Wear testing and particle characterisation of sequentially crosslinked polyethylene acetabular liners using different femoral head sizes.

Carmen Zietz1, Christian Fabry, Lars Middelborg, Gerhard Fulda, Wolfram Mittelmeier, Rainer Bader.   

Abstract

Larger femoral heads lead to a decreased risk of total hip dislocation and an improved range of motion. However, the larger diameter is associated with increased wear rates. The low wear rates of crosslinked polyethylene opens up the possibility of using larger heads. The aim of this experimental study was to evaluate the wear of conventional non-crosslinked versus sequentially crosslinked polyethylene liners in combination with different ceramic head sizes (28, 36, 44 mm). Wear testing was performed in a hip simulator according to ISO 14242. Wear particles from the polyethylene liners were characterized after wear testing. The wear measurements revealed a significant increase in the wear of crosslinked polyethylene liners with larger heads. By sequential crosslinking, however, the gravimetric wear using larger heads was reduced to a fractional amount of the wear using conventional polyethylene. Significant differences were observed for particle morphology but not for the number of particles when comparing non-crosslinked and crosslinked polyethylene.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23615788     DOI: 10.1007/s10856-013-4936-z

Source DB:  PubMed          Journal:  J Mater Sci Mater Med        ISSN: 0957-4530            Impact factor:   3.896


  47 in total

1.  Third-body wear of highly cross-linked polyethylene in a hip simulator.

Authors:  Charles R Bragdon; Murali Jasty; Orhun K Muratoglu; Daniel O O'Connor; William H Harris
Journal:  J Arthroplasty       Date:  2003-08       Impact factor: 4.757

2.  Reasons for revision hip surgery: a retrospective review.

Authors:  John C Clohisy; George Calvert; Frank Tull; Douglas McDonald; William J Maloney
Journal:  Clin Orthop Relat Res       Date:  2004-12       Impact factor: 4.176

3.  Fatigue crack propagation resistance of virgin and highly crosslinked, thermally treated ultra-high molecular weight polyethylene.

Authors:  Sara J Gencur; Clare M Rimnac; Steven M Kurtz
Journal:  Biomaterials       Date:  2005-11-21       Impact factor: 12.479

4.  Particle analysis for the determination of UHMWPE wear.

Authors:  M Scott; M Morrison; S R Mishra; S Jani
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2005-05       Impact factor: 3.368

5.  Evaluation of sequentially crosslinked ultra-high molecular weight polyethylene.

Authors:  M L Morrison; S Jani
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2009-07       Impact factor: 3.368

6.  Retrieval analyses of highly cross-linked polyethylene acetabular liners four and five years after implantation.

Authors:  K Knahr; M Pospischill; P Köttig; W Schneider; H Plenk
Journal:  J Bone Joint Surg Br       Date:  2007-08

7.  Cross-linked ultra-high-molecular weight polyethylene liner and ceramic femoral head in total hip arthroplasty: a prospective study at 5 years follow-up.

Authors:  Iñigo Orradre Burusco; Raquel Romero; Miriam Brun; Jose Javier López Blasco
Journal:  Arch Orthop Trauma Surg       Date:  2011-07-08       Impact factor: 3.067

8.  Unexpected failure of highly cross-linked polyethylene acetabular liner.

Authors:  Warit Waewsawangwong; Stuart B Goodman
Journal:  J Arthroplasty       Date:  2011-05-20       Impact factor: 4.757

9.  Effect of femoral head size on polyethylene wear and synovitis after total hip arthroplasty: a sonographic and radiographic study of 39 patients.

Authors:  Sarunas Tarasevicius; Otto Robertsson; Uldis Kesteris; Romas Jonas Kalesinskas; Hans Wingstrand
Journal:  Acta Orthop       Date:  2008-08       Impact factor: 3.717

10.  5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets: two randomized studies using radiostereometric analysis.

Authors:  Georgios Digas; Johan Kärrholm; Jonas Thanner; Peter Herberts
Journal:  Acta Orthop       Date:  2007-12       Impact factor: 3.717

View more
  2 in total

1.  Quantifying wear depth in hip prostheses using a 3D optical scanner.

Authors:  Katherine A Hollar; Daniel S Ferguson; John B Everingham; Jillian L Helms; Kevin J Warburton; Trevor J Lujan
Journal:  Wear       Date:  2017-10-16       Impact factor: 3.892

2.  Influence of Accelerated Aging on the Wear Behavior of Cross-Linked Polyethylene Liners-A Hip Simulator Study.

Authors:  Rebecca H Dammer; Carmen Zietz; Jörn Reinders; Michael Teske; Rainer Bader
Journal:  J Clin Med       Date:  2022-08-01       Impact factor: 4.964

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.