Literature DB >> 21903309

Influence of assembly procedure and material combination on the strength of the taper connection at the head-neck junction of modular hip endoprostheses.

Annelie Rehmer1, Nicholas E Bishop, Michael M Morlock.   

Abstract

BACKGROUND: A stable fixation between femoral head and endoprosthesis taper is necessary to prevent relative motions and corrosion at the taper junction. Although the importance of the component assembly has been recognised, no definitive instructions are available. The purpose of this study was to assess the influence of assembly force, assembly tool and number of hammer strokes on the taper junction strength of various material combinations.
METHODS: Co-Cr29-Mo (n=10) and Ti-6Al-4V (n=10) neck tapers were assembled with Co-Cr and Al(2)O(3) ceramic heads either by push-on or by impaction with single or multiple hammer blows. The strength of the taper-head connection was evaluated by measuring the head pull-off forces according to ISO 7206-10 and the turn-off moment capacity.
FINDINGS: The taper strength linearly increased with assembly forces (P<0.001). Co-Cr heads combined with Co-Cr tapers showed significantly lower pull-off forces and turn-off moments than the combination with Ti tapers (0.001<P<0.025). Multiple impaction did not increase taper strength (0.063 <P<0.995). Ceramic and Co-Cr heads showed similar fixation patterns on Ti tapers. Turn-off moments varied between 6 Nm and 19 Nm, dependent on material combination and assembly force.
INTERPRETATION: It is suggested that sufficient head-taper junction strength in all bearing conditions is achieved by impaction forces of at least 4 kN. A single impact is sufficient to achieve fixation. Special attention should be paid to the assembly of Co-Cr heads on Co-Cr tapers.
Copyright © 2011 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21903309     DOI: 10.1016/j.clinbiomech.2011.08.002

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  30 in total

Review 1.  What Factors Drive Taper Corrosion?

Authors:  Robin Pourzal; Hannah J Lundberg; Deborah J Hall; Joshua J Jacobs
Journal:  J Arthroplasty       Date:  2018-03-30       Impact factor: 4.757

2.  Modelling changes in modular taper micromechanics due to surgeon assembly technique in total hip arthroplasty.

Authors:  Jonathan A Gustafson; Robin Pourzal; Brett R Levine; Joshua J Jacobs; Hannah J Lundberg
Journal:  Bone Joint J       Date:  2020-07       Impact factor: 5.082

Review 3.  What is the trouble with trunnions?

Authors:  Christina I Esposito; Timothy M Wright; Stuart B Goodman; Daniel J Berry
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

4.  Contact conditions for total hip head-neck modular taper junctions with microgrooved stem tapers.

Authors:  Maren Bechstedt; Jonathan A Gustafson; Steven P Mell; Julian Gührs; Michael M Morlock; Brett R Levine; Hannah J Lundberg
Journal:  J Biomech       Date:  2020-02-25       Impact factor: 2.712

Review 5.  What Surgeons Need to Know About Adverse Local Tissue Reaction in Total Hip Arthroplasty.

Authors:  Deborah J Hall; Robin Pourzal; Joshua J Jacobs
Journal:  J Arthroplasty       Date:  2020-01-15       Impact factor: 4.757

6.  The effect of manufacturing tolerances on the mechanical environment of taper junctions in modular TKR.

Authors:  Kyle Snethen; Jorge Hernandez; Melinda Harman
Journal:  J Mech Behav Biomed Mater       Date:  2019-04-18

Review 7.  What are the current clinical issues in wear and tribocorrosion?

Authors:  Daniel J Berry; Matthew P Abdel; John J Callaghan
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

Review 8.  Corrosion of the Head-Stem Taper Junction-Are We on the Verge of an Epidemic?: Review Article.

Authors:  Michael Morlock; Dennis Bünte; Julian Gührs; Nicholas Bishop
Journal:  HSS J       Date:  2016-09-20

9.  Does Surface Topography Play a Role in Taper Damage in Head-neck Modular Junctions?

Authors:  Robin Pourzal; Deborah J Hall; Nguyen Q Ha; Robert M Urban; Brett R Levine; Joshua J Jacobs; Hannah J Lundberg
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

10.  Fixation strength of taper connection at head-neck junction in retrieved carbon fiber-reinforced PEEK hip stems.

Authors:  Ichiro Nakahara; Masaki Takao; Shunichi Bandoh; Nobuhiko Sugano
Journal:  J Artif Organs       Date:  2014-09-05       Impact factor: 1.731

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.