Literature DB >> 25411131

Design of parallel transmission radiofrequency pulses robust against respiration in cardiac MRI at 7 Tesla.

Sebastian Schmitter1, Xiaoping Wu1, Kâmil Uğurbil1, Pierre-François Van de Moortele1.   

Abstract

PURPOSE: Two-spoke parallel transmission (pTX) radiofrequency (RF) pulses have been demonstrated in cardiac MRI at 7T. However, current pulse designs rely on a single set of B1(+)/B0 maps that may not be valid for subsequent scans acquired at another phase of the respiration cycle because of organ displacement. Such mismatches may yield severe excitation profile degradation.
METHODS: B1(+)/B0 maps were obtained, using 16 transmit channels at 7T, at three breath-hold positions: exhale, half-inhale, and inhale. Standard and robust RF pulses were designed using maps obtained at exhale only, and at multiple respiratory positions, respectively. Excitation patterns were analyzed for all positions using Bloch simulations. Flip-angle homogeneity was compared in vivo in cardiac CINE acquisitions.
RESULTS: Standard one- and two-spoke pTX RF pulses are sensitive to breath-hold position, primarily due to B1(+) alterations, with high dependency on excitation trajectory for two spokes. In vivo excitation inhomogeneity varied from nRMSE = 8.2% (exhale) up to 32.5% (inhale) with the standard design; much more stable results were obtained with the robust design with nRMSE = 9.1% (exhale) and 10.6% (inhale).
CONCLUSION: A new pTX RF pulse design robust against respiration induced variations of B1(+)/B0 maps is demonstrated and is expected to have a positive impact on cardiac MRI in breath-hold, free-breathing, and real-time acquisitions.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  7 Tesla; RF pulse design; breath-hold; cardiac MRI; pTX; respiration; spoke pulses

Mesh:

Year:  2014        PMID: 25411131      PMCID: PMC4437923          DOI: 10.1002/mrm.25512

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  43 in total

1.  Inner-volume imaging in vivo using three-dimensional parallel spatially selective excitation.

Authors:  Johannes T Schneider; Raffi Kalayciyan; Martin Haas; Sarah R Herrmann; Wolfgang Ruhm; Jürgen Hennig; Peter Ullmann
Journal:  Magn Reson Med       Date:  2012-06-21       Impact factor: 4.668

2.  Design of a radiative surface coil array element at 7 T: the single-side adapted dipole antenna.

Authors:  A J E Raaijmakers; O Ipek; D W J Klomp; C Possanzini; P R Harvey; J J W Lagendijk; C A T van den Berg
Journal:  Magn Reson Med       Date:  2011-05-31       Impact factor: 4.668

3.  kT -points: short three-dimensional tailored RF pulses for flip-angle homogenization over an extended volume.

Authors:  M A Cloos; N Boulant; M Luong; G Ferrand; E Giacomini; D Le Bihan; A Amadon
Journal:  Magn Reson Med       Date:  2011-05-16       Impact factor: 4.668

4.  Comparison between eight- and sixteen-channel TEM transceive arrays for body imaging at 7 T.

Authors:  C J Snyder; L Delabarre; S Moeller; J Tian; C Akgun; P-F Van de Moortele; P J Bolan; K Ugurbil; J T Vaughan; G J Metzger
Journal:  Magn Reson Med       Date:  2011-11-18       Impact factor: 4.668

5.  3D radial sampling and 3D affine transform-based respiratory motion correction technique for free-breathing whole-heart coronary MRA with 100% imaging efficiency.

Authors:  Himanshu Bhat; Lan Ge; Sonia Nielles-Vallespin; Sven Zuehlsdorff; Debiao Li
Journal:  Magn Reson Med       Date:  2011-01-10       Impact factor: 4.668

6.  7 Tesla (T) human cardiovascular magnetic resonance imaging using FLASH and SSFP to assess cardiac function: validation against 1.5 T and 3 T.

