Literature DB >> 21866071

The silk flow-diverting stent in the endovascular treatment of complex intracranial aneurysms: technical aspects and midterm results in 24 consecutive patients.

Olli I Tähtinen1, Hannu I Manninen, Ritva L Vanninen, Janne Seppänen, Tero Niskakangas, Jaakko Rinne, Leo Keski-Nisula.   

Abstract

BACKGROUND: The flow-diverting stent is a new option in endovascular therapy specifically designed for the endovascular reconstruction of a segmentally diseased artery. The safety of flow-diverting stents is still equivocal.
OBJECTIVE: To evaluate the technical aspects, thromboembolic events, adjunctive therapies, and midterm results in patients with complex intracranial aneurysms treated with a flow-diverting stent (Silk; Balt Extrusion, Montmorency, France).
METHODS: We retrospectively examined angiographic images and clinical reports of 24 consecutive patients (29 stents) treated (n = 23) or attempted to treat (n = 1) with a flow-diverting device in 2 Finnish centers between March 2009 and October 2010.
RESULTS: The primary technical success rate was 67% (16/24). Adjunctive therapies were required in 6 (25%) patients, including 4 cases where intra-arterial abciximab was administered for the treatment of intraprocedural thromboembolic events. Technique-related complication rate and the 30-day mortality rate were each 4% (1/24). Follow-up imaging revealed 1 case of delayed in-stent thrombosis resulting in permanent disability of the patient, 1 asymptomatic occlusion, and 1 asymptomatic stenosis of the stented artery. Complete occlusion of the aneurysm with fully patent parent artery was observed in 16 of the 23 aneurysms (70%) where follow-up images were available.
CONCLUSION: Many previously untreatable cerebral aneurysms may be successfully treated with the Silk flow-diverting stent, but the associated risk of thromboembolic events is justifiable only if conventional endovascular or surgical treatment options are not applicable. Perioperative thromboembolic events should be prepared for and treated without unnecessary delays because they frequently respond to adjunctive medical therapy.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 21866071     DOI: 10.1227/NEU.0b013e31823387d4

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  23 in total

Review 1.  Endovascular treatment of cerebral aneurysms using flow-diverter devices: A systematic review.

Authors:  Francesco Briganti; Giuseppe Leone; Mariano Marseglia; Giuseppe Mariniello; Ferdinando Caranci; Arturo Brunetti; Francesco Maiuri
Journal:  Neuroradiol J       Date:  2015-08-27

2.  Silk flow-diverter stent for the treatment of intracranial aneurysms: a series of 58 patients with emphasis on long-term results.

Authors:  B Lubicz; O Van der Elst; L Collignon; B Mine; F Alghamdi
Journal:  AJNR Am J Neuroradiol       Date:  2014-11-06       Impact factor: 3.825

3.  Flow diversion in the treatment of carotid injury and carotid-cavernous fistula after transsphenoidal surgery.

Authors:  Daniela Iancu; Cheemum Lum; Muhammad E Ahmed; Rafael Glikstein; Marlise P Dos Santos; Howard Lesiuk; Mohamed Labib; Amin B Kassam
Journal:  Interv Neuroradiol       Date:  2015-05-26       Impact factor: 1.610

4.  In vitro reproduction of device deformation leading to thrombotic complications and failure of flow diversion.

Authors:  Laurent Estrade; Alina Makoyeva; Tim E Darsaut; Jimmy Ghostine; Evgueni Kouznetsov; Igor Salazkin; Daniel Roy; Alain Weill; Jean Raymond
Journal:  Interv Neuroradiol       Date:  2013-12-18       Impact factor: 1.610

5.  Influencing factors of immediate angiographic results in intracranial aneurysms patients after endovascular treatment.

Authors:  Jia-Xiang Chen; Ling-Feng Lai; Kuang Zheng; Guo-Xiong Li; Xu-Ying He; Liang-Ping Li; Chuan-Zhi Duan
Journal:  J Neurol       Date:  2015-06-24       Impact factor: 4.849

6.  International retrospective study of the pipeline embolization device: a multicenter aneurysm treatment study.

Authors:  D F Kallmes; R Hanel; D Lopes; E Boccardi; A Bonafé; S Cekirge; D Fiorella; P Jabbour; E Levy; C McDougall; A Siddiqui; I Szikora; H Woo; F Albuquerque; H Bozorgchami; S R Dashti; J E Delgado Almandoz; M E Kelly; R Turner; B K Woodward; W Brinjikji; G Lanzino; P Lylyk
Journal:  AJNR Am J Neuroradiol       Date:  2014-10-29       Impact factor: 3.825

7.  Treatment of complex intracranial aneurysms using flow-diverting silk® stents. An analysis of 32 consecutive patients.

Authors:  Ramazan Buyukkaya; Hasan Kocaeli; Nalan Yildirim; Hakan Cebeci; Cüneyt Erdogan; Bahattin Hakyemez
Journal:  Interv Neuroradiol       Date:  2014-12-05       Impact factor: 1.610

8.  Stent-assisted embolization of recurrent or residual intracranial aneurysms.

Authors:  Olli I Tähtinen; Hannu I Manninen; Ritva L Vanninen; Riitta Rautio; Arto Haapanen; Janne Seppänen; Tero Niskakangas; Jaakko Rinne; Leo Keski-Nisula
Journal:  Neuroradiology       Date:  2013-07-17       Impact factor: 2.804

9.  A novel flow-diverting device (Tubridge) for the treatment of 28 large or giant intracranial aneurysms: a single-center experience.

Authors:  Y Zhou; P-F Yang; Y-B Fang; Y Xu; B Hong; W-Y Zhao; Q Li; R Zhao; Q-H Huang; J-M Liu
Journal:  AJNR Am J Neuroradiol       Date:  2014-04-10       Impact factor: 3.825

10.  Meta-Analysis of the Efficiency and Prognosis of Intracranial Aneurysm Treated with Flow Diverter Devices.

Authors:  Gengfan Ye; Meng Zhang; Lin Deng; Xiaohui Chen; Yunyan Wang
Journal:  J Mol Neurosci       Date:  2016-01-27       Impact factor: 3.444

View more