Towards high-resolution 4D flow MRI in the human aorta using kt-GRAPPA and B1+ shimming at 7T.

Sebastian Schmitter1, Susanne Schnell2, Kâmil Uğurbil1, Michael Markl2,3, Pierre-François Van de Moortele1.   

Abstract

PURPOSE: To evaluate the feasibility of aortic 4D flow magnetic resonance imaging (MRI) at 7T with improved spatial resolution using kt-GRAPPA acceleration while restricting acquisition time and to address radiofrequency (RF) excitation heterogeneities with B1+ shimming.
MATERIALS AND METHODS: 4D flow MRI data were obtained in the aorta of eight subjects using a 16-channel transmit/receive coil array at 7T. Flow quantification and acquisition time were compared for a kt-GRAPPA accelerated (R = 5) and a standard GRAPPA (R = 2) accelerated protocol. The impact of different dynamic B1+ shimming strategies on flow quantification was investigated. Two kt-GRAPPA accelerated protocols with 1.2 × 1.2 × 1.2 mm(3) and 1.8 × 1.8 × 2.4 mm(3) spatial resolution were compared.
RESULTS: Using kt-GRAPPA, we achieved a 4.3-fold reduction in net acquisition time resulting in scan times of about 10 minutes. No significant effect on flow quantification was observed compared to standard GRAPPA with R = 2. Optimizing the B1+ fields for the aorta impacted significantly (P <  0.05) the flow quantification while specific B1+ settings were required for respiration navigators. The high-resolution protocol yielded similar flow quantification, but allowed the depiction of branching vessels.
CONCLUSION: 7T in combination with B1+ shimming allows for high-resolution 4D flow MRI acquisitions in the human aorta, while kt-GRAPPA limits total scan times without affecting flow quantification. J. Magn. Reson. Imaging 2016;44:486-499.
© 2016 Wiley Periodicals, Inc.

Entities:  

Keywords:  4D flow MRI; human aorta; kt-GRAPPA

Mesh:

Year:  2016        PMID: 26841070      PMCID: PMC4947022          DOI: 10.1002/jmri.25164

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  43 in total

1.  Flow patterns in the aortic root and the aorta studied with time-resolved, 3-dimensional, phase-contrast magnetic resonance imaging: implications for aortic valve-sparing surgery.

Authors:  John-Peder Escobar Kvitting; Tino Ebbers; Lars Wigström; Jan Engvall; Christian L Olin; Ann F Bolger
Journal:  J Thorac Cardiovasc Surg       Date:  2004-06       Impact factor: 5.209

2.  In vitro validation of phase-contrast flow measurements at 3 T in comparison to 1.5 T: precision, accuracy, and signal-to-noise ratios.

Authors:  Joachim Lotz; Rolf Döker; Ralph Noeske; Meike Schüttert; Roland Felix; Michael Galanski; Matthias Gutberlet; Gerd Peter Meyer
Journal:  J Magn Reson Imaging       Date:  2005-05       Impact factor: 4.813

3.  Parallel MRI with extended and averaged GRAPPA kernels (PEAK-GRAPPA): optimized spatiotemporal dynamic imaging.

Authors:  Bernd Jung; Peter Ullmann; Matthias Honal; Simon Bauer; Jürgen Hennig; Michael Markl
Journal:  J Magn Reson Imaging       Date:  2008-11       Impact factor: 4.813

4.  Quantification and visualization of flow in the Circle of Willis: time-resolved three-dimensional phase contrast MRI at 7 T compared with 3 T.

Authors:  P van Ooij; J J M Zwanenburg; F Visser; C B Majoie; E vanBavel; J Hendrikse; A J Nederveen
Journal:  Magn Reson Med       Date:  2012-05-22       Impact factor: 4.668

5.  Modular 32-channel transceiver coil array for cardiac MRI at 7.0T.

Authors:  Andreas Graessl; Wolfgang Renz; Fabian Hezel; Matthias A Dieringer; Lukas Winter; Celal Oezerdem; Jan Rieger; Peter Kellman; Davide Santoro; Tomasz D Lindel; Tobias Frauenrath; Harald Pfeiffer; Thoralf Niendorf
Journal:  Magn Reson Med       Date:  2013-07-31       Impact factor: 4.668

6.  Wall shear stress and flow patterns in the ascending aorta in patients with bicuspid aortic valves differ significantly from tricuspid aortic valves: a prospective study.

