Evaluation of the Cosmed K2 portable telemetric oxygen uptake analyzer.

C G Crandall1, S L Taylor, P B Raven.   

Abstract

The purpose of this study was to determine the validity and evaluate the accuracy of a portable telemetric oxygen uptake analyzer (K2). Two experiments were carried out: a) using a mechanical lung, the accuracy of the K2 to measure oxygen fractions and minute ventilation following 10 and 60 min of warm-up was determined; and b) two maximal graded exercise tests (GXT) on 15 subjects, one with the K2 system and the other with a standardized breath-by-breath (BBB) system, while heart rate (HR), minute ventilation (VE), and oxygen uptake (VO2) were compared. Following 10-min warm-up prior to calibration, the K2 underestimated the true oxygen fraction as early as 5 min into the test, and this value continued to decrease throughout the 30-min test. After 60 min of warm-up prior to calibration, the K2 accurately measured the true oxygen fraction for the first 15 min; at minute 20, and on to minute 30, the K2 underestimated the oxygen fraction. Ventilation volumes were not affected by warm-up time. Minute ventilation during the K2 GXT was significantly higher than VE for the BBB test. No significant differences were found between the HRs obtained with the BBB or K2 systems. No differences in VO2 for any stage of the GXT were identified between the K2 device, BBB device or when a respiratory exchange ratio (RER) correction factor was applied to the K2 derived values. However, the RER correction factor did minimize the VO2 differences between the BBB and K2 systems. Therefore, we conclude that the K2 accurately measures VO2 during a GXT; however, its accuracy can be compromised by limitations inherent to the system.

Entities:  

Mesh:

Year:  1994        PMID: 8133730

Source DB:  PubMed          Journal:  Med Sci Sports Exerc        ISSN: 0195-9131            Impact factor:   5.411


  7 in total

Review 1.  Automated metabolic gas analysis systems: a review.

Authors:  D J Macfarlane
Journal:  Sports Med       Date:  2001       Impact factor: 11.136

2.  Development of a sports specific aerobic capacity test for karate - a pilot study.

Authors:  David Nunan
Journal:  J Sports Sci Med       Date:  2006-07-01       Impact factor: 2.988

3.  Comparison of gas exchange data using the Aquatrainer system and the facemask with Cosmed K4b2 during exercise in healthy subjects.

Authors:  Mathieu Gayda; Laurent Bosquet; Martin Juneau; Thibaut Guiraud; Jean Lambert; Anil Nigam
Journal:  Eur J Appl Physiol       Date:  2010-01-05       Impact factor: 3.078

4.  Physical activity energy expenditure measured using a portable telemetric device in comparison with a mass spectrometer.

Authors:  C Maiolo; G Melchiorri; L Iacopino; S Masala; A De Lorenzo
Journal:  Br J Sports Med       Date:  2003       Impact factor: 13.800

5.  Validation of a new portable metabolic system during an incremental running test.

Authors:  Víctor Díaz; Pedro José Benito; Ana Belén Peinado; María Alvarez; Carlos Martín; Valter Di Salvo; Fabio Pigozzi; Nicola Maffulli; Fracisco Javier Calderón
Journal:  J Sports Sci Med       Date:  2008-12-01       Impact factor: 2.988

6.  Assessing Cardiorespiratory Fitness of Soccer Players: Is Test Specificity the Issue?-A Review.

Authors:  Monèm Jemni; Mohammad Shoaib Prince; Julien S Baker
Journal:  Sports Med Open       Date:  2018-06-19

7.  Validity of a Multi-Sensor Armband for Estimating Energy Expenditure during Eighteen Different Activities.

Authors:  Paige Dudley; David R Bassett; Dinesh John; Scott E Crouter
Journal:  J Obes Weight Loss Ther       Date:  2012-08-29
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.