Trunnion Failure of the Recalled Low Friction Ion Treatment Cobalt Chromium Alloy Femoral Head.

Kenneth L Urish1, Brian R Hamlin2, Anton Y Plakseychuk2, Timothy J Levison2, Genymphas B Higgs3, Steven M Kurtz4, Anthony M DiGioia2.   

Abstract

BACKGROUND: Gross trunnion failure (GTF) is a rare complication in total hip arthroplasty (THA) reported across a range of manufacturers. Specific lots of the Stryker low friction ion treatment (LFIT) anatomic cobalt chromium alloy (CoCr) V40 femoral head were recalled in August 2016. In part, the recall was based out of concerns for disassociation of the femoral head from the stem and GTF.
METHODS: We report on 28 patients (30 implants) with either GTF (n = 18) or head-neck taper corrosion (n = 12) of the LFIT CoCr femoral head and the Accolade titanium-molybdenum-zirconium-iron alloy femoral stems. All these cases were associated with adverse local tissue reactions requiring revision of the THA.
RESULTS: In our series, a conservative estimate of the incidence of failure was 4.7% (n = 636 total implanted) at 8.0 ± 1.4 years from the index procedure. Failures were associated with a high-offset 127° femoral stem neck angle and increased neck lengths; 43.3% (13 of 30) of the observed failures included implant sizes outside the voluntary recall (27.8% [5 of 18] of the GTF and 75.0% [8 of 12] of the taper corrosion cases). Serum cobalt and chromium levels were elevated (cobalt: 8.4 ± 7.0 μg/mL; chromium: 3.4 ± 3.3 μ/L; cobalt/chromium ratio: 3.7). The metal artifact reduction sequence magnetic resonance imaging demonstrated large cystic fluid collections typical with adverse local tissue reactions. During revision, a pseudotumor was observed in all cases. Pathology suggested a chronic inflammatory response. Impending GTF could be diagnosed based on aspiration of black synovial fluid and an oblique femoral head as compared with the neck taper on radiographs.
CONCLUSION: In our series of the recalled LFIT CoCr femoral head, the risk of impending GTF or head-neck taper corrosion should be considered as a potential diagnosis in a painful LFIT femoral head and Accolade titanium-molybdenum-zirconium-iron alloy THA with unknown etiology. Almost half of the failures we observed included sizes outside of the voluntary recall.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  adverse local tissue reaction (ALTR); gross trunnion failure; head-neck taper corrosion; recall; total hip arthroplasty; trunnion corrosion

Mesh:

Substances:

Year:  2017        PMID: 28478184      PMCID: PMC5572503          DOI: 10.1016/j.arth.2017.03.075

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  30 in total

1.  MRI, Retrieval Analysis, and Histologic Evaluation of Adverse Local Tissue Reaction in Metal-on-Polyethylene Total Hip Arthroplasty.

Authors:  Scott R Nodzo; Christina I Esposito; Hollis G Potter; Chitranjan S Ranawat; Timothy M Wright; Douglas E Padgett
Journal:  J Arthroplasty       Date:  2016-12-08       Impact factor: 4.757

2.  Inflammatory pseudotumor associated with femoral nerve palsy following metal-on-metal resurfacing of the hip. A case report.

Authors:  Robert A E Clayton; Ian Beggs; Donald M Salter; M Helen Grant; James T Patton; Daniel E Porter
Journal:  J Bone Joint Surg Am       Date:  2008-09       Impact factor: 5.284

3.  Catastrophic Femoral Head-Stem Trunnion Dissociation Secondary to Corrosion.

Authors:  Laura Matsen Ko; Antonia F Chen; Gregory K Deirmengian; William J Hozack; Peter F Sharkey
Journal:  J Bone Joint Surg Am       Date:  2016-08-17       Impact factor: 5.284

4.  Risk stratification algorithm for management of patients with metal-on-metal hip arthroplasty: consensus statement of the American Association of Hip and Knee Surgeons, the American Academy of Orthopaedic Surgeons, and the Hip Society.

Authors:  Young-Min Kwon; Adolph V Lombardi; Joshua J Jacobs; Thomas K Fehring; Courtland G Lewis; Miguel E Cabanela
Journal:  J Bone Joint Surg Am       Date:  2014-01-01       Impact factor: 5.284

5.  Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement.

