Literature DB >> 21828152

Reproducibility of cornea measurements in anterior segment OCT images of normal eyes and eyes with bullous keratopathy analyzed with the Zhongshan Assessment Program.

Joy B Chan1, Leonard H Yuen, Elaine H Huang, Hla M Htoon, Mingguang He, Tin Aung, Donald T Tan, Jodhbir S Mehta.   

Abstract

PURPOSE: To determine the interobserver and intraobserver measurement reproducibility of cornea parameters of both normal eyes and eyes with bullous keratopathy (BK) obtained with the Zhongshan Assessment Program (ZAP) on anterior segment optical coherence tomography (AS-OCT) images.
METHODS: A comparative study was carried out on 24 healthy volunteers and 25 subjects with BK. AS-OCT images were independently analyzed by two examiners. Parameters examined: anterior chamber depth (ACD), central corneal thickness (CCT), posterior corneal curvature (PCC), and posterior corneal arc length (PCAL). Interobserver and intraobserver reproducibility of these parameters was calculated in terms of limits of agreement (mean of differences ± 1.96SD of differences).
RESULTS: In the normal group, both horizontal and vertical ACD were successfully measured in 23 of 24 (96%) images. The mean bias for two measurements by two different observers ranged from 0.003 to 0.117 mm for ACD, PCC, and PCAL measurements and from 0.013 to 2.25 μm for CCT measurements, and there were no differences between the two observers (P > 0.05). Mean bias for two measurements by the same grader ranged from 0.005 to 0.327 mm for ACD, PCC, and PCAL measurements and 1.46 to 2.53 μm for CCT measurements. There was no difference between the two observations (P > 0.05). Similar results were found in the BK group.
CONCLUSIONS: There was high inter- and intraobserver reproducibility for normal and pathologic corneas using the ZAP software. The ZAP software may serve as a new investigatory tool for accurately evaluating the anterior segment and corneal parameters for corneal procedures.

Entities:  

Mesh:

Year:  2011        PMID: 21828152     DOI: 10.1167/iovs.10-6411

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  11 in total

1.  Association between baseline angle width and induced angle opening following prophylactic laser peripheral iridotomy.

Authors:  Roland Y Lee; Toshimitsu Kasuga; Qi N Cui; Guofu Huang; Mingguang He; Shan C Lin
Journal:  Invest Ophthalmol Vis Sci       Date:  2013-05-01       Impact factor: 4.799

2.  Comparison of anterior segment morphology following prophylactic laser peripheral iridotomy in Caucasian and Chinese eyes.

Authors:  Roland Y Lee; Toshimitsu Kasuga; Qi N Cui; Guofu Huang; Mingguang He; Shan C Lin
Journal:  Clin Exp Ophthalmol       Date:  2013-10-29       Impact factor: 4.207

3.  Corneal flap morphological analysis using anterior segment optical coherence tomography in laser in situ keratomileusis with femtosecond lasers versus mechanical microkeratome.

Authors:  Xiao-Xiao Zhang; Xing-Wu Zhong; Jun-Shu Wu; Zheng Wang; Ke-Ming Yu; Quan Liu; Bin Yang
Journal:  Int J Ophthalmol       Date:  2012-02-18       Impact factor: 1.779

4.  Development of a novel in vivo corneal fibrosis model in the dog.

Authors:  K M Gronkiewicz; E A Giuliano; K Kuroki; F Bunyak; A Sharma; L B C Teixeira; C W Hamm; R R Mohan
Journal:  Exp Eye Res       Date:  2015-10-09       Impact factor: 3.467

5.  Association between baseline iris thickness and prophylactic laser peripheral iridotomy outcomes in primary angle-closure suspects.

Authors:  Roland Y Lee; Toshimitsu Kasuga; Qi N Cui; Travis C Porco; Guofu Huang; Mingguang He; Shan C Lin
Journal:  Ophthalmology       Date:  2014-02-15       Impact factor: 12.079

6.  Patency of small laser iridotomy evaluated using anterior-segment optical coherence tomography.

Authors:  Seigo Nakabayashi; Motofumi Kawai; Toru Yamaguchi; Akitoshi Yoshida
Journal:  Clin Ophthalmol       Date:  2014-03-21

7.  Comparison of anterior segment optical tomography parameters measured using a semi-automatic software to standard clinical instruments.

Authors:  Marcus Ang; Wesley Chong; Huiqi Huang; Wan Ting Tay; Tien Yin Wong; Ming-Guang He; Tin Aung; Jodhbir S Mehta
Journal:  PLoS One       Date:  2013-06-04       Impact factor: 3.240

8.  Reproducibility and age-related changes of ocular parametric measurements in rabbits.

Authors:  Andri K Riau; Neil Y S Tan; Romesh I Angunawela; Hla M Htoon; Shyam S Chaurasia; Jodhbir S Mehta
Journal:  BMC Vet Res       Date:  2012-08-19       Impact factor: 2.741

9.  Determinants of posterior corneal biometric measurements in a multi-ethnic Asian population.

Authors:  Marcus Ang; Wesley Chong; Huiqi Huang; Tien Yin Wong; Ming-Guang He; Tin Aung; Jodhbir S Mehta
Journal:  PLoS One       Date:  2014-07-09       Impact factor: 3.240

10.  Cross-sectional, observational study of anterior segment parameters using anterior segment optical coherence tomography in North Indian Population.

Authors:  Latika Khatri Dalal; Renu Dhasmana; Amit Maitreya
Journal:  Ann Afr Med       Date:  2017 Jul-Sep
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.