| Literature DB >> 24149962 |
Víctor Díaz1, Pedro José Benito, Ana Belén Peinado, María Alvarez, Carlos Martín, Valter Di Salvo, Fabio Pigozzi, Nicola Maffulli, Fracisco Javier Calderón.
Abstract
We tested a new portable metabolic system, the Jaeger Oxycon Mobile (OM) at a range of running speeds. Six subjects carried out, in random order, two incremental tests on a treadmill, one of them using the OM, and the other using the Jaeger Oxycon Pro (OP). There are systematic errors in the measurements of oxygen consumption (VO2) and respiratory exchange ratio (RER) with the OM. Production of CO2 (VCO2) tends to be overestimated by the OM, although the differences are not significant. Ventilation (VE) showed very similar values in both analyzers. Data of VO2 and RER were corrected with a regression equation which minimised the differences among the devices. The portable metabolic system OM makes systematic errors in measurements of VO2 and RER which can be adjusted with a regression analysis to obtain data comparable to those obtained by fixed systems. Key pointsPortable metabolic systems are frequently used to explore various physiological ventilatory variables in field tests.There are systematic errors in the measurements of oxygen consumption (VO2) and respiratory exchange ratio (RER) with the Jaeger Oxycon Mobile (OM) portable metabolic systemProduction of CO2 (VCO2) tends to be overestimated by the OMData of VO2 and RER can be corrected with a regression equationThe portable metabolic system OM makes systematic errors in measurements of VO2 and RER which can be adjusted with a regression analysis to obtain data comparable to those obtained by fixed systems.Entities:
Keywords: Portable metabolic chart; accuracy; consistency; oxygen consumption; running
Year: 2008 PMID: 24149962 PMCID: PMC3761920
Source DB: PubMed Journal: J Sports Sci Med ISSN: 1303-2968 Impact factor: 2.988