis. Two observers independently measured the in-stent lumen and the attenuation values in the center of the stents. The artificial lumen narrowing (ALN) was calculated using the measured in-stent lumen and the nominal diameter of the plastic tube.
RESULTS: Interobserver correlation was excellent for in-stent lumen (0.86) and attenuation measurements (0.91). There was no significant difference neither for ALN (SP: 54.7-62.8%; HP: 55.8-64.0%) nor attenuation (SP: 356-395 Hounsfield units [HU]; HP: 352-384 HU) between SP and HP mode. For both modes, the orientation of the stent relative to the z-axis significantly affected ALN and attenuation (each P < .001). CT volume dose index was significantly lower using HP mode as compared to SP mode (P < .001).
CONCLUSION: The HP mode in DSCT provides visualization of the coronary in-stent lumen comparable to that measured in SP mode while reducing applied radiation dose in a stationary phantom model.
Copyright © 2010 AUR. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20801697     DOI: 10.1016/j.acra.2010.06.017

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  5 in total

1.  Influence of cardiac motion on stent lumen visualization in third generation dual-source CT employing a pulsatile heart model.

Authors:  Nils Petri; Tobias Gassenmaier; Thomas Allmendinger; Thomas Flohr; Wolfram Voelker; Thorsten A Bley
Journal:  Br J Radiol       Date:  2016-11-29       Impact factor: 3.039

2.  Next generation coronary CT angiography: in vitro evaluation of 27 coronary stents.

Authors:  Tobias Gassenmaier; Nils Petri; Thomas Allmendinger; Thomas Flohr; David Maintz; Wolfram Voelker; Thorsten A Bley
Journal:  Eur Radiol       Date:  2014-07-21       Impact factor: 5.315

3.  Dual-Source CT Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent Types.

Authors:  Michael Köhler; Matthias C Burg; Alexander C Bunck; Walter Heindel; Harald Seifarth; David Maintz
Journal:  Radiol Res Pract       Date:  2011-07-18

4.  A systematic approach for analysis, interpretation, and reporting of coronary CTA studies.

Authors:  Christoph A Karlo; Sebastian Leschka; Paul Stolzmann; Nicola Glaser-Gallion; Simon Wildermuth; Hatem Alkadhi
Journal:  Insights Imaging       Date:  2012-05-01

5.  In-vitro assessment of coronary artery stents in 256-multislice computed tomography angiography.

Authors:  Florian André; Dirk Müller; Grigorios Korosoglou; Waldemar Hosch; Hans-Ulrich Kauczor; Hugo A Katus; Henning Steen
Journal:  BMC Res Notes       Date:  2014-01-14
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.