| Literature DB >> 23761174 |
Steven M Kurtz1, Sevi B Kocagöz, Josa A Hanzlik, Richard J Underwood, Jeremy L Gilbert, Daniel W MacDonald, Gwo-Chin Lee, Michael A Mont, Matthew J Kraay, Gregg R Klein, Javad Parvizi, Clare M Rimnac.
Abstract
BACKGROUND: Previous studies regarding modular head-neck taper corrosion were largely based on cobalt chrome (CoCr) alloy femoral heads. Less is known about head-neck taper corrosion with ceramic femoral heads. QUESTIONS/PURPOSES: We asked (1) whether ceramic heads resulted in less taper corrosion than CoCr heads; (2) what device and patient factors influence taper fretting corrosion; and (3) whether the mechanism of taper fretting corrosion in ceramic heads differs from that in CoCr heads.Entities:
Mesh:
Substances:
Year: 2013 PMID: 23761174 PMCID: PMC3773155 DOI: 10.1007/s11999-013-3096-2
Source DB: PubMed Journal: Clin Orthop Relat Res ISSN: 0009-921X Impact factor: 4.176
Patient and device information corresponding to the ceramic-metal taper cohort
| Case number | Implantation time (years) | Age at insertion (years) | Sex | BMI (kg/m2) | Maximum UCLA score | Revision reason | Head size (mm) | Lateral offset (mm) | Manufacturer | Stem type | Stem alloy | Stem taper angle | Flexural rigidity 109·Nm2 |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 0.5 | 75 | M | 27 | 7 | Loosening | 36 | 41.5 | Stryker | Accolade | TMZF | V40 | 89 |
| 2 | 0.5 | 31 | M | 33 | 5 | Infection | 36 | 39 | Stryker | Accolade | TMZF | V40 | 89 |
| 3 | 0.5 | 48 | M | 42 | 2 | Loosening | 36 | 50.3 | Stryker | Accolade | TMZF | V40 | 89 |
| 4 | 0.6 | N/A | M | N/A | N/A | Loosening | 32 | 44 | Stryker | Accolade | TMZF | V40 | 89 |
| 5 | 0.6 | 39 | M | 41 | N/A | Infection | 36 | 37 | Stryker | Accolade | TMZF | V40 | 89 |
| 6 | 0.6 | 46 | F | 19 | 4 | Infection | 36 | 49 | DePuy | Corail | Ti-6Al-4V | 12/14 | 185 |
| 7 | 0.7 | 49 | M | 27 | 2 | Loosening | 32 | 40 | Zimmer | M/L Taper | Ti-6Al-4V | 12/14 | 202 |
| 8 | 0.8 | 52 | M | 38 | 6 | Loosening | 32 | 41 | Zimmer | M/L Taper | Ti-6Al-4V | 12/14 | 202 |
| 9 | 0.8 | 49 | M | 35 | 5 | Loosening | 32 | 48 | Stryker | Accolade | TMZF | V40 | 89 |
| 10 | 0.9 | 49 | M | N/A | N/A | Loosening | 32 | 44.2 | Stryker | Accolade | TMZF | V40 | 89 |
| 11 | 1.0 | 53 | M | 27 | 4 | Infection | 36 | 39 | Stryker | Omnifit | Ti-6Al-4V | V40 | 185 |
| 12 | 1.0 | 53 | M | 28 | 7 | Loosening | 36 | 33.4 | Stryker | Accolade | TMZF | V40 | 89 |
| 13 | 1.0 | 55 | M | N/A | 5 | Loosening | 36 | 39 | Stryker | Accolade | TMZF | V40 | 89 |
| 14 | 1.1 | 43 | M | 21 | 8 | Infection | 36 | 58 | DePuy | Corail | Ti-6Al-4V | 12/14 | 185 |
