Using GafChromic film to estimate the effective dose from dental cone beam CT and panoramic radiography.

A Al-Okshi1, M Nilsson, A Petersson, M Wiese, C Lindh.   

Abstract

OBJECTIVES: To demonstrate the feasibility of GafChromic(®) XR-QA2 (ISP Corp., Wayne, NJ) as a dosemeter when performing measurements of the effective dose from three cone beam CT (CBCT) units and to compare the doses from examinations of three common dental clinical situations. A second aim was to compare the radiation doses for three digital panoramic units with the doses for the CBCT units.
METHODS: The CBCT units used were Veraviewepocs 3De(®) (J Morita MFG Corp., Kyoto, Japan), ProMax(®) 3D (Planmeca, Helsinki, Finland) and NewTom VGi(®) (Quantitative Radiology, Verona, Italy). GafChromic XR-QA2 films were placed between the selected layers of the head and neck of a tissue-equivalent human skull (RANDO(®) phantom; The Phantom Laboratory, Salem, NY). The exposure parameters were set using the automatic exposure control function of the units. Depending on the availability, medium and smaller field of view (FOV) scanning modes were used. The effective dose was estimated using the 2007 International Commission on Radiological Protection formalism.
RESULTS: The lowest effective dose of a CBCT unit was observed for ProMax 3D, FOV 4 × 5 cm (10 μSv), the highest for NewTom VGi, FOV 8 × 8 cm-high resolution (129 μSv). The range of effective doses for digital panoramic machines measured was 8-14 μSv.
CONCLUSIONS: This study demonstrates the feasibility of using radiochromic films for dental CBCT and panoramic dosimetry.

Entities:  

Keywords:  cone beam computed tomography; film dosimetry; radiation dosage

Mesh:

Year:  2013        PMID: 23610090      PMCID: PMC3699970          DOI: 10.1259/dmfr.20120343

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  18 in total

1.  Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CAT.

Authors:  J B Ludlow; L E Davies-Ludlow; S L Brooks; W B Howerton
Journal:  Dentomaxillofac Radiol       Date:  2006-07       Impact factor: 2.419

2.  Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry.

Authors:  L Paelinck; W De Neve; C De Wagter
Journal:  Phys Med Biol       Date:  2006-12-18       Impact factor: 3.609

3.  Measurement of energy dependence for XRCT radiochromic film.

Authors:  Martin J Butson; Tsang Cheung; Peter K N Yu
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

4.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

5.  Radiochromic films for dental CT dosimetry: a feasibility study.

Authors:  O Rampado; S D Bianchi; A Peruzzo Cornetto; V Rossetti; R Ropolo
Journal:  Phys Med       Date:  2012-06-30       Impact factor: 2.685

6.  The evolution and application of dental maxillofacial imaging modalities.

Authors:  Stuart C White; Michael J Pharoah
Journal:  Dent Clin North Am       Date:  2008-10

7.  Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology.

Authors:  John B Ludlow; Marija Ivanovic
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2008-05-27

8.  Computed tomography dose measurements with radiochromic films and a flatbed scanner.

Authors:  O Rampado; E Garelli; R Ropolo
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

9.  Patient risk related to common dental radiographic examinations: the impact of 2007 International Commission on Radiological Protection recommendations regarding dose calculation.

Authors:  John B Ludlow; Laura E Davies-Ludlow; Stuart C White
Journal:  J Am Dent Assoc       Date:  2008-09       Impact factor: 3.634

10.  A study of GafChromic XR Type R film response with reflective-type densitometers and economical flatbed scanners.

Authors:  G Thomas; R Y L Chu; Frank Rabe
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

View more
  12 in total

1.  Radiation dose from X-ray examinations of impacted canines: cone beam CT vs two-dimensional imaging.

Authors:  Nils Kadesjö; Randi Lynds; Mats Nilsson; Xie-Qi Shi
Journal:  Dentomaxillofac Radiol       Date:  2018-01-16       Impact factor: 2.419

Review 2.  Effective dose of cone beam CT (CBCT) of the facial skeleton: a systematic review.

Authors:  A Al-Okshi; C Lindh; H Salé; M Gunnarsson; M Rohlin
Journal:  Br J Radiol       Date:  2015-01       Impact factor: 3.039

3.  Comparison of the accuracy of full head cone beam CT images obtained using a large field of view and stitched images.

Authors:  Mehmet Ozgur Ozemre; Ayse Gulsahi
Journal:  Dentomaxillofac Radiol       Date:  2018-06-19       Impact factor: 2.419

4.  Effective dose reduction using collimation function in digital panoramic radiography and possible clinical implications in dentistry.

Authors:  Daniel Benchimol; Juha Koivisto; Nils Kadesjö; Xie-Qi Shi
Journal:  Dentomaxillofac Radiol       Date:  2018-05-31       Impact factor: 2.419

5.  Presurgical evaluation of bony implant sites using panoramic radiography and cone beam computed tomography-influence of medical education.

Authors:  Michael Dau; Amir Edalatpour; Ralf Schulze; Bilal Al-Nawas; Abdulmonem Alshihri; Peer W Kämmerer
Journal:  Dentomaxillofac Radiol       Date:  2016-10-19       Impact factor: 2.419

6.  Thyroid Radiation Dose to Patients from Diagnostic Radiology Procedures over Eight Decades: 1930-2010.

Authors:  Lienard A Chang; Donald L Miller; Choonsik Lee; Dunstana R Melo; Daphnée Villoing; Vladimir Drozdovitch; Isabelle Thierry-Chef; Sarah J Winters; Michael Labrake; Charles F Myers; Hyeyeun Lim; Cari M Kitahara; Martha S Linet; Steven L Simon
Journal:  Health Phys       Date:  2017-12       Impact factor: 1.316

Review 7.  Effective dose of dental CBCT-a meta analysis of published data and additional data for nine CBCT units.

Authors:  J B Ludlow; R Timothy; C Walker; R Hunter; E Benavides; D B Samuelson; M J Scheske
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

8.  Effective dose estimation in cone-beam computed tomography for dental use by Monte-Carlo simulation optimizing calculation numbers using a step-and-shoot method.

Authors:  Yoshihiro Ozaki; Hiroshi Watanabe; Tohru Kurabayashi
Journal:  Dentomaxillofac Radiol       Date:  2021-04-30       Impact factor: 3.525

Review 9.  Indications for 3-D diagnostics and navigation in dental implantology with the focus on radiation exposure: a systematic review.

Authors:  Burkhard Kunzendorf; Hendrik Naujokat; Jörg Wiltfang
Journal:  Int J Implant Dent       Date:  2021-05-27

10.  Fully customized placement of orthodontic miniplates: a novel clinical technique.

Authors:  Jan Hourfar; Georgios Kanavakis; Peter Goellner; Björn Ludwig
Journal:  Head Face Med       Date:  2014-05-03       Impact factor: 2.151

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.