Cardiac imaging at 7 Tesla: Single- and two-spoke radiofrequency pulse design with 16-channel parallel excitation.

Sebastian Schmitter1, Lance DelaBarre, Xiaoping Wu, Andreas Greiser, Dingxin Wang, Edward J Auerbach, J Thomas Vaughan, Kâmil Uğurbil, Pierre-François Van de Moortele.   

Abstract

PURPOSE: Higher signal to noise ratio (SNR) and improved contrast have been demonstrated at ultra-high magnetic fields (≥7 Tesla [T]) in multiple targets, often with multi-channel transmit methods to address the deleterious impact on tissue contrast due to spatial variations in B1 (+) profiles. When imaging the heart at 7T, however, respiratory and cardiac motion, as well as B0 inhomogeneity, greatly increase the methodological challenge. In this study we compare two-spoke parallel transmit (pTX) RF pulses with static B1 (+) shimming in cardiac imaging at 7T.
METHODS: Using a 16-channel pTX system, slice-selective two-spoke pTX pulses and static B1 (+) shimming were applied in cardiac CINE imaging. B1 (+) and B0 mapping required modified cardiac triggered sequences. Excitation homogeneity and RF energy were compared in different imaging orientations.
RESULTS: Two-spoke pulses provide higher excitation homogeneity than B1 (+) shimming, especially in the more challenging posterior region of the heart. The peak value of channel-wise RF energy was reduced, allowing for a higher flip angle, hence increased tissue contrast. Image quality with two-spoke excitation proved to be stable throughout the entire cardiac cycle.
CONCLUSION: Two-spoke pTX excitation has been successfully demonstrated in the human heart at 7T, with improved image quality and reduced RF pulse energy when compared with B1 (+) shimming.
Copyright © 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  7T; RF pulse design; cardiac CINE imaging; pTX; parallel transmission; spoke excitation

Mesh:

Year:  2013        PMID: 24038314      PMCID: PMC3960017          DOI: 10.1002/mrm.24935

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  35 in total

1.  7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images.

Authors:  J T Vaughan; M Garwood; C M Collins; W Liu; L DelaBarre; G Adriany; P Andersen; H Merkle; R Goebel; M B Smith; K Ugurbil
Journal:  Magn Reson Med       Date:  2001-07       Impact factor: 4.668

2.  Transmit SENSE.

Authors:  Ulrich Katscher; Peter Börnert; Christoph Leussler; Johan S van den Brink
Journal:  Magn Reson Med       Date:  2003-01       Impact factor: 4.668

3.  Efficient high-frequency body coil for high-field MRI.

Authors:  J T Vaughan; G Adriany; C J Snyder; J Tian; T Thiel; L Bolinger; H Liu; L DelaBarre; K Ugurbil
Journal:  Magn Reson Med       Date:  2004-10       Impact factor: 4.668

4.  Parallel excitation with an array of transmit coils.

Authors:  Yudong Zhu
Journal:  Magn Reson Med       Date:  2004-04       Impact factor: 4.668

5.  B(1) destructive interferences and spatial phase patterns at 7 T with a head transceiver array coil.

Authors:  Pierre-François Van de Moortele; Can Akgun; Gregor Adriany; Steen Moeller; Johannes Ritter; Christopher M Collins; Michael B Smith; J Thomas Vaughan; Kāmil Uğurbil
Journal:  Magn Reson Med       Date:  2005-12       Impact factor: 4.668

6.  Spatial domain method for the design of RF pulses in multicoil parallel excitation.

Authors:  William Grissom; Chun-yu Yip; Zhenghui Zhang; V Andrew Stenger; Jeffrey A Fessler; Douglas C Noll
Journal:  Magn Reson Med       Date:  2006-09       Impact factor: 4.668

7.  Exploring the limits of RF shimming for high-field MRI of the human head.

Authors:  Weihua Mao; Michael B Smith; Christopher M Collins
Journal:  Magn Reson Med       Date:  2006-10       Impact factor: 4.668