| 15 | 1.2 | 68 | F | 39 | 4 | Instability | 32 | 38 | Zimmer | M/L Taper | Ti-6Al-4V | 12/14 | 202 |
| 16 | 1.2 | 50 | M | 34 | 6 | Pain | 32 | 38 | DePuy | Tri-Lock | Ti-6Al-4V | 12/14 | 185 |
| 17 | 1.2 | 55 | F | 31 | 8 | Loosening | 28 | 44 | Stryker | Accolade | TMZF | V40 | 89 |
| 18 | 1.4 | 41 | M | 44 | 6 | Loosening | 32 | 34 | Stryker | Accolade | TMZF | V40 | 89 |
| 19 | 1.5 | 57 | M | 30 | 3 | Infection | 32 | 42 | Smith & Nephew | Anthology | Ti-6Al-4V | 12/14 | 185 |
| 20 | 1.6 | 48 | M | 29 | 4 | Infection | 36 | 39 | Stryker | Accolade | TMZF | V40 | 89 |
| 21 | 1.7 | N/A | M | N/A | 6 | Loosening | 32 | 49 | Stryker | Accolade | TMZF | V40 | 89 |
| 22 | 1.7 | 50 | F | N/A | N/A | Loosening | 32 | 32.1 | Stryker | Accolade | TMZF | V40 | 89 |
| 23 | 1.9 | 53 | F | 27 | N/A | Malalignment | 28 | 40 | Stryker | Accolade | TMZF | V40 | 89 |
| 24 | 1.9 | 63 | M | 28 | 4 | Loosening | 36 | 49 | Stryker | Accolade | TMZF | V40 | 89 |
| 25 | 2.0 | 53 | M | N/A | N/A | Loosening | 32 | 38 | Stryker | Accolade | TMZF | V40 | 89 |
| 26 | 2.0 | 61 | M | 28 | 6 | Loosening | 36 | 48 | Stryker | Accolade | TMZF | V40 | 89 |
| 27 | 2.1 | 50 | F | 26 | 7 | Loosening | 32 | 32 | Stryker | Accolade | TMZF | V40 | 89 |
| 28 | 2.3 | 32 | M | 34 | 10 | Loosening | 32 | 49 | Stryker | Accolade | TMZF | V40 | 89 |
| 29 | 2.3 | 72 | F | 22 | N/A | Malposition | 32 | 44 | Zimmer | Versys FMT | Ti-6Al-4V | 12/14 | 202 |
| 30 | 2.3 | 34 | M | 34 | 3 | Infection | 32 | 45 | Smith & Nephew | Echelon | CoCr | 12/14 | 358 |
| 31 | 2.4 | 71 | M | N/A | N/A | Loosening | 32 | 38 | Stryker | Accolade | TMZF | V40 | 89 |
| 32 | 2.5 | 51 | M | 29 | 2 | Infection | 32 | 40 | Smith & Nephew | Echelon | CoCr | 12/14 | 353 |
| 33 | 2.5 | 57 | F | 21 | 6 | Loosening | 28 | 36 | Encore | Linear | Ti-6Al-4V | 12/14 | 220 |
| 34 | 2.5 | 49 | M | 24 | 6 | Infection | 32 | 27 | DePuy | Tri-Lock | Ti-6Al-4V | 12/14 | 185 |
| 35 | 2.9 | N/A | F | 25 | 5 | Loosening | 32 | 41 | Zimmer | Versys FMT | Ti-6Al-4V | 12/14 | 190 |
| 36 | 3.4 | 59 | M | 28 | 6 | Loosening | 36 | 52 | Stryker | Accolade | TMZF | V40 | 89 |
| 37 | 3.5 | 48 | F | 32 | 7 | Loosening | 32 | 50 | Stryker | Accolade | TMZF | V40 | 89 |
| 38 | 3.6 | 76 | F | 22 | 8 | Periprosthetic fracture | 28 | 32 | Stryker | Accolade | TMZF | V40 | 89 |
| 39 | 4.5 | 56 | M | 52 | N/A | Loosening | 36 | 38 | Stryker | Accolade | TMZF | V40 | 89 |
| 40 | 5.1 | 42 | F | 26 | 6 | Malalignment | 28 | 44 | Wright Medical | Perfecta | Ti-6Al-4V | 12/14 | 252 |
| 41 | 5.4 | 66 | M | 28 | 7 | Loosening | 28 | 34 | Stryker | Accolade | TMZF | V40 | 89 |
| 42 | 5.6 | 46 | M | 18 | 2 | Pain | 36 | 59 | DePuy | Corail | Ti-6Al-4V | 12/14 | 179 |