Authors:  A J Hart; S A Sabah; A S Bandi; P Maggiore; P Tarassoli; B Sampson; J A Skinner
Journal:  J Bone Joint Surg Br       Date:  2011-10

6.  A High Prevalence of Corrosion at the Head-Neck Taper with Contemporary Zimmer Non-Cemented Femoral Hip Components.

Authors:  Brian J McGrory; Johanna MacKenzie; George Babikian
Journal:  J Arthroplasty       Date:  2015-02-19       Impact factor: 4.757

7.  Gross trunnion failure after primary total hip arthroplasty.

Authors:  Samik Banerjee; Jeffrey J Cherian; James V Bono; Steven M Kurtz; Rudolph Geesink; R Michael Meneghini; Ronald E Delanois; Michael A Mont
Journal:  J Arthroplasty       Date:  2014-11-26       Impact factor: 4.757

8.  Do serologic and synovial tests help diagnose infection in revision hip arthroplasty with metal-on-metal bearings or corrosion?

Authors:  Paul H Yi; Michael B Cross; Mario Moric; Brett R Levine; Scott M Sporer; Wayne G Paprosky; Joshua J Jacobs; Craig J Della Valle
Journal:  Clin Orthop Relat Res       Date:  2015-02       Impact factor: 4.176

9.  Metal-on-metal hip arthroplasty: does early clinical outcome justify the chance of an adverse local tissue reaction?

Authors:  Charles A Engh; Henry Ho; Charles A Engh
Journal:  Clin Orthop Relat Res       Date:  2009-09-01       Impact factor: 4.176

10.  Ceramic Heads Decrease Metal Release Caused by Head-taper Fretting and Corrosion.

Authors:  Sevi B Kocagoz; Richard J Underwood; Daniel W MacDonald; Jeremy L Gilbert; Steven M Kurtz
Journal:  Clin Orthop Relat Res       Date:  2016-02-04       Impact factor: 4.176

View more
  14 in total

1.  Impending spontaneous head neck dissociation caused by anteverted cup: A case report on reverse bottle opener effect and review of literature.

Authors:  Ramneek Mahajan; Piyush Suresh Nashikkar; Varun Khanna
Journal:  J Clin Orthop Trauma       Date:  2020-11-01

2.  Do Well-functioning THAs Retrieved at Autopsy Exhibit Evidence of Fretting and Corrosion?

Authors:  Jeffrey Lange; Amanda Wach; Chelsea N Koch; Robert H Hopper; Henry Ho; Charles A Engh; Timothy M Wright; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2018-10       Impact factor: 4.176

Review 3.  Trunnion Corrosion in Total Hip Arthroplasty-Basic Concepts.

Authors:  Kenneth L Urish; Nicholas John Giori; Jack E Lemons; William M Mihalko; Nadim Hallab
Journal:  Orthop Clin North Am       Date:  2019-04-16       Impact factor: 2.472

4.  Blood cobalt ion level in patients with different sizes of cobalt/chrome femoral head with the Accolade TMZF stem.

Authors:  Seiya Ishii; Yasuhiro Homma; Takehisa Matsukawa; Tomonori Baba; Kazuo Kaneko; Muneaki Ishijima
Journal:  Int Orthop       Date:  2022-07-15       Impact factor: 3.479

5.  Diagnosis and Management of Adverse Reactions to Metal Debris.

Authors:  Richard A Wawrose; Kenneth L Urish
Journal:  Oper Tech Orthop       Date:  2019-10-01

Review 6.  Gross trunnion failure in metal on polyethylene total hip arthroplasty-a systematic review of literature.

Authors:  Tungish Bansal; Sameer Aggarwal; Mandeep Singh Dhillon; Sandeep Patel
Journal:  Int Orthop       Date:  2020-01-04       Impact factor: 3.075

7.  Comparison of Asymptomatic and Symptomatic Adverse Local Tissue Reaction in Patients With Head-Neck Taper Corrosion.

Authors:  Margaret A Weber; Matthew J Snyder; Kalain K Workman; Margaret M Sims; Clair N Smith; Deepak Kumar; Akshay V Daji; Camilo G Borrero; Andrew C Cordle; Anthony M DiGioia; Brian R Hamlin; Anton Y Plakseychuk; Kenneth L Urish
Journal:  J Arthroplasty       Date:  2020-09-17       Impact factor: 4.435

8.  What Is the Risk of THA Revision for ARMD in Patients with Non-metal-on-metal Bearings? A Study from the Australian National Joint Replacement Registry.

Authors:  R N de Steiger; Alesha Hatton; Yi Peng; Stephen Graves
Journal:  Clin Orthop Relat Res       Date:  2020-06       Impact factor: 4.755

Review 9.  What the Surgeon Can Do to Reduce the Risk of Trunnionosis in Hip Arthroplasty: Recommendations from the Literature.

Authors:  Claude B Rieker; Peter Wahl
Journal:  Materials (Basel)       Date:  2020-04-21       Impact factor: 3.623

10.  A case of bilateral hip mechanically assisted crevice corrosion after staged total hip arthroplasty.

Authors:  Carl L Herndon; Roshan P Shah; H John Cooper; Jeffrey A Geller
Journal:  Arthroplast Today       Date:  2018-06-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.