In vitro reproduction of device deformation leading to thrombotic complications and failure of flow diversion.

Laurent Estrade1, Alina Makoyeva, Tim E Darsaut, Jimmy Ghostine, Evgueni Kouznetsov, Igor Salazkin, Daniel Roy, Alain Weill, Jean Raymond.   

Abstract

Flow diverters (FDs) are increasingly used for complex intracranial aneurysms. As these self-expanding devices are deployed across an aneurysm neck, they can undergo deformations. The potential clinical consequences of FD deformations remain unclear. We describe an immediate thrombotic complication attributed to a stereotypical stenotic deformation of an FD extremity that can occur when landing zones are of insufficient length. This case is supplemented with in vitro studies showing the relationship between i) the length of the landing zones and ii) discrepancies between the diameter of the device and recipient vessel, and the severity of FD stenosis. In vitro, a shorter landing zone was associated with a progressive stenotic deformation of the terminal ends of all FDs studied. This deformation was more pronounced when the diameter of the device was oversized compared to the size of the recipient tube. In our clinical case, the presence of this deformation led to an immediate thrombotic complication, requiring deployment of a second stent to correct the observed stenosis. In addition, treatment failure ultimately led to a fatal rupture, a failure that can be explained by residual flows through a more porous transition zone, another characteristic FD deformation which occurs when they are oversized as compared to the parent vessel, but free to expand at the level of the aneurysm. Proper selection of device diameter and length of the landing zone is important, and may decrease the incidence of deformation-related complications.

Entities:  

Keywords:  aneurysm; complication; deformation; flow diverter; in-stent stenosis

Mesh:

Year:  2013        PMID: 24355146      PMCID: PMC3902741          DOI: 10.1177/159101991301900405

Source DB:  PubMed          Journal:  Interv Neuroradiol        ISSN: 1591-0199            Impact factor:   1.610


  13 in total

1.  Intra-aneurysmal thrombosis as a possible cause of delayed aneurysm rupture after flow-diversion treatment.

Authors:  Z Kulcsár; E Houdart; A Bonafé; G Parker; J Millar; A J P Goddard; S Renowden; G Gál; B Turowski; K Mitchell; F Gray; M Rodriguez; R van den Berg; A Gruber; H Desal; I Wanke; D A Rüfenacht
Journal:  AJNR Am J Neuroradiol       Date:  2010-11-11       Impact factor: 3.825

2.  The varying porosity of braided self-expanding stents and flow diverters: an experimental study.

Authors:  A Makoyeva; F Bing; T E Darsaut; I Salazkin; J Raymond
Journal:  AJNR Am J Neuroradiol       Date:  2012-08-09       Impact factor: 3.825

3.  The silk flow-diverting stent in the endovascular treatment of complex intracranial aneurysms: technical aspects and midterm results in 24 consecutive patients.

Authors:  Olli I Tähtinen; Hannu I Manninen; Ritva L Vanninen; Janne Seppänen; Tero Niskakangas; Jaakko Rinne; Leo Keski-Nisula
Journal:  Neurosurgery       Date:  2012-03       Impact factor: 4.654

4.  Stents and flow diverters in the treatment of aneurysms: device deformation in vivo may alter porosity and impact efficacy.

Authors:  Fabrice Bing; Tim E Darsaut; Igor Salazkin; Alina Makoyeva; Guylaine Gevry; Jean Raymond
Journal:  Neuroradiology       Date:  2012-08-16       Impact factor: 2.804

5.  Immediate and midterm results following treatment of unruptured intracranial aneurysms with the pipeline embolization device.

Authors:  W McAuliffe; V Wycoco; H Rice; C Phatouros; T J Singh; J Wenderoth
Journal:  AJNR Am J Neuroradiol       Date:  2011-10-06       Impact factor: 3.825

6.  Treatment of intracranial aneurysms by functional reconstruction of the parent artery: the Budapest experience with the pipeline embolization device.

Authors:  I Szikora; Z Berentei; Z Kulcsar; M Marosfoi; Z S Vajda; W Lee; A Berez; P K Nelson
Journal:  AJNR Am J Neuroradiol       Date:  2010-02-11       Impact factor: 3.825

7.  Rescue, combined, and stand-alone thrombectomy in the management of large vessel occlusion stroke using the solitaire device: a prospective 50-patient single-center study: timing, safety, and efficacy.