| 43 | 5.9 | 58 | M | 37 | 7 | Loosening | 32 | 49 | Stryker | Accolade | TMZF | V40 | 89 |
| 44 | 6.3 | 43 | M | 27 | 10 | Osteolysis | 32 | 35 | Smith & Nephew, Richards | Spectron EF | CoCr | 12/14 | 353 |
| 45 | 6.3 | 47 | F | 32 | 8 | Infection | 32 | 39 | Stryker | Accolade | TMZF | V40 | 89 |
| 46 | 7.6 | 47 | F | 21 | N/A | Infection | 32 | 45 | DePuy | Solution | CoCr | 12/14 | 676 |
| 47 | 8.8 | 47 | F | 21 | N/A | Infection | 32 | 46 | DePuy | Solution | CoCr | 12/14 | 676 |
| 48 | 14.3 | 52 | F | 38 | N/A | Femoral loosening | 32 | 36 | Wright Medical, Dow Corning Wright | Infinity | Ti-6Al-4V | 12/14 | 226 |
| 49 | 14.6 | 64 | F | 34 | N/A | Loosening | 32 | 46 | Zimmer | Muller modified | CoCr | 12/14 | 385 |
| 50 | 17.8 | 48 | M | 33 | 8 | PE wear | 32 | 41 | Smith & Nephew, Richards | Opti-Fix | Ti-6Al-4V | 14/16 | 328 |
BMI = body mass index; N/A = not available; M = male; F = female; PE = polyethylene; CoCr = cobalt-chromium.
Patient and device information corresponding to the matched metal-metal taper cohort
| Case number | Implantation time (years) | Age at insertion (years) | Sex | BMI (kg/m2) | Maximum UCLA score | Revision reason | Head size (mm) | Lateral offset (mm) | Manufacturer | Stem type | Stem alloy | Trunnion type | Flexural rigidity 109·Nm2 |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 0.9 | 51 | F | 23 | 6 | Infection | 28 | 38 | Stryker | Accolade | TMZF | V40 | 89 |
| 2 | 0.5 | 70 | F | 32 | 3 | Loosening | 32 | 38 | Stryker | Accolade | TMZF | V40 | 89 |
| 3 | 0.6 | 79 | F | N/A | N/A | Loosening | 32 | 46 | Stryker | Accolade | TMZF | V40 | 89 |
| 4 | 0.6 | 64 | M | 31 | 7 | Loosening | 40 | 44 | Stryker | Accolade | TMZF | V40 | 89 |
| 5 | 0.7 | 46 | M | 37 | 8 | Loosening | 32 | 39 | Stryker | Accolade | TMZF | V40 | 89 |
| 6 | 1.0 | N/A | F | 35 | N/A | Infection | 32 | 44 | Zimmer | Versys FMT | Ti-6Al-4V | 12/14 | 202 |
| 7 | 0.5 | 59 | M | N/A | 8 | Infection | 36 | 44 | Zimmer | M/L Taper | Ti-6Al-4V | 12/14 | 202 |
| 8 | 0.8 | 52 | F | 28 | 6 | Infection | 36 | 48 | Zimmer | M/L Taper | Ti-6Al-4V | 12/14 | 202 |
| 9 | 1.4 | 42 | M | 27 | 5 | Infection | 32 | 58 | Stryker | Accolade | TMZF | V40 | 247 |
| 10 | 1.0 | 61 | M | 29 | 5 | Loosening | 36 | 55 | Stryker | Accolade | TMZF | V40 | 89 |
| 11 | 1.1 | 65 | F | 33 | 4 | Infection | 36 | 39 | Zimmer | Versys FMT | Ti-6Al-4V | 12/14 | 202 |
| 12 | 1.1 | 59 | F | 22 | N/A | Infection | 36 | 32 | Stryker | Accolade | TMZF | V40 | 89 |
| 13 | 0.8 | 67 | M | 38 | 4 | Infection | 36 | 33 | Stryker | Accolade | TMZF | V40 | 89 |