Authors:  Christian Meierhofer; Eike Philipp Schneider; Christine Lyko; Andrea Hutter; Stefan Martinoff; Michael Markl; Alfred Hager; John Hess; Heiko Stern; Sohrab Fratz
Journal:  Eur Heart J Cardiovasc Imaging       Date:  2012-12-09       Impact factor: 6.875

7.  Comparison of blood flow velocity quantification by 4D flow MR imaging with ultrasound at the carotid bifurcation.

Authors:  A Harloff; T Zech; F Wegent; C Strecker; C Weiller; M Markl
Journal:  AJNR Am J Neuroradiol       Date:  2013-02-14       Impact factor: 3.825

8.  Assessment of fluctuating velocities in disturbed cardiovascular blood flow: in vivo feasibility of generalized phase-contrast MRI.

Authors:  Petter Dyverfeldt; John-Peder Escobar Kvitting; Andreas Sigfridsson; Jan Engvall; Ann F Bolger; Tino Ebbers
Journal:  J Magn Reson Imaging       Date:  2008-09       Impact factor: 4.813

Review 9.  4D flow imaging with MRI.

Authors:  Zoran Stankovic; Bradley D Allen; Julio Garcia; Kelly B Jarvis; Michael Markl
Journal:  Cardiovasc Diagn Ther       Date:  2014-04

10.  Aortic 4D flow: quantification of signal-to-noise ratio as a function of field strength and contrast enhancement for 1.5T, 3T, and 7T.

Authors:  Aaron T Hess; Malenka M Bissell; Ntobeko A B Ntusi; Andrew J M Lewis; Elizabeth M Tunnicliffe; Andreas Greiser; Aurelien F Stalder; Jane M Francis; Saul G Myerson; Stefan Neubauer; Matthew D Robson
Journal:  Magn Reson Med       Date:  2014-06-16       Impact factor: 4.668

View more
  4 in total

1.  k-t accelerated aortic 4D flow MRI in under two minutes: Feasibility and impact of resolution, k-space sampling patterns, and respiratory navigator gating on hemodynamic measurements.

Authors:  Emilie Bollache; Alex J Barker; Ryan Scott Dolan; James C Carr; Pim van Ooij; Rouzbeh Ahmadian; Alex Powell; Jeremy D Collins; Julia Geiger; Michael Markl
Journal:  Magn Reson Med       Date:  2017-03-07       Impact factor: 4.668

Review 2.  Progress in Imaging the Human Torso at the Ultrahigh Fields of 7 and 10.5 T.

Authors:  Kamil Uğurbil; Pierre-Francois Van de Moortele; Andrea Grant; Edward J Auerbach; Arcan Ertürk; Russell Lagore; Jutta M Ellermann; Xiaoxuan He; Gregor Adriany; Gregory J Metzger
Journal:  Magn Reson Imaging Clin N Am       Date:  2021-02       Impact factor: 2.266

3.  Electromagnetic simulation of a 16-channel head transceiver at 7 T using circuit-spatial optimization.

Authors:  Xin Li; Jullie W Pan; Nikolai I Avdievich; Hoby P Hetherington; Joseph V Rispoli
Journal:  Magn Reson Med       Date:  2021-02-03       Impact factor: 3.737

Review 4.  Evolution of UHF Body Imaging in the Human Torso at 7T: Technology, Applications, and Future Directions.

Authors:  M Arcan Erturk; Xiufeng Li; Pierre-Fancois Van de Moortele; Kamil Ugurbil; Gregory J Metzger
Journal:  Top Magn Reson Imaging       Date:  2019-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.