Literature DB >> 27457491

Evaluation of occurring complications after flow diverter treatment of elastase-induced aneurysm in rabbits using micro-CT and MRI at 9.4 T.

Andreas Simgen1, Désirée Ley2, Christian Roth3, Giorgio Franco Maria Cattaneo4, Ruben Mühl-Benninghaus2, Andreas Müller5, Heiko Körner2, Yoo-Jin Kim6, Bruno Scheller7, Wolfgang Reith2, Umut Yilmaz2.   

Abstract

INTRODUCTION: Flow diverters are increasingly being used to treat intracranial aneurysms. This study evaluates occurring complications of flow-diverting devices in the treatment of experimental aneurysms, involving the use of micro-CT and small animal MRI at 9.4 T, in correlation to angiographic and histological findings.
METHODS: We previously published two preclinical studies, in which we assessed two different flow diverters in the treatment of elastase-induced aneurysms. Devices have been implanted across the aneurysm neck as well as in the abdominal aorta. From these studies, a total of 65 devices (prototype FD (n = 30) and Derivo embolization device (n = 35)) additionally underwent micro-CT and MRI after angiographic follow-up and before being histologically examined.
RESULTS: The different architectures of both devices were precisely comparable due to high-resolution micro-CT imaging. Micro-CT revealed wire fractures in nine cases (30 %) only with the prototype FD. In three cases (10 %), severe wire fractures correlated with an in-stent stenosis due to intimal hyperplasia. Other complications, like distal stent occlusions and post-stent stenosis, were seen in both groups and verified with both imaging techniques. Osseous metaplasia were correlated to calcifications seen with micro-CT. MRI enabled visualization of the position of the implanted devices relative to the aneurysm and revealed incomplete aneurysm neck coverage with the prototype FD in two cases (6.7 %).
CONCLUSION: Micro-CT and 9.4-T MRI are valid to discover and understand occurring complications of flow diverters in the preclinical phase and can serve as evaluation tools to minimize complication rates of endovascular devices in the future.

Entities:  

Keywords:  9.4-T MRI; Aneurysm; Animal studies; Flow diverter; Micro-CT

Mesh:

Substances:

Year:  2016        PMID: 27457491     DOI: 10.1007/s00234-016-1730-x

Source DB:  PubMed          Journal:  Neuroradiology        ISSN: 0028-3940            Impact factor:   2.804


  35 in total

1.  Treatment of intra-cranial aneurysms with the SILK flow diverter: 2 years' experience with 28 patients at a single center.

Authors:  Shimon Maimon; Lior Gonen; Erez Nossek; Ido Strauss; Ronen Levite; Zvi Ram
Journal:  Acta Neurochir (Wien)       Date:  2012-03-09       Impact factor: 2.216

2.  Flow-diverter silk stent for the treatment of intracranial aneurysms: 1-year follow-up in a multicenter study.

Authors:  J Berge; A Biondi; P Machi; H Brunel; L Pierot; J Gabrillargues; K Kadziolka; X Barreau; V Dousset; A Bonafé
Journal:  AJNR Am J Neuroradiol       Date:  2012-02-02       Impact factor: 3.825

3.  Intra-aneurysmal thrombosis as a possible cause of delayed aneurysm rupture after flow-diversion treatment.

Authors:  Z Kulcsár; E Houdart; A Bonafé; G Parker; J Millar; A J P Goddard; S Renowden; G Gál; B Turowski; K Mitchell; F Gray; M Rodriguez; R van den Berg; A Gruber; H Desal; I Wanke; D A Rüfenacht
Journal:  AJNR Am J Neuroradiol       Date:  2010-11-11       Impact factor: 3.825

Review 4.  The SILK flow diverter in the treatment of intracranial aneurysms.

Authors:  Santosh B Murthy; Shreyansh Shah; Aditi Shastri; Chethan P Venkatasubba Rao; Eric M Bershad; Jose I Suarez
Journal:  J Clin Neurosci       Date:  2013-10-13       Impact factor: 1.961

5.  Endovascular creation of an in vivo bifurcation aneurysm model in rabbits.

Authors:  H J Cloft; T A Altes; W F Marx; R J Raible; S B Hudson; G A Helm; J W Mandell; M E Jensen; J E Dion; D F Kallmes
Journal:  Radiology       Date:  1999-10       Impact factor: 11.105

6.  Fracture of a carotid stent: an unexpected complication.

Authors:  Arif Rahim Valibhoy; Bibombe Patrice Mwipatayi; Kishore Sieunarine
Journal:  J Vasc Surg       Date:  2007-03       Impact factor: 4.268

7.  Evaluation of a newly designed flow diverter for the treatment of intracranial aneurysms in an elastase-induced aneurysm model, in New Zealand white rabbits.