Authors:  Vincent Costalat; Paolo Machi; Kyriakos Lobotesis; Igor Maldonado; Jean François Vendrell; Carlos Riquelme; Isabelle Mourand; Didier Milhaud; Chérif Héroum; Pierre-François Perrigault; Caroline Arquizan; Alain Bonafé
Journal:  Stroke       Date:  2011-05-19       Impact factor: 7.914

8.  Mechanisms of occlusion and recanalization in canine carotid bifurcation aneurysms embolized with platinum coils: an alternative concept.

Authors:  J Raymond; T Darsaut; I Salazkin; G Gevry; F Bouzeghrane
Journal:  AJNR Am J Neuroradiol       Date:  2008-01-17       Impact factor: 3.825

9.  Definitive reconstruction of circumferential, fusiform intracranial aneurysms with the pipeline embolization device.

Authors:  David Fiorella; Henry H Woo; Felipe C Albuquerque; Peter K Nelson
Journal:  Neurosurgery       Date:  2008-05       Impact factor: 4.654

10.  Curative endovascular reconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience.

Authors:  Pedro Lylyk; Carlos Miranda; Rosana Ceratto; Angel Ferrario; Esteban Scrivano; Hugh Ramirez Luna; Aaron L Berez; Quang Tran; Peter K Nelson; David Fiorella
Journal:  Neurosurgery       Date:  2009-04       Impact factor: 4.654

View more
  13 in total

Review 1.  Flow-diverter devices in the treatment of intracranial aneurysms: A meta-analysis and systematic review.

Authors:  Xianli Lv; Hongchao Yang; Peng Liu; Youxiang Li
Journal:  Neuroradiol J       Date:  2016-02-02

2.  Y-crossing of braided stents with stents and flow diverters does not cause significant stenosis in bench-top studies.

Authors:  Alina Makoyeva; Tim E Darsaut; Igor Salazkin; Jean Raymond
Journal:  Interv Neuroradiol       Date:  2013-12-18       Impact factor: 1.610

3.  Filament inadequate wall apposition of the different ends of flow diverters in the abdominal aorta of rabbits.

Authors:  Kuizhong Wang; Peng Jin; Peigang Lu; Qiang Liu; Bo Li; Zhenqiang Hao
Journal:  Neuroradiol J       Date:  2019-11-27

4.  Late recurrence of a completely occluded large intracranial aneurysm treated with a Tubridge flow diverter.

Authors:  Xiaoxi Zhang; Nan Lv; Chi Wang; Wei Cao; Jianmin Liu; Qinghai Huang
Journal:  BMJ Case Rep       Date:  2016-06-21

5.  Re-treatment of residual aneurysms after flow diversion: An experimental study.

Authors:  Robert Fahed; Tim E Darsaut; Marc Kotowski; Igor Salazkin; Jean Raymond
Journal:  Neuroradiol J       Date:  2018-03-07

6.  Deployment of flow diverter devices: prediction of foreshortening and validation of the simulation in 18 clinical cases.

Authors:  Robert Kellermann; Steffen Serowy; Oliver Beuing; Martin Skalej
Journal:  Neuroradiology       Date:  2019-08-31       Impact factor: 2.804

7.  Is there an association between flow diverter fish mouthing and delayed-type hypersensitivity to metals?-a case-control study.

Authors:  Naci Kocer; Prabath Kumar Mondel; Elif Yamac; Ayse Kavak; Osman Kizilkilic; Civan Islak
Journal:  Neuroradiology       Date:  2017-09-05       Impact factor: 2.804

Review 8.  Flow diversion: what can clinicians learn from animal models?

Authors:  Robert Fahed; Tim E Darsaut; Jean-Christophe Gentric; Behzad Farzin; Igor Salazkin; Guylaine Gevry; Jean Raymond
Journal:  Neuroradiology       Date:  2017-02-27       Impact factor: 2.804

9.  Evaluation of occurring complications after flow diverter treatment of elastase-induced aneurysm in rabbits using micro-CT and MRI at 9.4 T.

Authors:  Andreas Simgen; Désirée Ley; Christian Roth; Giorgio Franco Maria Cattaneo; Ruben Mühl-Benninghaus; Andreas Müller; Heiko Körner; Yoo-Jin Kim; Bruno Scheller; Wolfgang Reith; Umut Yilmaz
Journal:  Neuroradiology       Date:  2016-07-25       Impact factor: 2.804

10.  Virtual-versus-Real Implantation of Flow Diverters: Clinical Potential and Influence of Vascular Geometry.

Authors:  P Bouillot; O Brina; H Yilmaz; M Farhat; G Erceg; K-O Lovblad; M I Vargas; Z Kulcsar; V M Pereira
Journal:  AJNR Am J Neuroradiol       Date:  2016-06-30       Impact factor: 3.825

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.