Authors:  J J Suttie; L Delabarre; A Pitcher; P F van de Moortele; S Dass; C J Snyder; J M Francis; G J Metzger; P Weale; K Ugurbil; S Neubauer; M Robson; T Vaughan
Journal:  NMR Biomed       Date:  2011-07-19       Impact factor: 4.044

7.  Highly accelerated real-time cardiac cine MRI using k-t SPARSE-SENSE.

Authors:  Li Feng; Monvadi B Srichai; Ruth P Lim; Alexis Harrison; Wilson King; Ganesh Adluru; Edward V R Dibella; Daniel K Sodickson; Ricardo Otazo; Daniel Kim
Journal:  Magn Reson Med       Date:  2012-08-06       Impact factor: 4.668

8.  Cardiac imaging at 7 Tesla: Single- and two-spoke radiofrequency pulse design with 16-channel parallel excitation.

Authors:  Sebastian Schmitter; Lance DelaBarre; Xiaoping Wu; Andreas Greiser; Dingxin Wang; Edward J Auerbach; J Thomas Vaughan; Kâmil Uğurbil; Pierre-François Van de Moortele
Journal:  Magn Reson Med       Date:  2013-09-10       Impact factor: 4.668

9.  Evaluation of a subject specific dual-transmit approach for improving B1 field homogeneity in cardiovascular magnetic resonance at 3T.

Authors:  Ramkumar Krishnamurthy; Amol Pednekar; Marc Kouwenhoven; Benjamin Cheong; Raja Muthupillai
Journal:  J Cardiovasc Magn Reson       Date:  2013-08-06       Impact factor: 5.364

Review 10.  Standardized cardiovascular magnetic resonance (CMR) protocols 2013 update.

Authors:  Christopher M Kramer; Jörg Barkhausen; Scott D Flamm; Raymond J Kim; Eike Nagel
Journal:  J Cardiovasc Magn Reson       Date:  2013-10-08       Impact factor: 5.364

View more
  9 in total

Review 1.  Parallel Transmission for Ultrahigh Field MRI.

Authors:  Cem M Deniz
Journal:  Top Magn Reson Imaging       Date:  2019-06

2.  IMPULSE: A scalable algorithm for design of minimum specific absorption rate parallel transmit RF pulses.

Authors:  Mihir Pendse; Riccardo Stara; Mohammad Mehdi Khalighi; Brian Rutt
Journal:  Magn Reson Med       Date:  2018-11-13       Impact factor: 4.668

3.  Motion-robust cardiac B1+ mapping at 3T using interleaved bloch-siegert shifts.

Authors:  Sebastian Weingärtner; Fabian Zimmer; Gregory J Metzger; Kâmil Uğurbil; Pierre-Francois Van de Moortele; Mehmet Akçakaya
Journal:  Magn Reson Med       Date:  2016-09-07       Impact factor: 4.668

Review 4.  Progress in Imaging the Human Torso at the Ultrahigh Fields of 7 and 10.5 T.

Authors:  Kamil Uğurbil; Pierre-Francois Van de Moortele; Andrea Grant; Edward J Auerbach; Arcan Ertürk; Russell Lagore; Jutta M Ellermann; Xiaoxuan He; Gregor Adriany; Gregory J Metzger
Journal:  Magn Reson Imaging Clin N Am       Date:  2021-02       Impact factor: 2.266

Review 5.  Parallel transmission for ultrahigh-field imaging.

Authors:  Francesco Padormo; Arian Beqiri; Joseph V Hajnal; Shaihan J Malik
Journal:  NMR Biomed       Date:  2015-05-19       Impact factor: 4.044

Review 6.  Evolution of UHF Body Imaging in the Human Torso at 7T: Technology, Applications, and Future Directions.

Authors:  M Arcan Erturk; Xiufeng Li; Pierre-Fancois Van de Moortele; Kamil Ugurbil; Gregory J Metzger
Journal:  Top Magn Reson Imaging       Date:  2019-06

7.  Extended RF shimming: Sequence-level parallel transmission optimization applied to steady-state free precession MRI of the heart.

Authors:  Arian Beqiri; Anthony N Price; Francesco Padormo; Joseph V Hajnal; Shaihan J Malik
Journal:  NMR Biomed       Date:  2017-02-14       Impact factor: 4.044

8.  Diaphragm position can be accurately estimated from the scattering of a parallel transmit RF coil at 7 T.

Authors:  Aaron T Hess; Elizabeth M Tunnicliffe; Christopher T Rodgers; Matthew D Robson
Journal:  Magn Reson Med       Date:  2017-08-03       Impact factor: 4.668

9.  Cardiac functional magnetic resonance imaging at 7T: Image quality optimization and ultra-high field capabilities.

Authors:  El-Sayed H Ibrahim; V Emre Arpinar; L Tugan Muftuler; Jadranka Stojanovska; Andrew S Nencka; Kevin M Koch
Journal:  World J Radiol       Date:  2020-10-28
  9 in total