Authors:  Andreas Simgen; Desiree Ley; Christian Roth; Umut Yilmaz; Heiko Körner; Ruben Mühl-Benninghaus; Yoo-Jin Kim; Bruno Scheller; Wolfgang Reith
Journal:  Neuroradiology       Date:  2013-11-14       Impact factor: 2.804

8.  Stenting for carotid artery stenosis: fractures, proposed etiology and the need for surveillance.

Authors:  Adrian James Ling; Patrice Mwipatayi; Tarun Gandhi; Kishore Sieunarine
Journal:  J Vasc Surg       Date:  2008-04-28       Impact factor: 4.268

9.  Aneurysm rupture after endovascular flow diversion: the possible role of persistent flows through the transition zone associated with device deformation.

Authors:  T E Darsaut; E Rayner-Hartley; A Makoyeva; I Salazkin; F Berthelet; J Raymond
Journal:  Interv Neuroradiol       Date:  2013-05-21       Impact factor: 1.610

Review 10.  Stent usage in the treatment of intracranial aneurysms: past, present and future.

Authors:  Bradley A Gross; Kai U Frerichs
Journal:  J Neurol Neurosurg Psychiatry       Date:  2012-11-08       Impact factor: 10.154

View more
  6 in total

1.  Intravenous versus intra-arterial digital subtraction angiography: Occlusion rate and complication assessment of experimental aneurysms after flow diverter treatment in rabbits.

Authors:  Andreas Simgen; Toshiki Tomori; Hagen Bomberg; Umut Yilmaz; Christian Roth; Wolfgang Reith; Ruben Mühl-Bennighaus
Journal:  Interv Neuroradiol       Date:  2018-11-04       Impact factor: 1.610

2.  Retrieval of Migrated Volume Coils Using Different Clot Retrievers in a Porcine Model.

Authors:  Andreas Simgen; Toshiki Tomori; Ruben Mühl-Benninghaus; Hagen Bomberg; Umut Yilmaz; Heiko Körner; Matthias W Laschke; Michael D Menger; Wolfgang Reith
Journal:  Clin Neuroradiol       Date:  2017-06-01       Impact factor: 3.649

3.  Downstream vascular changes after flow-diverting device deployment in a rabbit model.

Authors:  Praveen Kolumam Parameswaran; Daying Dai; Yong-Hong Ding; Matthew W Urban; Logan Manlove; Venkatachalem Sathish; Juan R Cebral; David F Kallmes; Ramanathan Kadirvel
Journal:  J Neurointerv Surg       Date:  2018-11-10       Impact factor: 5.836

4.  Analysis of Flow Dynamics and Outcomes of Cerebral Aneurysms Treated with Intrasaccular Flow-Diverting Devices.

Authors:  J R Cebral; B J Chung; F Mut; J Chudyk; C Bleise; E Scrivano; P Lylyk; R Kadirvel; D Kallmes
Journal:  AJNR Am J Neuroradiol       Date:  2019-08-08       Impact factor: 3.825

5.  Solitaire Stentectomy Using a Stent-Retriever Technique in a Porcine Model.

Authors:  Andreas Simgen; Michael Kettner; Frida Juliane Webelsiep; Toshiki Tomori; Ruben Mühl-Benninghaus; Umut Yilmaz; Pervinder Bhogal; Matthias W Laschke; Michael D Menger; Wolfgang Reith; Philipp Dietrich
Journal:  Clin Neuroradiol       Date:  2020-05-18       Impact factor: 3.649

6.  Different Rescue Approaches of Migrated Woven Endobridge (WEB) Devices: an Animal Study.

Authors:  Andreas Simgen; Michael Kettner; Philipp Dietrich; Toshiki Tomori; Ruben Mühl-Benninghaus; Pervinder Bhogal; Matthias W Laschke; Michael D Menger; Wolfgang Reith; Umut Yilmaz
Journal:  Clin Neuroradiol       Date:  2020-03-12       Impact factor: 3.649

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.