| 14 | 1.2 | 47 | F | 40 | 2 | Infection | 36 | 65 | DePuy | Corail | Ti-6Al-4V | 12/14 | 196 |
| 15 | 1.0 | 65 | M | 26 | 5 | Loosening | 32 | 51 | Zimmer | M/L Taper | Ti-6Al-4V | 12/14 | 202 |
| 16 | 1.3 | 67 | F | 25 | 2 | Loosening | 32 | 40 | DePuy | Tri-Lock | Ti-6Al-4V | 12/14 | 185 |
| 17 | 1.0 | 56 | M | 30 | 8 | Loosening | 40 | 45 | Stryker | Accolade | TMZF | V40 | 89 |
| 18 | 1.4 | 46 | F | N/A | 2 | Loosening | 28 | 34 | Stryker | Accolade | TMZF | V40 | 89 |
| 19 | 1.5 | 83 | M | 23 | 8 | Infection | 32 | 49 | Zimmer | M/L Taper | Ti-6Al-4V | 12/14 | 202 |
| 20 | 0.8 | 53 | F | N/A | 7 | Loosening | 32 | 41 | Stryker | Accolade | TMZF | V40 | 89 |
| 21 | 1.8 | 55 | M | 36 | 4 | Loosening | 36 | 46 | Stryker | Accolade | TMZF | V40 | 89 |
| 22 | 1.8 | 82 | M | 27 | 3 | Loosening | 32 | 33 | Stryker | Accolade | TMZF | V40 | 89 |
| 23 | 1.7 | 57 | F | 29 | 4 | Infection | 40 | 43 | Stryker | Accolade | TMZF | V40 | 89 |
| 24 | 1.9 | 57 | M | 28 | N/A | Loosening | 32 | 48 | Stryker | Accolade | TMZF | V40 | 89 |
| 25 | 1.0 | 76 | F | 31 | 4 | Loosening | 32 | 44 | Stryker | Accolade | TMZF | V40 | 89 |
| 26 | 2.2 | N/A | F | 22 | 4 | Loosening | 28 | 47 | Stryker | Accolade | TMZF | V40 | 236 |
| 27 | 2.0 | 59 | F | 43 | 6 | Loosening | 36 | 34 | Stryker | Accolade | TMZF | V40 | 89 |
| 28 | 1.9 | 45 | M | 45 | 7 | Loosening | 32 | 44 | Stryker | Accolade | TMZF | V40 | 89 |
| 29 | 2.0 | 63 | M | 35 | 6 | Infection | 32 | 49 | Zimmer | Versys FMT | Ti-6Al-4V | 12/14 | 202 |
| 30 | 2.5 | N/A | M | 23 | 2 | Infection | 40 | 42 | Zimmer | Versys Beaded Fullcoat | CoCr | 12/14 | 385 |
| 31 | 1.1 | 57 | F | 23 | N/A | Infection | 36 | 35 | Stryker | Accolade | TMZF | V40 | 89 |
| 32 | 2.5 | 76 | M | 22 | N/A | Infection | 28 | 46 | Zimmer | Harris Precoat | CoCr | 6° | 180 |
| 33 | 2.3 | 49 | F | 35 | 6 | Infection | 40 | 65 | DePuy | Corail | Ti-6Al-4V | 12/14 | 202 |
| 34 | 2.2 | 61 | F | 23 | 7 | N/A | 28 | 42 | DePuy | Tri-Lock | Ti-6Al-4V | 12/14 | 185 |
| 35 | 2.3 | 63 | F | 51 | 4 | Infection | 32 | 39 | Zimmer | Versys FMT | Ti-6Al-4V | 12/14 | 202 |
| 36 | 2.7 | 55 | M | 29 | 6 | Loosening | 32 | 50 | Stryker | Accolade | TMZF | V40 | 89 |
| 37 | 2.3 | 74 | M | 34 | 9 | Loosening | 32 | 48 | Stryker | Accolade | TMZF | V40 | 89 |
| 38 | 3.6 | 50 | M | 20 | Pain | 36 | 38 | Stryker | Accolade | TMZF | V40 | 89 | |
| 39 | 4.0 | 49 | F | 30 | 6 | Periprosthetic fracture | 28 | 39 | Stryker | Accolade | TMZF | V40 | 89 |
| 40 | 5.2 | 66 | F | 37 | N/A | N/A | 28 | 50 | Biomet | Taperloc | Ti-6Al-4V | Type I | 132 |
| 41 | 5.0 | 13 | F | 21 | 3 | Infection | 22.2 | 39 | Stryker | Accolade | TMZF | V40 | 89 |