8.  Fast-kz three-dimensional tailored radiofrequency pulse for reduced B1 inhomogeneity.

Authors:  Suwit Saekho; Chun-yu Yip; Douglas C Noll; Fernando E Boada; V Andrew Stenger
Journal:  Magn Reson Med       Date:  2006-04       Impact factor: 4.668

9.  High frequency volume coils for clinical NMR imaging and spectroscopy.

Authors:  J T Vaughan; H P Hetherington; J O Otu; J W Pan; G M Pohost
Journal:  Magn Reson Med       Date:  1994-08       Impact factor: 4.668

10.  Inversion recovery at 7 T in the human myocardium: measurement of T(1), inversion efficiency and B(1) (+).

Authors:  Christopher T Rodgers; Stefan K Piechnik; Lance J Delabarre; Pierre-François Van de Moortele; Carl J Snyder; Stefan Neubauer; Matthew D Robson; J Thomas Vaughan
Journal:  Magn Reson Med       Date:  2012-11-29       Impact factor: 4.668

View more
  20 in total

Review 1.  Magnetic resonance imaging at ultrahigh fields.

Authors:  Kamil Ugurbil
Journal:  IEEE Trans Biomed Eng       Date:  2014-03-25       Impact factor: 4.538

Review 2.  RF pulse methods for use with surface coils: Frequency-modulated pulses and parallel transmission.

Authors:  Michael Garwood; Kamil Uğurbil
Journal:  J Magn Reson       Date:  2018-04-26       Impact factor: 2.229

3.  Quadrature transmit array design using single-feed circularly polarized patch antenna for parallel transmission in MR imaging.

Authors:  Yong Pang; Baiying Yu; Daniel B Vigneron; Xiaoliang Zhang
Journal:  Quant Imaging Med Surg       Date:  2014-02

4.  Mitigating transmit B 1 inhomogeneity in the liver at 7T using multi-spoke parallel transmit RF pulse design.

Authors:  Xiaoping Wu; Sebastian Schmitter; Edward J Auerbach; Kâmil Uğurbil; Pierre-François Van de Moortele
Journal:  Quant Imaging Med Surg       Date:  2014-02

5.  First in-vivo human imaging at 10.5T: Imaging the body at 447 MHz.

Authors:  Xiaoxuan He; M Arcan Ertürk; Andrea Grant; Xiaoping Wu; Russell L Lagore; Lance DelaBarre; Yiğitcan Eryaman; Gregor Adriany; Eddie J Auerbach; Pierre-François Van de Moortele; Kâmil Uğurbil; Gregory J Metzger
Journal:  Magn Reson Med       Date:  2019-12-17       Impact factor: 4.668

Review 6.  Imaging at ultrahigh magnetic fields: History, challenges, and solutions.

Authors:  Kamil Uğurbil
Journal:  Neuroimage       Date:  2017-07-08       Impact factor: 6.556

7.  Design of parallel transmission radiofrequency pulses robust against respiration in cardiac MRI at 7 Tesla.

Authors:  Sebastian Schmitter; Xiaoping Wu; Kâmil Uğurbil; Pierre-François Van de Moortele
Journal:  Magn Reson Med       Date:  2014-11-19       Impact factor: 4.668

Review 8.  Advanced flow MRI: emerging techniques and applications.

Authors:  M Markl; S Schnell; C Wu; E Bollache; K Jarvis; A J Barker; J D Robinson; C K Rigsby
Journal:  Clin Radiol       Date:  2016-03-02       Impact factor: 2.350

9.  A 16-channel combined loop-dipole transceiver array for 7 Tesla body MRI.

Authors:  M Arcan Ertürk; Alexander J E Raaijmakers; Gregor Adriany; Kâmil Uğurbil; Gregory J Metzger
Journal:  Magn Reson Med       Date:  2016-02-17       Impact factor: 4.668

10.  Towards high-resolution 4D flow MRI in the human aorta using kt-GRAPPA and B1+ shimming at 7T.

Authors:  Sebastian Schmitter; Susanne Schnell; Kâmil Uğurbil; Michael Markl; Pierre-François Van de Moortele
Journal:  J Magn Reson Imaging       Date:  2016-02-03       Impact factor: 4.813

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.