Literature DB >> 29881207

Wear performance of cobalt chromium, ceramic, and oxidized zirconium on highly crosslinked polyethylene at mid-term follow-up.

Matthew G Teeter1,2,3, Colin J MacLean1, Lyndsay E Somerville1, James L Howard1, Richard W McCalden1, Brent A Lanting1, Edward M Vasarhelyi1.   

Abstract

Patients with THA using a ceramic head and using an OxZr head were each matched to patients with a CoCr head. Mean implantation time was 5 years. There was no difference in steady state wear rate between the ceramic (0.066 ± 0.050 mm/year) and CoCr match groups (0.052 ± 0.041 mm/year), or between the OxZr (0.022 ± 0.029 mm/year) and CoCr match groups (0.048 ± 0.071 mm/year). Follow-up into the second decade will be necessary before any changes in THA wear rate from using ceramic or OxZr bearings may be appreciated clinically with available imaging techniques.

Entities:  

Keywords:  Ceramic; Cobalt chromium; Femoral head; Highly crosslinked polyethylene; Oxidized zirconium; Total hip arthroplasty; Wear

Year:  2018        PMID: 29881207      PMCID: PMC5990298          DOI: 10.1016/j.jor.2018.05.018

Source DB:  PubMed          Journal:  J Orthop        ISSN: 0972-978X


  26 in total

1.  Wear and Osteolysis of Highly Crosslinked Polyethylene at 10 to 14 Years: The Effect of Femoral Head Size.

Authors:  Paul F Lachiewicz; Elizabeth S Soileau; John M Martell
Journal:  Clin Orthop Relat Res       Date:  2016-02       Impact factor: 4.176

2.  Determination of polyethylene wear in total hip replacements with use of digital radiographs.

Authors:  J M Martell; S Berdia
Journal:  J Bone Joint Surg Am       Date:  1997-11       Impact factor: 5.284

Review 3.  What is the trouble with trunnions?

Authors:  Christina I Esposito; Timothy M Wright; Stuart B Goodman; Daniel J Berry
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

4.  Total hip arthroplasty using highly cross-linked polyethylene in patients younger than 50 years with minimum 10-year follow-up.

Authors:  Nikola Babovic; Robert T Trousdale
Journal:  J Arthroplasty       Date:  2013-03-15       Impact factor: 4.757

5.  Clinical Impact of Trunnion Wear After Total Hip Arthroplasty.

Authors:  Brent Lanting; Douglas D R Naudie; Richard W McCalden
Journal:  JBJS Rev       Date:  2016-08-23

Review 6.  Clinical and hip simulator comparisons of ceramic-on-polyethylene and metal-on-polyethylene wear.

Authors:  I C Clarke; A Gustafson
Journal:  Clin Orthop Relat Res       Date:  2000-10       Impact factor: 4.176

7.  Highly cross-linked polyethylene decreases the rate of revision of total hip arthroplasty compared with conventional polyethylene at 13 years' follow-up.

Authors:  S A Hanna; L Somerville; R W McCalden; D D Naudie; S J MacDonald
Journal:  Bone Joint J       Date:  2016-01       Impact factor: 5.082

8.  Mid-term survivorship and clinical outcomes of cobalt-chrome and oxidized zirconium on highly crosslinked polyethylene.

Authors:  Stephen M Petis; Edward M Vasarhelyi; Brent A Lanting; James L Howard; Douglas D R Naudie; Lyndsay E Somerville; Richard W McCalden
Journal:  Can J Surg       Date:  2016-02       Impact factor: 2.089

9.  Wear rate of highly cross-linked polyethylene in total hip arthroplasty. A randomized controlled trial.

Authors:  Richard W McCalden; Steven J MacDonald; Cecil H Rorabeck; Robert B Bourne; David G Chess; Kory D Charron
Journal:  J Bone Joint Surg Am       Date:  2009-04       Impact factor: 5.284

Review 10.  Fracture of ceramic bearing surfaces following total hip replacement: a systematic review.

Authors:  Francesco Traina; Marcello De Fine; Alberto Di Martino; Cesare Faldini
Journal:  Biomed Res Int       Date:  2013-06-13       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.