| 42 | 5.2 | 46 | F | 21 | 7 | Periprosthetic fracture | 28 | 64 | DePuy | Corail | Ti-6Al-4V | 12/14 | 207 |
| 43 | 5.0 | 48 | M | 24 | 2 | Infection | 28 | 45 | Stryker | Accolade | TMZF | V40 | 89 |
| 44 | 5.2 | 51 | F | N/A | 2 | Loosening | 28 | 34 | Smith & Nephew | Spectron | CoCr | 12/14 | 353 |
| 45 | 9.5 | 48 | F | 21 | 3 | Infection | 36 | 35 | Stryker | Accolade | CoCr | V40 | 236 |
| 46 | 9.3 | 53 | M | 33 | 8 | Component fracture | 28 | 66 | DePuy | AML | CoCr | 14/16, or 12/14 | 374 |
| 47 | 10.1 | 51 | M | 32 | 8 | Loosening | 32 | 40 | DePuy | AML | CoCr | 14/16, or 12/14 | 676 |
| 48 | 13.1 | 64 | M | 29 | 8 | PE wear | 28 | 43 | Biomet | Taperloc | Ti-6Al-4V | Type I | 112 |
| 49 | 15.0 | 49 | M | 35 | 8 | Periprosthetic fracture | 32 | 52 | Zimmer | Anatomic | Ti-6Al-4V | 12/14 | 116 |
| 50 | 17.4 | 16 | M | 32 | 9 | Loosening | 28 | 36 | Zimmer | Harris-Galante | Ti-6Al-4V | 6° | 101 |
BMI = body mass index; N/A = not available; F = female; M = male; PE = polyethylene; CoCr = cobalt-chromium.
Fig. 1The femoral stem taper fretting and corrosion damage scores for the matched ceramic and CoCr head cohorts are shown. The damage scores were significantly lower for the ceramic cohort (p = 0.03).
Fig. 2Some examples of stem taper fretting and corrosion scores for the ceramic head cohort are shown. The median score for this cohort was 2.
Fig. 3Examples of stem taper fretting and corrosion scores for the CoCr head cohort are shown. The median score for this cohort was 3.
Fig. 4A boxplot illustrating femoral stem taper fretting and corrosion score versus stem alloy for the ceramic and metal head cohorts is presented.
Fig. 5A–ESEMs of five different design and materials for the male taper of ceramic-metal trunnions. (A) TMZF (Stryker Orthopaedics, Mahwah, NJ, USA) × 35 BEC, (B) Ti-6Al-4V (Zimmer, Inc, Warsaw, IN, USA) × 100 SEI, (C) Ti-6Al-4V (Wright Medical Technology, Inc, Arlington, TN, USA) × 220 BEC, (D) Co-Cr-Mo (DePuy Orthopaedics, Inc, Warsaw, IN, USA) × 100 BEC, (E) Co-Ni-Cr-Mo (Zimmer) × 100 BEC. SEI = secondary electron imaging; BEC = backscattered electron contrast image. A is a ground surface, whereas B–E have machining grooves present. Also shown are fretting scars and corrosion and biological debris present. For grooved implants, only the groove tips show evidence of fretting corrosion damage.
Fig. 6A–DBackscattered electron micrographs of (A) TMZF, (B) Ti-6Al-4V, (C) Co-Cr-Mo, and (D) CoNiCrMo alloy tapers used in conjunction with ceramic femoral heads. Each image shows fretting damage and some corrosion debris present. In C, the damage has a distinctly corrosion-like appearance emanating from a machining ridge.